QUANTUM M50 ST151Recalls & reported problems

In2bones · QUANTUM Tibial Implant Long Right Size 5 · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03701337601461), received between Sep 18, 2025 and Sep 18, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Fragmentation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2025Material Fragmentation; Material Fragmentation

    MEDICAL DEVICE DESCRIPTION: THE QUANTUM® TOTAL ANKLE PROSTHESIS IS A FIXED-BEARING SEMI-CONSTRAINT ANKLE PROTHESIS WITH 2 COMPONENTS EXISTING IN DIFFERENT SIZES AND MODELS AND COMPOSED OF: - A TIBIAL COMPONENT COMPOSED OF A METALLIC (TA6V) TIBIAL TRAY IMPLANT FIXED TO A POLYMER (UHWMPE) INSERT, - A METALLIC (COCR) TALAR IMPLANT REPRODUCING THE TALUS DOME ANATOMY. WHEN ALL COMPONENTS ARE IMPLANTED, WITH THE INSERT RIGIDLY LOCKED TO THE TIBIAL TRAY, THE POLYETHYLENE INSERT ACTS AS A SINGLE BEARING ALONG THE TALAR DOME, ENABLING FLEXION/EXTENSION AND ROTATION MOVEMENT AT THE REPLACED ANKLE JOINT. EACH COMPONENT OF THE QUANTUM® TOTAL ANKLE PROSTHESIS EXISTS IN DIFFERENT SIZES AND MODELS. EVENT DESCRIPTION : SURGICAL INTERVENTION TO INSTALL A QUANTUM TOTAL ANKLE PROSTHESIS WITH PATIENT SPECIFIC CUTTING GUIDES. DUE TO DIFFICULTIES IN HAVING A COMPLETE IMPACTION OF THE TIBIAL IMPLANT (SCOPIC CONTROL), THE IMPLANT WAS REMOVED MANUALLY (WITHOUT USING INSTRUMENT) IN ORDER TO BE REPOSITIONED AFTER ADDITIONAL PREPARATION WITH THE FIXATION CROSS PREPARATION INSTRUMENTS. THE IMPLANT WAS PLACED ON THE INSTRUMENTATION TABLE, AND THEN IT WAS NOTICED THAT PART OF THE ANTEROMEDIAL COATING OF THE IMPL

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23084903 ↗

Other Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer devices

All Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer devices →

Sources and updates