Skip to content

AH PLUS JET 667005Recalls & reported problems

Dentsply De Trey GmbH · AH Plus Jet Starter Kit is a 2 component paste/paste root canal sealer based on epoxy-amine resin. AH Plus Jet is received in a syringe delivery for homogeneous mixing of both pastes. With its supporting accessories (Mixing/Intraoral Tips, Mixing Pad and Organizer Tray) it can be classified as a convenience kit. · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI D0106670051), received between Oct 19, 2023 and Oct 19, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem1
  • Patient-Device Incompatibility1

Patient problems

  • Caustic/Chemical Burns1
  • Hypersensitivity/Allergic reaction1
  • Numbness1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 19, 2023Patient-Device Incompatibility; Patient Device Interaction Problem; Patient-Device Incompatibility; Patient Device Interaction Problem

    IN THIS EVENT IT IS REPORTED THAT AH PLUS JET START KIT MAIL.NA THAT WAS USED IN TREATMENT HAD A MALFUNCTION WITH THE AUTO MIXING TIP. THE UNMIXED SEALER WENT OUT OF THE APEX AND REPORTEDLY CAUSED A CHEMICAL BURN OF PATIENT'S MENTAL NERVE - RIGHT SIDE OF FACE. THE PATIENT REPORTEDLY DEVELOPED NUMBNESS IN THE RIGHT HALF OF THE CHIN AND LOWER LIP AND PAIN ON THE WHOLE RIGHT SIDE OF THEIR FACE TWO DAYS AFTER EXPOSURE TO AH PLUS JET DURING ROOT CANAL TREATMENT. THE SYMPTOMS ARE STILL ONGOING. THE DENTIST ALLEGES THIS WAS DUE TO AN ISSUE WITH THE AUTO-MIXING TIP, WHICH WOULD NOT HAVE MIXED THE PRODUCT CORRECTLY, IN TURN LEADING TO UNMIXED MATERIAL COMING INTO CONTACT WITH THE NERVE AND CAUSING A CHEMICAL BURN. CUSTOMER/DENTIST REPORTS THAT THIS WAS NOT AN ALLERGIC REACTION.

    Patient problems: Caustic/Chemical Burns; Hypersensitivity/Allergic reaction; Numbness

    MAUDE report 17967638 ↗

Other RESIN, ROOT CANAL FILLING devices

All RESIN, ROOT CANAL FILLING devices →