Albumin Gen.2 08056714190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613336120958), received between Jul 23, 2025 and Sep 17, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (3 reports).
3 other recalls cover devices cleared under the same FDA submission (K033009); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results3
- Incorrect, Inadequate or Imprecise Result or Readings2
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50805531 ↗MalfunctionJan 28, 2026Low Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ALBUMIN BCP RESULTS FOR 25 PATIENT SAMPLES ON A COBAS PURE C 303 ANALYTICAL UNIT. THE CUSTOMER QUESTIONED THE RESULTS AS THE ALBUMIN RESULTS WERE LOW BUT THE TOTAL PROTEIN RESULTS WERE NORMAL IN PATIENTS THAT DO NOT HAVE KIDNEY DISEASE. REFER TO THE ATTACHMENT TO THE MEDWATCH FOR ALL PATIENT DATA. THE UNITS OF MEASUREMENT FOR TOTAL PROTEIN WERE NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24199417 ↗MalfunctionNov 19, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THERE WAS AN ALLEGATION OF QUESTIONABLE ALBUMIN BCP RESULTS FROM THE COBAS PURE C 303 ANALYTICAL UNIT THAT DID NOT MATCH THE CLINICAL PICTURE. FOR MULTIPLE PATIENT SAMPLES, THE ALBUMIN RESULTS WERE LOW, BUT THE TOTAL PROTEIN RESULTS WERE NORMAL WITHOUT ANY SPECIFIC DIAGNOSIS RELATED TO LOW ALBUMIN RESULTS. TWO REPRESENTATIVE EXAMPLES: EXAMPLE 1: ON (B)(6) 2025, THE ALBUMIN RESULT WAS 33.80 G/L AND THE TOTAL PROTEIN RESULT WAS 69.9 G/L. EXAMPLE 2: ON (B)(6) 2025, THE ALBUMIN RESULT WAS 37.3 G/L AND THE TOTAL PROTEIN RESULT WAS 75.1 G/L. THE SAMPLES WERE REPEATED ON ANOTHER COBAS C303 AND A COBAS C503 ANALYZERS, AND THE RESULTS WERE IDENTICAL. THE SAMPLES WERE REPEATED USING THE "ALB-G" ASSAY, AND THE RESULTS WERE SLIGHTLY HIGHER BUT STILL LOW. NO SPECIFIC DATA WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23593030 ↗MalfunctionAug 18, 2025Low Test Results; Low Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ALBUMIN BCP RESULTS FROM THE COBAS PURE C 303 ANALYTICAL UNIT. THE INITIAL RESULT WAS 27G/L. THE SAMPLE WAS THEN SENT TO ANOTHER LAB AS PART OF THE NETWORKS IQAP AND WAS RETESTED ON A COBAS C 701 WITH A RESULT OF 37.3 G/L. THIS RESULT WAS BELIEVED TO BE CORRECT, AND THE REPORTED RESULT WAS AMENDED. AN EDTA SAMPLE FROM THE SAME VENIPUNCTURE OF THE PATIENT WAS TESTE,D AND THE RESULT WAS 40 G/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22826934 ↗MalfunctionJul 23, 2025Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results
THE INITIAL REPORTER RECEIVED QUESTIONABLE ALBUMIN BCP RESULTS FROM SOME PATIENT SAMPLES TESTED ON THE COBAS C 503 ANALYTICAL UNIT. THE RESULTS FROM THE ANALYZER WERE COMPARED TO TWO COMPETITOR ANALYZERS (COMPETITOR 1 AND COMPETITOR 2). COMPETITOR 1'S LABORATORY METHOD IS UNKNOWN; COMPETITOR 2 USES THE BROMOCRESOL GREEN (BCG) LABORATORY METHOD. THE REPORTER PROVIDED THE FOLLOWING EXAMPLES OF QUESTIONABLE RESULTS: PATIENT SAMPLE 1 THE INITIAL RESULT FROM THE ANALYZER IS 29 G/L. THE REPEAT RESULT FROM COMPETITOR 1 IS 36 G/L. THE REPEAT RESULT FROM COMPETITOR 2 IS 36 G/L. PATIENT SAMPLE 2 THE INITIAL RESULT FROM THE ANALYZER IS 28 G/L. THE REPEAT RESULT FROM COMPETITOR 1 IS 34 G/L. THE REPEAT RESULT FROM COMPETITOR 2 IS 34 G/L. PATIENT SAMPLE 3 THE INITIAL RESULT FROM THE ANALYZER IS 41 G/L. THE REPEAT RESULT FROM COMPETITOR 1 IS 47 G/L. THE REPEAT RESULT FROM COMPETITOR 2 IS 48 G/L. THE REPORTER STATED THAT THE DIFFERENT RESULTS FROM THE DIFFERENT ANALYZERS MAKE PATIENT CARE AND TREATMENT DIFFICULT FOR DOCTORS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22574524 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033009. One submission often covers many models, so check the product and lot details in each record.
- Z-0104-2021 · Aug 27, 2020 · TerminatedAlbumin Gen.2, Catalog 05166861190
- Z-0152-2020 · Sep 11, 2019 · TerminatedRoche Diagnostics cobas Integra Albumin Gen.2 Catalog Number: 05166861190 - Product Usage: In vitro test for the quantitative determination of the albumin in human serum and plasm
- Z-0011-2009 · Jun 11, 2008 · TerminatedRoche Diagnostics ALBT2, Tina-quant Albumin Gen. 2, COBAS INTEGRA, cobas c systems, Roche Diagnostics, Indianapolis, IN.; 04469658190. An in-vitro diagnostic reagent system intend
Other BROMCRESOL GREEN DYE-BINDING, ALBUMIN devices
- VITROS 6844293Ortho-Clinical Diagnostics, Inc. · BROMCRESOL GREEN DYE-BINDING, ALBUMIN2 recalls0 reports
- Albumin Gen2 5166861190Roche Diagnostics GmbH · BROMCRESOL GREEN DYE-BINDING, ALBUMIN8 reports
- Albumin Gen.2 08056692190Roche Diagnostics GmbH · BROMCRESOL GREEN DYE-BINDING, ALBUMIN6 reports
- Albumin Gen2 3183688122Roche Diagnostics GmbH · BROMCRESOL GREEN DYE-BINDING, ALBUMIN5 reports
- VITROS 1149764Ortho-Clinical Diagnostics, Inc. · BROMCRESOL GREEN DYE-BINDING, ALBUMIN3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page