Albumin Gen2 3183688122Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04015630918898), received between Jul 20, 2021 and Jul 31, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (4 reports).

3 other recalls cover devices cleared under the same FDA submission (K033009); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results4
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Incorrect Measurement1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 31, 2025Incorrect Measurement; Incorrect Measurement

    THERE WAS AN ALLEGATION OF 10 QUESTIONABLE ALBUMIN GEN.2 RESULTS FROM THE COBAS 6000 C (501) MODULE. ONLY THE DATA FOR 2 OF THE PATIENTS WERE PROVIDED, AND ARE A REPORTABLE MALFUNCTION. PATIENT 1'S INITIAL RESULT WAS 0.42 G/DL, AND THE REPEAT RESULT FROM ANOTHER INSTRUMENT WAS 4.22 G/DL PATIENT 2'S INITIAL RESULT WAS 0.47 G/DL, AND THE REPEAT RESULT FROM ANOTHER INSTRUMENT WAS 4.78 G/DL. THE REPEAT RESULTS WERE DEEMED TO BE CORRECT. THE INITIAL RESULTS WERE REPEATED BECAUSE THE DOCTOR OR TECHNICIAN QUESTIONED THE CRITICAL RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22676057 ↗
  • MalfunctionSep 11, 2023Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results; Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results

    WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 20 PATIENTS' SAMPLES TESTED WITH ALBUMIN (ALB2) ASSAY ON A COBAS 8000 COBAS C502 ANALYZER. THE CUSTOMER STATED THAT THE RESULTS FOR THE 20 PATIENTS' SAMPLE WERE LESS THAN 3 G/DL WHILE THEIR PREVIOUS RESULTS WERE CLOSE TO 5 G/DL. DISCREPANT RESULTS FOR 2 PATIENTS' SAMPLES WERE PROVIDED. SAMPLE 1: INITIAL RESULT: 2.44 G/DL. REPEAT RESULT: 4.3 G/DL. SAMPLE 2: INITIAL RESULT: 2.66 G/DL. REPEAT RESULT: 4.56 G/DL. THE REPEAT RESULTS WERE DEEMED TO BE CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17719453 ↗
  • MalfunctionJan 4, 2023Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results; Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results

    THE INITIAL REPORTER STATED THEY RAN PATIENT SAMPLES AS PART OF A LOT-TO-LOT COMPARISON ON THE COBAS 8000 C 502 MODULE WHEN STARTING A NEW LOT OF ALB2 ALBUMIN GEN.2. TWO PATIENT SAMPLES HAD DISCREPANT RESULTS FOR ALBUMIN. WHEN TESTED WITH THE OLD LOT (LOT 64818101), THE FIRST SAMPLE INITIALLY RESULTED IN AN ALBUMIN VALUE OF 2.60 G/DL AND THIS VALUE WAS REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLE WAS REPEATED TWICE WITH THE NEW LOT (LOT 657589), RESULTING IN VALUES OF 4.37 G/DL AND 4.36 G/DL. THE REPEAT VALUES WERE DEEMED CORRECT. WHEN TESTED WITH THE OLD LOT (LOT 64818101), THE SECOND SAMPLE INITIALLY RESULTED IN AN ALBUMIN VALUE OF 2.54 G/DL AND THIS VALUE WAS REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLE WAS REPEATED TWICE WITH THE NEW LOT (LOT 657589), RESULTING IN VALUES OF 4.80 G/DL AND 4.79 G/DL. THE REPEAT VALUES WERE DEEMED CORRECT. THE SERIAL NUMBER OF THE C 502 ANALYZER IS (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16096673 ↗
  • MalfunctionOct 31, 2022Low Test Results; Low Test Results

    THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH ALB2 ALBUMIN GEN.2 ON A COBAS 8000 C 502 MODULE. THE SAMPLE INITIALLY RESULTED IN AN ALBUMIN VALUE OF 1.31 G/DL AND THIS VALUE WAS REPORTED OUTSIDE OF THE LABORATORY. AS THE VALUE WAS CONSIDERED CRITICAL, THE SAMPLE WAS REPEATED, RESULTING IN A VALUE OF 1.84 G/DL. THE REPEAT VALUE WAS DEEMED CORRECT. THE ALBUMIN REAGENT LOT NUMBER AND EXPIRATION DATE WERE REQUESTED, BUT NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15698042 ↗
  • MalfunctionJul 20, 2021Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results

    THE INITIAL REPORTER RECEIVED QUESTIONABLE ALB2 ALBUMIN GEN.2 RESULTS FOR ONE PATIENT THE COBAS 6000 C501 MODULE SERIAL NUMBER (B)(4) COMPARED TO A COMPETITOR¿S PLATFORM. THE REPORTER STATED THE SAME SAMPLE WAS TESTED ON 3 COBAS SYSTEMS (TWICE ON A 6000 AND ONCE ON AN 8000) AND ALL THREE WERE NON-COMPARABLE WITH THE COMPETITOR PLATFORM. THE REPORTER PROVIDED THE FOLLOWING EXAMPLES: THE C501 RESULT WAS 0.8 G/DL. THE COMPETITOR RESULT WAS 3.2 G/DL. THE C501 RESULT WAS 0.02 G/DL. THE COMPETITOR RESULT WAS 2.42 G/DL. THE QUESTIONABLE RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12196697 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K033009. One submission often covers many models, so check the product and lot details in each record.

  • Z-0104-2021 · Aug 27, 2020 · Terminated
    Albumin Gen.2, Catalog 05166861190
  • Z-0152-2020 · Sep 11, 2019 · Terminated
    Roche Diagnostics cobas Integra Albumin Gen.2 Catalog Number: 05166861190 - Product Usage: In vitro test for the quantitative determination of the albumin in human serum and plasm
  • Z-0011-2009 · Jun 11, 2008 · Terminated
    Roche Diagnostics ALBT2, Tina-quant Albumin Gen. 2, COBAS INTEGRA, cobas c systems, Roche Diagnostics, Indianapolis, IN.; 04469658190. An in-vitro diagnostic reagent system intend

Other BROMCRESOL GREEN DYE-BINDING, ALBUMIN devices

All BROMCRESOL GREEN DYE-BINDING, ALBUMIN devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page