Skip to content

Idys®-LLIF 3DTi 51430013-SRecalls & reported problems

Clariance · LLIF 3DTi Plate · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03700780634521), received between Jun 14, 2021 and Jun 14, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Disconnection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Disconnection1
  • Entrapment of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 14, 2021Disconnection; Entrapment of Device

    SURGEON INSERTED THE CAGE + PLATE ASSEMBLY. HE WAS UNHAPPY WITH THE CAGE POSITION. HE DECIDED TO REMOVE THE CAGE + PLATE ASSEMBLY. WHILE EXTRACTING THE CAGE + PLATE ASSEMBLY AND USING SLAP-HAMMER, THE PLATE DISCONNECTED FROM CAGE. AS A RESULT, CAGE REMAINED IN THE DISC SPACE. AN INSTRUMENT AVAILABLE DURING SURGERY WAS USED TO REMOVE CAGE OUT OF THE DISC SPACE. THE SURGEON RE-INSERTED THE CAGE + PLATE ASSEMBLY AND THE PLATE DISCONNECTED AGAIN. AS A CONSEQUENCE, THE PRODUCT WAS USED WITHOUT PLATE. NO HARM WAS CAUSED TO THE PATIENT. AS ORIGINALLY PLANNED, A POSTERIOR FIXATION WAS USED TO BACK-UP THE CAGE. DELAY IN SURGERY WAS ABOUT 15 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11994829 ↗

Other Intervertebral fusion device with integrated fixation, lumbar devices

All Intervertebral fusion device with integrated fixation, lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page