ROI-C MC9001RRecalls & reported problems

LDR Spine Usa, Inc. · ROI-C IMPLANT HOLDER · Class 2

All models of ROI-C (39)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 03662663043168), received between Dec 3, 2020 and May 28, 2025 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Fracture” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture11
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Foreign Body In Patient1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 28, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING AN ROI-C PROCEDURE, WHILE TRYING TO DISENGAGE THE IMPLANT FROM THE IMPLANT HOLDER, THE HOOK ON THE IMPLANT HOLDER BROKE OFF AND REMAINED IN THE CAGE. THEY TRIED TO REMOVE IT, BUT THERE WASN'T ENOUGH SPACE OR GRIP TO GRAB IT. SO THEY DECIDED TO LEAVE IT IN PLACE AND CLOSE THE PATIENT. THE HOOK TIP REMAINS STUCK IN THE CERVICAL CAGE BETWEEN VERTEBRAE.

    Patient problems: Foreign Body In Patient

    MAUDE report 22098466 ↗
  • MalfunctionDec 11, 2024Fracture; Fracture

    A HOSPITAL REPORTED THAT THE SMALL HOOK ON A ROI-C IMPLANT HOLDER TWISTED AND EVENTUALLY BROKE DURING SURGERY. THERE WAS NO REPORTED PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20903880 ↗
  • MalfunctionNov 5, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE HOOK AT THE TIP OF A ROI-C IMPLANT HOLDER FRACTURED INTRA-OPERATIVELY WHEN THE DOCTOR IMPACTED THE CAGE AND WHEN HE TRIED TO REMOVE THE PIECE THAT LOCKS THE CAGE. IT IS NOT BELIEVED THAT THE PIECE WAS RETAINED IN THE PATIENT SINCE IT DID NOT SHOW UP IN ANY X-RAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20610154 ↗
  • MalfunctionJun 5, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE HOOK TIP OF A ROI-C IMPLANT HOLDER FRACTURED WHEN REMOVING FROM THE CAGE INTRA-OPERATIVELY. THE TIP WAS REMOVED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT PATIENT IMPACTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19469268 ↗
  • MalfunctionMar 30, 2023Fracture; Fracture

    IT WAS REPORTED THAT A ROI-C IMPLANT HOLDER BROKE DURING A PROCEDURE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16646366 ↗
  • MalfunctionMar 30, 2023Fracture; Fracture

    IT WAS REPORTED THAT AN ROI-C IMPLANT HOLDER BROKE INTRA-OPERATIVELY. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16645821 ↗
  • MalfunctionMar 29, 2023Fracture; Fracture

    IT WAS REPORTED THAT THE HOOK AT THE TIP OF AN ROI-C IMPLANT HOLDER WAS FRACTURED OFF INTRA-OPERATIVELY WHEN THE SURGEON WAS DETACHING IT FROM THE IMPLANT. THERE WERE NO PATIENT IMPACTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16640355 ↗
  • MalfunctionAug 29, 2022Fracture; Physical Resistance/Sticking; Fracture; Physical Resistance/Sticking

    IT WAS REPORTED THE THE HOOK BROKE OFF AN ROI-C IMPLANT HOLDER AND A PLATE BECAME STUCK IN THE HOLDER INTRA-OPERATIVELY. THERE WERE NO PATIENT IMPACTS. THIS IS REPORT TWO OF TWO FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15311160 ↗
  • MalfunctionDec 22, 2021Fracture; Fracture

    IT WAS REPORTED THAT DURING THE SURGERY AFTER INSERTING THE IMPLANT, THE SURGEON WENT TO REMOVE THE HOLDER AND THE HOOK BROKE OFF. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13060724 ↗
  • MalfunctionDec 17, 2021Fracture; Fracture

    IT WAS REPORTED THAT THE HOOK ON THE TIP OF AN IMPLANT INSERTER BROKE OFF INTRA-OPERATIVELY. THE HOOK WAS RECOVERED WITHOUT PATIENT IMPACTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13026223 ↗
  • MalfunctionDec 3, 2020Fracture; Fracture

    IT WAS REPORTED THAT DURING THE SURGERY, THE HOOK ON THE IMPLANT HOLDER FRACTURED. AN ALTERNATIVE IMPLANT HOLDER WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO IMPACTS TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10937357 ↗

Other Intervertebral Fusion Device With Integrated Fixation, Cervical devices

All Intervertebral Fusion Device With Integrated Fixation, Cervical devices →