Albumin Gen.2 08056714190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613336120958), received between Jul 23, 2025 and Sep 17, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (3 reports).

3 other recalls cover devices cleared under the same FDA submission (K033009); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results3
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50805531 ↗
  • MalfunctionJan 28, 2026Low Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE ALBUMIN BCP RESULTS FOR 25 PATIENT SAMPLES ON A COBAS PURE C 303 ANALYTICAL UNIT. THE CUSTOMER QUESTIONED THE RESULTS AS THE ALBUMIN RESULTS WERE LOW BUT THE TOTAL PROTEIN RESULTS WERE NORMAL IN PATIENTS THAT DO NOT HAVE KIDNEY DISEASE. REFER TO THE ATTACHMENT TO THE MEDWATCH FOR ALL PATIENT DATA. THE UNITS OF MEASUREMENT FOR TOTAL PROTEIN WERE NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24199417 ↗
  • MalfunctionNov 19, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THERE WAS AN ALLEGATION OF QUESTIONABLE ALBUMIN BCP RESULTS FROM THE COBAS PURE C 303 ANALYTICAL UNIT THAT DID NOT MATCH THE CLINICAL PICTURE. FOR MULTIPLE PATIENT SAMPLES, THE ALBUMIN RESULTS WERE LOW, BUT THE TOTAL PROTEIN RESULTS WERE NORMAL WITHOUT ANY SPECIFIC DIAGNOSIS RELATED TO LOW ALBUMIN RESULTS. TWO REPRESENTATIVE EXAMPLES: EXAMPLE 1: ON (B)(6) 2025, THE ALBUMIN RESULT WAS 33.80 G/L AND THE TOTAL PROTEIN RESULT WAS 69.9 G/L. EXAMPLE 2: ON (B)(6) 2025, THE ALBUMIN RESULT WAS 37.3 G/L AND THE TOTAL PROTEIN RESULT WAS 75.1 G/L. THE SAMPLES WERE REPEATED ON ANOTHER COBAS C303 AND A COBAS C503 ANALYZERS, AND THE RESULTS WERE IDENTICAL. THE SAMPLES WERE REPEATED USING THE "ALB-G" ASSAY, AND THE RESULTS WERE SLIGHTLY HIGHER BUT STILL LOW. NO SPECIFIC DATA WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23593030 ↗
  • MalfunctionAug 18, 2025Low Test Results; Low Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE ALBUMIN BCP RESULTS FROM THE COBAS PURE C 303 ANALYTICAL UNIT. THE INITIAL RESULT WAS 27G/L. THE SAMPLE WAS THEN SENT TO ANOTHER LAB AS PART OF THE NETWORKS IQAP AND WAS RETESTED ON A COBAS C 701 WITH A RESULT OF 37.3 G/L. THIS RESULT WAS BELIEVED TO BE CORRECT, AND THE REPORTED RESULT WAS AMENDED. AN EDTA SAMPLE FROM THE SAME VENIPUNCTURE OF THE PATIENT WAS TESTE,D AND THE RESULT WAS 40 G/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22826934 ↗
  • MalfunctionJul 23, 2025Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results

    THE INITIAL REPORTER RECEIVED QUESTIONABLE ALBUMIN BCP RESULTS FROM SOME PATIENT SAMPLES TESTED ON THE COBAS C 503 ANALYTICAL UNIT. THE RESULTS FROM THE ANALYZER WERE COMPARED TO TWO COMPETITOR ANALYZERS (COMPETITOR 1 AND COMPETITOR 2). COMPETITOR 1'S LABORATORY METHOD IS UNKNOWN; COMPETITOR 2 USES THE BROMOCRESOL GREEN (BCG) LABORATORY METHOD. THE REPORTER PROVIDED THE FOLLOWING EXAMPLES OF QUESTIONABLE RESULTS: PATIENT SAMPLE 1 THE INITIAL RESULT FROM THE ANALYZER IS 29 G/L. THE REPEAT RESULT FROM COMPETITOR 1 IS 36 G/L. THE REPEAT RESULT FROM COMPETITOR 2 IS 36 G/L. PATIENT SAMPLE 2 THE INITIAL RESULT FROM THE ANALYZER IS 28 G/L. THE REPEAT RESULT FROM COMPETITOR 1 IS 34 G/L. THE REPEAT RESULT FROM COMPETITOR 2 IS 34 G/L. PATIENT SAMPLE 3 THE INITIAL RESULT FROM THE ANALYZER IS 41 G/L. THE REPEAT RESULT FROM COMPETITOR 1 IS 47 G/L. THE REPEAT RESULT FROM COMPETITOR 2 IS 48 G/L. THE REPORTER STATED THAT THE DIFFERENT RESULTS FROM THE DIFFERENT ANALYZERS MAKE PATIENT CARE AND TREATMENT DIFFICULT FOR DOCTORS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22574524 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K033009. One submission often covers many models, so check the product and lot details in each record.

  • Z-0104-2021 · Aug 27, 2020 · Terminated
    Albumin Gen.2, Catalog 05166861190
  • Z-0152-2020 · Sep 11, 2019 · Terminated
    Roche Diagnostics cobas Integra Albumin Gen.2 Catalog Number: 05166861190 - Product Usage: In vitro test for the quantitative determination of the albumin in human serum and plasm
  • Z-0011-2009 · Jun 11, 2008 · Terminated
    Roche Diagnostics ALBT2, Tina-quant Albumin Gen. 2, COBAS INTEGRA, cobas c systems, Roche Diagnostics, Indianapolis, IN.; 04469658190. An in-vitro diagnostic reagent system intend

Other BROMCRESOL GREEN DYE-BINDING, ALBUMIN devices

All BROMCRESOL GREEN DYE-BINDING, ALBUMIN devices →