VITROS 6844293Recalls & reported problems
Ortho-Clinical Diagnostics, Inc. · VITROS XT Chemistry Products ALB-TP Slides · Class 2
All models of VITROS (71)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Jul 10, 2024Class IIOpen, Classified
The reason for this recall is QuidelOrtho has determined that during normal use the VITROS XT ALB-TP Slides may shed dust and debris as they travel through the MicroSlide Subsystem of the VITROS XT 3400 or VITROS XT 7600 analyzer. This dust and debris may then settle onto VITROS Na+ (Sodium) Slides as they are processed through the same MicroSlide Subsystem, potentially resulting in an increase in non-reproducible, positively, or negatively biased Na+ results. The level of bias will depend on where the dust and debris settle on the Na+ MicroSlide and may impact Na+ measurements to varying degrees inside and outside of the measuring range depending on the amount of contamination. All of which could vary from no/low bias with no patient harm, to extreme bias (falsely elevated or reduced results) potentially resulting in incorrect diagnosis; treatment; and/or a serious patient health hazard (for example, hypernatremia or hyponatremia). This issue may also potentially generate error codes preventing the analyzer from processing the MicroSlide Assays and cause delay of the patient results. All current and expired VITROS XT Chemistry Products ALB-TP slides are affected by this issue.
Z-2684-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Dec 15, 2020Class IITerminated
Eltrombopag Interference for VITROS Chemistry Products TP Slides and VITROS XT Chemistry Products ALB-TP Slides
Z-0845-2021 · Root cause: Device Design · Recall details · FDA record ↗
FDA clearance and classification
The FDA regulates product code CIX (bromcresol green dye-binding, albumin) under 21 CFR 862.1035 as a Class II device. 21 CFR 862.1035 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the VITROS 6844293
Has the VITROS 6844293 been recalled?
Yes. 2 FDA recall records name this device's identifier: 1 still open and 1 closed (terminated) by the FDA. The most recent, Z-2684-2024 (Class II), began on Jul 10, 2024: The reason for this recall is QuidelOrtho has determined that during normal use the VITROS XT ALB-TP Slides may shed dust and debris as they travel through the MicroSlide Subsystem of the VITROS XT 3400 or VITROS XT… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
Who makes the VITROS 6844293, and is it still sold?
The FDA's device database lists Ortho-Clinical Diagnostics, Inc. as the company responsible for it (brand VITROS, model 6844293). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BROMCRESOL GREEN DYE-BINDING, ALBUMIN devices
- Albumin Gen2 5166861190Roche Diagnostics GmbH · BROMCRESOL GREEN DYE-BINDING, ALBUMIN8 reports
- Albumin Gen2 3183688122Roche Diagnostics GmbH · BROMCRESOL GREEN DYE-BINDING, ALBUMIN5 reports
- Albumin Gen.2 08056714190Roche Diagnostics GmbH · BROMCRESOL GREEN DYE-BINDING, ALBUMIN4 reports
- Albumin Gen.2 08056692190Roche Diagnostics GmbH · BROMCRESOL GREEN DYE-BINDING, ALBUMIN3 reports
- VITROS 1149764Ortho-Clinical Diagnostics, Inc. · BROMCRESOL GREEN DYE-BINDING, ALBUMIN3 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page