Tornier Perform MWE413Recalls & reported problems
Tornier · CortiLoc™ PEGGED GLENOID TRIAL · Class 1
All models of Tornier Perform (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 03700386974441), received between Jul 6, 2020 and Mar 29, 2022 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
- Fracture1
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Bone Fracture(s)1
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 29, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE M40 CORTILOC TRIAL BROKE OFF INSIDE THE PATIENT. THE PHYSICIAN PLACED A THREADED GUIDE WIRE INTO THE PIECE TO PULL IT OUT OF THE PATIENTS GLENOID. IT WAS SUCCESSFULLY REMOVED. NO PATIENT IMPACT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13938596 ↗MalfunctionJan 6, 2022Break; Break
IT WAS REPORTED THAT THE TRIAL BROKE WHEN THE SURGEON WAS REMOVING THE TRIAL. IT BROKE AGAINST THE POSTERIOR RETRACTOR. NO PATIENT IMPACT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13182465 ↗InjuryJul 6, 2020Material Fragmentation; Material Fragmentation
THE INFERIOR POSTERIOR (IN A LEFT SHOULDER) PERIPHERAL PEG BROKE OFF AND STAYED IN THE GLENOID WHEN THE TRIAL WAS REMOVED. IN THE PROCESS TO RETRIEVE THE PEG FROM THE GLENOID, THE INFERIOR POSTERIOR QUADRANT OF THE GLENOID BROKE OFF AND DR (B)(6) HAD TO CONVERT FROM AN ANATOMIC TO A REVERSE DUE TO THE LOSE OF GLENOID BONE. SURGERY WAS COMPLETED BY SWITCHING TO PERFORM A REVERSE SHOULDER ARTHROPLASTY. THE PATIENT REQUIRED A REVERSE INSTEAD OF AN ANATOMIC SHOULDER. IT ADDED 45 MINUTES TO THE CASE AND ULTIMATELY MADE DR (B)(6) SWITCH FROM AN ANATOMIC TO A REVERSE ARTHROPLASTY.
Patient problems: Bone Fracture(s); Device Embedded In Tissue or Plaque
MAUDE report 10232167 ↗
Other Orthopedic manual surgical instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page