BIOROOT RCS 01E0300Recalls & reported problems
Septodont Sas Specialites Septodont · BIOACTIVE ROOT CANAL SEALER · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03760140703373), received between Jul 4, 2017 and Oct 23, 2017 — classified as 2 injuries.
The most frequently reported specific device problem is “Extrusion” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Extrusion1
Patient problems
- Necrosis1
- Swelling1
- Tingling1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 23, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
SPONTANEOUS REPORT, (B)(4). INITIAL INFORMATION RECEIVED ON 21-SEP-2017 FROM THE DENTIST BY PHONE. THE DENTIST REPORTED THAT A (B)(6) FEMALE PATIENT, WITH MEDICAL HISTORY OF ALLERGY MULTIPLE, PENICILLIN AND ASPIRIN ALLERGY, DENTAL FISTULA AND BROKEN TOOTH #21, RECEIVED AN UNKNOWN DOSE OF THE SUSPECTED SEPTANEST 40 MG/ML WITH ADRENALINE 1/200.000 SOLUTION FOR INJECTION (ARTICAINE HYDROCHLORIDE; EPINEPHRINE BITARTRATE), BY DENTAL ROUTE FOR LOCAL ANAESTHESIA BEFORE A ROOT CANAL PROCEDURE WITH THE MEDICAL DEVICE BIOROOT RCS (BATCH B18997, EXP 08/2018), ON (B)(6) 2017. THE PATIENT HAD PREVIOUSLY BEEN TREATED WITH SEPTANEST 40 MG/ML WITH ADRENALINE 1/200.000, WITHOUT COMPLICATIONS. ON (B)(6) 2017, AT THE END OF DENTAL PROCEDURE, THE DENTIST NOTED A GINGIVAL SWELLING WITHOUT INFLAMMATORY REACTION (NO REDNESS), AND THE PATIENT EXPERIENCED SOFT TISSUE NECROSIS RATHER WHITISH APPEARANCE. DURING THE ROOT CANAL TREATMENT, THE DENTIST USED A DISINFECTANT (PAROEX) FOR CLEANING THE ROOT, WITHOUT REACTION OF THE PATIENT. THE DENTIST CONFIRMED THERE WAS NO PERIAPICAL EXTRUSION OF PRODUCT. ON (B)(6) 2017, THE DENTIST SAW AGAIN THE PATIENT AND EXTENSION OF SWELLING FROM CHEEK TO EYE WAS NOTED. THE DE
Patient problems: Necrosis; Swelling
MAUDE report 6968516 ↗InjuryJul 4, 2017Extrusion; Extrusion
ADDITIONAL INFORMATION RECEIVED ON (B)(4) 2017: THE DENTIST REPORTED THE BATCH NUMBER OF BIOROOT WAS (B)(4). HE STATED THAT HE HAS USED BIOROOT SEVERAL TIMES BEFORE ON SEVERAL PATIENTS. THE DENTIST SAW THE PATIENT AGAIN ON (B)(6) 2017. HE REPORTED THAT THE SURGEON DID NOT REMOVE THE PRODUCT WHICH WAS CLOSE AROUND THE NERVE AS HE THOUGHT IT WAS NOT NECESSARY. THE PATIENT HAD RECOVERED FROM NUMBNESS OF LIPS AND CHIN. CAUSALITY ASSESSMENT ON (B)(4) 2017 ON ADDITIONAL INFORMATION RECEIVED ON (B)(4) 2017: SERIOUSNESS: SERIOUS (OTHER MEDICALLY IMPORTANT CONDITION AND REQUIRED INTERVENTION TO PREVENT PERMANENT IMPAIRMENT/DAMAGE). EXPECTEDNESS: PARESTHESIA ORAL: UNLISTED EU, UNEXPECTED FR, DEVICE EXTRUSION: UNLISTED EU, UNEXPECTED FR, CAUSALITY: LATENCY: POSSIBLE. RECOGNIZED ASSOCIATION: NO. ANALYSIS: LIP PARAESTHESIA MAY BE DUE TO DENTAL PROCEDURE CONSIDERING POSSIBLE DIRECT NERVE TRAUMA, CHEMICAL TOXICITY OF IRRIGATION SOLUTION OR COMPRESSION OF THE NERVE CAUSED BY HEMATOMA, OEDEMA OR OVERFILLING SPREADING OF BIOROOT. IN THIS CASE, THE DENTIST REPORTED BIOROOT EXTRUSION, AND COMPRESSION OF THE EXTERNAL BRANCH OF THE MANDIBULAR NERVE WAS CONFIRMED BY SCAN. ACCORDING TO THE LAST INFORMATIO
Patient problems: Tingling
MAUDE report 6685396 ↗
Other Resin, Root Canal Filling devices
- AH Plus BioCermanic Sealer AHPBIOSEALTulsa Dental Products LLC · RESIN, ROOT CANAL FILLING46 reports
- AH Plus BioCermanic Sealer AHPBIOREFILLTulsa Dental Products LLC · RESIN, ROOT CANAL FILLING10 reports
- AH PLUS ROOTCANAL 60620110Dentsply De Trey GmbH · RESIN, ROOT CANAL FILLING4 reports
- AH PLUS JET 60620115Dentsply De Trey GmbH · RESIN, ROOT CANAL FILLING3 reports
- AH PLUS JET 667005Dentsply De Trey GmbH · RESIN, ROOT CANAL FILLING1 report
- AH PLUS ROOTCANAL 667003Dentsply De Trey GmbH · RESIN, ROOT CANAL FILLING1 report
- RibbonTulsa Dental Products LLC · RESIN, ROOT CANAL FILLING1 report
- ProRoot MTA PROROOT5WTulsa Dental Products LLC · RESIN, ROOT CANAL FILLING1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.