C-CURVE B20520015Recalls & reported problems
Medicrea International · C-Curve Screws Ø3.5mm x L15mmX2 · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 03613720275688), received between Jun 5, 2026 and Jun 8, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Use of Device Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 8, 2026Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING ACDF (ANTERIOR CERVICAL DISCECTOMY AND FUSION) PROCEDURE. IT WAS REPORTED THAT THE SCREW WAS BROKEN. NO FRAGMENTS OF THE IMPLANT OR INSTRUMENT REMAINED IN THE PATIENT. THE PRODUCT WAS EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE SCREW BREAK HAPPENED DURING IMPLANTATION ITSELF AND THE PRODUCT EXPLANTED ON THE SAME DAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25432123 ↗MalfunctionJun 8, 2026Break; Use of Device Problem
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT DURING A C5-C6 DECOMPRESSION AND FUSION PROCEDURE, ONE OF THE SCREWS BROKE INTRA-OPERATIVELY AND ENDED UP IN THE DISC SPACE. THIS REQUIRED ADDITIONAL PROCEDURE TIME TO REMOVE THE BROKEN SCREW AND REPLACE ALL IMPLANTS WITH NEW ONES. THE PROCEDURE INVOLVED NOT USING THE INSERTER AS A GUIDE FOR DRILLING THE SCREW HOLES, CONTRARY TO OFFICIAL RECOMMENDATIONS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25430982 ↗MalfunctionJun 8, 2026Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT THE FIRST SCREW - THREAD INSIDE SCREW (IN SLOT WHERE SCREWDRIVER INSERTS) THREAD CAME OFF IN LITTLE METAL PIECES AND WAS UNABLE TO ADVANCE SCREW ANY FURTHER AND ALSO UNABLE TO REMOVE SCREW (SCREW AND CAGE AS IT WAS NOW FIXED INSIDE). THE SECOND SCREW ON SECOND LEVEL, THE SCREW HEAD BROKE OFF WITH THE THREADED SHAFT IN PATIENT AND HEAD PART BROKE OFF DURING IMPLANTATION. SURGEON WAS ABLE TO REMOVE HEAD PART AND WAS UNABLE TO ADVANCE SCREW ANY FURTHER AND ALSO UNABLE TO REMOVE SCREW (SCREW AND CAGE AS IT WAS NOW FIXED INSIDE). CAGES IN PLACE BUT NOT AS FULL SEATED AS SURGEON WOULD PREFER. THERE WAS NO CONFIRMED PLAN FOR A REVISION SURGERY TO EXPLANT THE BROKEN SCREWS. THE PROCEDURE PERFORMED WAS A ANTERIOR CERVICAL DISCECTOMY AND FUSION. THE IMPLANTS WERE PLACED IN LEVELS C5¿C6 AND C6-C7. THERE WERE NO PATIENT SYMPTOMS REPORTED. NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25424384 ↗MalfunctionJun 5, 2026Break; Use of Device Problem
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT DURING A C5-C6 DECOMPRESSION AND FUSION PROCEDURE, ONE OF THE SCREWS BROKE INTRA-OPERATIVELY AND ENDED UP IN THE DISC SPACE. THIS REQUIRED ADDITIONAL PROCEDURE TIME TO REMOVE THE BROKEN SCREW AND REPLACE ALL IMPLANTS WITH NEW ONES. THE PROCEDURE INVOLVED NOT USING THE INSERTER AS A GUIDE FOR DRILLING THE SCREW HOLES, CONTRARY TO OFFICIAL RECOMMENDATIONS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25414028 ↗MalfunctionJun 5, 2026Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT THE FIRST SCREW - THREAD INSIDE SCREW (IN SLOT WHERE SCREWDRIVER INSERTS) THREAD CAME OFF IN LITTLE METAL PIECES AND WAS UNABLE TO ADVANCE SCREW ANY FURTHER AND ALSO UNABLE TO REMOVE SCREW (SCREW AND CAGE AS IT WAS NOW FIXED INSIDE). THE SECOND SCREW ON SECOND LEVEL, THE SCREW HEAD BROKE OFF WITH THE THREADED SHAFT IN PATIENT AND HEAD PART BROKE OFF DURING IMPLANTATION. SURGEON WAS ABLE TO REMOVE HEAD PART AND WAS UNABLE TO ADVANCE SCREW ANY FURTHER AND ALSO UNABLE TO REMOVE SCREW (SCREW AND CAGE AS IT WAS NOW FIXED INSIDE). CAGES IN PLACE BUT NOT AS FULL SEATED AS SURGEON WOULD PREFER. THERE WAS NO CONFIRMED PLAN FOR A REVISION SURGERY TO EXPLANT THE BROKEN SCREWS. THE PROCEDURE PERFORMED WAS A ANTERIOR CERVICAL DISCECTOMY AND FUSION. THE IMPLANTS WERE PLACED IN LEVELS C5¿C6 AND C6-C7. THERE WERE NO PATIENT SYMPTOMS REPORTED. NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25413756 ↗MalfunctionJun 5, 2026Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. PRE-OP DIAGNOSIS OF PATIENT WAS NON-SPECIFIC NEURAL COMPRESSION (NSNC). IT WAS REPORTED THAT, THIS WAS A REVISION OF A FAILED SURGERY. TO PERFORM THE PROCEDURE ANTERIOR CERVICAL DISCECTOMY AND FUSION, SURGEON HAD TO REMOVE THE TWO PREVIOUS C-CURVE IMPLANTS, CARRY OUT A DOUBLE CORPECTOMY, AND IMPLANT A MOS MESH. IT WAS IDENTIFIED THAT THERE WAS A PROBLEM WITH THE C-CURVE SCREW DURING THE REMOVAL OF THE IMPLANT. LEVELS IMPLANTED DURING INITIAL SURGERY WAS C3-C4 AND C6-C7. THERE WAS NO FRAGMENT OF BROKEN SCREWS LEFT INSIDE PATIENT BODY. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. ADDITIONAL INFORMATION RECEIVED STATING A WEEK AFTER THE FIRST INTERVENTION, THE PATIENT STILL COMPLAINED OF MYELOPATHY. THE FIRST SURGERY DID NOT SOLVE THE PATIENT'S PATHOLOGY. THE MAIN CAUSE OF THE NON-FUSION WAS TIME. ONLY 1 MONTH HAS PASSED BETWEEN THE FIRST AND THE SECOND INTERVENTION. THEREFORE, THE REASON FOR REOPERATION WAS MAINLY A PROBLEM WITH THE INDICATION AND SURGERY, NOT WITH THE IMPLANTED MATERIAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25412539 ↗
Other Intervertebral Fusion Device With Integrated Fixation, Cervical devices
- NA 03.617.902Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical103 reports
- NA 03.617.900Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical77 reports
- Divergence® Anterior Cervical Fusion System G6626525Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical75 reports
- Divergence® Anterior Cervical Fusion System G6626526Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical49 reports
- NA 03.617.905Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical29 reports
- NA 03.617.904Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical27 reports
- Divergence® Anterior Cervical Fusion System G6626746Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical21 reports
- NA 03.647.972Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page