Skip to content

ROI-C MC1442PRecalls & reported problems

LDR Spine Usa, Inc. · ROI-C Lordotic implant H6mm 12x14mm · Class 2

All models of ROI-C (39)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03662663016537), received between Jul 15, 2019 and Dec 3, 2025 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Break2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Code Available2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 3, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGERY TO EXTEND A CONSTRUCT, WHILE PUTTING THE ANCHORING PLATE IN, AN ROI-C IMPLANT BROKE WHERE IT CONNECTS TO THE INSERTER. THE SURGERY WAS COMPLETED WITH ANOTHER ROI-C IMPLANT WITHOUT ANY PATIENT OR PROCEDURAL COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23707083 ↗
  • MalfunctionMay 10, 2023Fracture; Fracture

    IT WAS REPORTED THAT AN ROI-C CAGE FRACTURED IN THE C2/3 DISC SPACE INTRA-OPERATIVELY WHILE INSTALLING THE SECOND PLATE. THE CAGE AND PLATES WERE REMOVED AND REPLACED WITH A NEW CAGE AND PLATES. THE FRACTURED PIECES OF THE CAGE WERE RECOVERED. THERE WERE NO PATIENT IMPACTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16903809 ↗
  • MalfunctionMar 31, 2023Fracture; Fracture

    IT WAS REPORTED THAT THE ANCHORING PLATE WOULD NOT ADVANCE THROUGH THE PATIENT'S BONE AT C4 OF A C4/5 ROI-C CONSTRUCT INTRAOPERATIVELY. WHILE REMOVING THE PLATE, THE CAGE CRACKED. BOTH COMPONENTS WERE REMOVED AND REPLACED WITH NEW ONES TO COMPLETE THE PROCEDURE. THE SURGEON SAID THE PATIENT HAD HARD, SCLEROTIC BONE SO HE USED THE AWL AT THAT LEVEL PRIOR TO RE-INSTALLING THE PLATES. THERE WERE NO PATIENT IMPACTS, BUT THERE WAS A DELAY OF 20 MINUTES TO REMOVE AND THE CAGE AND PLATE. THIS IS REPORT TWO OF TWO FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16658497 ↗
  • InjuryJul 15, 2019Break; Break

    ROI-C: 3 BROKEN CAGES AND BENT AWL. ACCORDING TO THE AVAILABLE DESCRIPTION , THE SURGEON WAS IMPACTING ANCHORING PLATE WITH IMPACTOR 1 THEN 2 AT C5-6 WHEN PEEK CAGE CRACKED (12X14 H6). HE REMOVED THE ANCHORING PLATES AND PEEK DEBRIS. HE PROCEEDED TO TRY IT AGAIN WITH A NEW PEEK CAGE AND PLATES (12 X14 H6) AND AGAIN THE PEEK CRACKED JUST LIKE BEFORE. HE REMOVED THE ANCHORING PLATES AND PEEK DEBRIS. THE REPORTER SUGGESTED THE AWL AND USING A NEW INSERTER AND HE AGREED TO DO BOTH. AFTER USING THE AWL AND NEW INSERTER HE IMPLANTED THE NEW PEEK AND PLATE (12X14 H6) CRACKED AGAIN FOLLOWING IMPACTOR 1 & 2. THE AWL WAS ALSO BENT AT THE TIP AFTER IMPACTING. SURGERY DELAYED 1 HOUR. NO PATIENT IMPACT OR SERIOUS INJURY WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED. INVESTIGATION ONGOING.

    Patient problems: No Code Available; No Known Impact Or Consequence To Patient

    MAUDE report 8788791 ↗
  • InjuryJul 15, 2019Break; Break

    IT WAS REPORTED THAT 3 CAGES BROKE WHILE IMPACTING THE ANCHORING PLATE. THE SURGERY WAS DELAYED OVER 30 MINUTES TO REMOVE AND REPLACE THE DEVICES. THIS IS REPORT 1 OF 3.

    Patient problems: No Code Available; No Known Impact Or Consequence To Patient

    MAUDE report 8788745 ↗

Other Intervertebral Fusion Device With Integrated Fixation, Cervical devices

All Intervertebral Fusion Device With Integrated Fixation, Cervical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page