STA R Max® (with cap piercing option) 59010Recalls & reported problems
Diagnostica Stago · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03607450590104), received between Oct 4, 2019 and Oct 16, 2019 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Use of Incorrect Control/Treatment Settings” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K082675, K093001, K151867, K983460); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Use of Incorrect Control/Treatment Settings1
Patient problems
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 16, 2019Use of Incorrect Control/Treatment Settings; Use of Incorrect Control/Treatment Settings
A CUSTOMER (B)(6)) CALLED THE (B)(6) ON (B)(6) 2019 TO REPORT A BIAS OF -100% ON STAGO 1 (S/N (B)(4)) VS. STAGO 2 (S/N (B)(4)) FOR PROTEIN C TESTING. THE PC RESULTS FROM (B)(6) 2018 UNTIL (B)(6) 2019 ON STAGO 1 (S/N (B)(4)) WERE RECOVERING APPROXIMATELY HALF OF EXPECTED LEVELS DUE TO AN INCORRECT DILUTION FACTOR SETTING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9195643 ↗InjuryOct 4, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
A CUSTOMER (CPL - SONIC) CALLED THE STAGO NORTH AMERICA HOTLINE ON (B)(6) 2019 TO REPORT A BIAS OF -100% ON STAGO 1 (S/N (B)(4)) VS. STAGO 2 (S/N (B)(4)) FOR PROTEIN C TESTING. THE PC-20 RESULTS FROM 19JUL2018 UNTIL 06SEP2019 ON STAGO 1 (S/N (B)(4)) WERE RECOVERING APPROXIMATELY HALF OF EXPECTED LEVELS DUE TO AN INCORRECT DILUTION FACTOR SETTING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9158233 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082675, K093001, K151867, K983460. One submission often covers many models, so check the product and lot details in each record.
- Z-1575-2023 · Mar 22, 2023 · Open, ClassifiedSTA R Max fully automatic clinical instrument designed to perform tests on human plasmas, Reference Numbers 59013 and 59026
- Z-0899-03 · Mar 5, 2003 · TerminatedSTA-Compact Hemostasis System with Cap piercing capability.
Other SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES devices
- ACL TOP 970 CL 00000280097Instrumentation Laboratory Company · System, Multipurpose For In Vitro Coagulation Studies2 recalls0 reports
- TEG5000 HEMOSTASIS ANALYZER 07-022Haemonetics Corporation · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall2 reports
- STA Satellite® 58104Diagnostica Stago · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall1 report
- ACL ELITE 00000880311Instrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- ACL ELITE PRO 00001000311RInstrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- ACL ELITE PRO 00001000311Instrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- ACL ELITE 00000880311RInstrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- Hemochron Whole Blood Microcoagulation System ELITEAccriva Diagnostics Holdings, Inc. · System, Multipurpose For In Vitro Coagulation Studies2 deaths13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page