ORTHOPEDIC MANUAL SURGICAL INSTRUMENT 951-0023Recalls & reported problems
Fx Solutions · DRILLING GUIDE DEPTH GAUGE · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03701037316894), received between May 5, 2025 and May 5, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Contamination /Decontamination Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
DURING A CASE WITH THE SURGEON ON (B)(6) 2025, THE DRILL GUIDE WAS DISCOVERED TO HAVE BONE DEBRIS OBSTRUCTING THE CANNULA. THE INSTRUMENT WAS FLUSHED WITH BETADINE WASH IN THE SURGICAL SUITE, AS THE SURGEON FELT THE DEBRIS WAS AT LEAST STERILE. THERE WAS NO SURGICAL DELAY. A PHOTO RETURNED DOES CONFIRM THE ISSUE. THE INSTRUMENT SET IN QUESTION HAS BEEN IDENTIFIED. IT WAS SHIPPED TO THE DISTRIBUTOR, AND IS NOT YET RETURNED TO FX SHOULDER SOLUTIONS. THE CASE PERFORMED ON 23-APR-2025 BY THE DISTRIBUTOR WAS THE FIRST CASE PERFORMED DURING THIS TIME. THE INSTRUMENT IN QUESTION WAS NOT ADEQUATELY INSPECTED OR CLEANED BY FX SHOULDER SOLUTIONS WAREHOUSE STAFF. CAPA-2025-0030 HAS BEEN OPENED TO ADDRESS THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21963608 ↗MalfunctionMay 5, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
DURING A CASE WITH THE SURGEON ON (B)(6) 2025, THE DRILL GUIDE WAS DISCOVERED TO HAVE BONE DEBRIS OBSTRUCTING THE CANNULA. THERE WERE 20 MINUTES OF SURGICAL DELAY AS THE SURGICAL TEAM USED ANOTHER SET OF INSTRUMENTATION TO CONTINUE THE SURGERY. THE INSTRUMENT SET IN QUESTION HAS BEEN IDENTIFIED. IT WAS SHIPPED TO THE DISTRIBUTOR 14-APR-2025, AND RETURNED TO FX SHOULDER SOLUTIONS ON 23-APR-2025. THE CASE PERFORMED ON 17-APR-2025 WAS THE ONLY CASE PERFORMED BY THE DISTRIBUTOR. THE INSTRUMENT IN QUESTION WAS NOT ADEQUATELY INSPECTED OR CLEANED BY FX SHOULDER SOLUTIONS WAREHOUSE STAFF. CAPA-2025-0030 HAS BEEN OPENED TO ADDRESS THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21963607 ↗
Other Orthopedic Manual Surgical Instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page