Skip to content

ComposiTCP ComposiTCP 60 round headRecalls & reported problems

S.B.M · Absorbable interference screw · Class 2 · Implant

All models of ComposiTCP (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03760154006675), received between Oct 4, 2024 and Oct 4, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 4, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING AN INITIAL HAMSTRING TENDON GRAFT PROCEDURE, TWO INTERFERENCE SCREWS FRACTURED. HALF OF ONE SCREW REMAINS IN THE PATIENT'S BONE TUNNEL AND IS INDICATED TO BE STABLE WITHOUT THE POTENTIAL FOR MIGRATION. THE SURGEON STATED THAT REMOVING IT WAS NOT POSSIBLE, AND THE CASE WAS COMPLETED WITH A BONE POST OUTSIDE THE TUNNEL FOR GRAFT FIXATION. THE CORRECT SURGICAL TECHNIQUE WAS USED. NO OTHER COMPLICATIONS OR IMPACTS WERE REPORTED DUE TO THIS MALFUNCTION. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 20370650 ↗

Other Fastener, Fixation, Biodegradable, Soft Tissue devices

All Fastener, Fixation, Biodegradable, Soft Tissue devices →