Porges Coloplast AA6C18Recalls & reported problems
Coloplast A/S · FOLYSIL® catheter straight tip CH FR 18 · Class 2
All models of Porges Coloplast (79)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03600040131992), received between Dec 18, 2018 and Mar 13, 2025 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 2, 2024Class IIOpen, Classified
A possible sterility issue was detected in Coloplast's facility on some Coloplast products.
Z-0844-2025 · Root cause: Packaging · Recall details · FDA record ↗
Most reported problems
Device problems
- Burst Container or Vessel2
- Difficult to Remove1
- Failure to Deflate1
- Material Fragmentation1
- Material Too Rigid or Stiff1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 13, 2025Material Fragmentation; Material Fragmentation
ACCORDING TO THE AVAILABLE INFORMATION, THE CATHETER TIP CAME LOOSE PRIOR TO CATHETERIZATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21600525 ↗MalfunctionJun 15, 2024Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION A CATHETER WAS INSERTED AND SHORTLY AFTER, THE CATHETER CAME OUT BECAUSE THE BALLOON WAS PIERCED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19543958 ↗InjuryDec 17, 2021Difficult to Remove; Material Too Rigid or Stiff; Failure to Deflate; Difficult to Remove; Material Too Rigid or Stiff; Failure to Deflate
ACCORDING TO THE AVAILABLE INFORMATION, THE NURSES WERE UNABLE TO DEFLATE THE BALLOON AND HAD TO SEND THE PATIENT TO THE HOSPITAL "A FEW TIMES" TO GET THE CATHETER REMOVED. THE BALLOON WAS HARD AND IT APPEARED THAT THE SOLUTION INSIDE HAD CRYSTALIZED. IT WAS UNCLEAR WHAT THE NURSES USED TO INFLATE IT, BUT THE DOCTOR WHO REMOVED THE CATHETER THOUGHT MAYBE NORMAL SALINE HAD BEEN USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13025530 ↗InjuryDec 18, 2018Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, THE CATHETER FELL OFF DUE TO BALLOON BURST. THE PHYSICIAN REPLACED THE DEVICE WITH A NEW CATHETER IMMEDIATELY, NO CLINICAL CONSEQUENCES REPORTED. NO RESIDUE OR PARTICLE REMAINS IN PATIENT'S BLADDER.
Patient problems: No Information
MAUDE report 8174798 ↗
Other CATHETER, RETENTION TYPE, BALLOON devices
- Bardex® All-Silicone Foley Catheter 165816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON512 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON476 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI16C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON1 death383 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON234 reports
- Bardex® Lubri-Sil® All-Silicone Temp Sensing Foley Catheter 119216MC. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON167 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175814C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON153 reports
- Bardex® All-Silicone Foley Catheter 165812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON151 reports
- SureStep™ Foley Tray System Bard® Lubri-Sil® Foley Catheter Tray A942216C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON2 recalls46 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page