Elecsys total PSA 08791686160Recalls & reported problems
Roche Diagnostics GmbH · Class 3
All models of Elecsys total PSA (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613336165997), received between Jan 27, 2023 and Dec 4, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect Measurement2
- Non Reproducible Results2
- High Test Results1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 4, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE INITIAL REPORTER RECEIVED A QUESTIONABLE ELECSYS TOTAL PSA (TOTAL PSA) RESULT FROM ONE PATIENT SAMPLE TESTED ON THE COBAS E411 RACK. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN. THE PHYSICIAN QUESTIONED THE RESULT AS IT WAS NOT REFLECTIVE OF THE PATIENT'S CONDITION. THE INITIAL RESULT FROM THE ANALYZER WAS 0.06 NG/ML. THE FIRST REPEAT RESULT FROM THE ANALYZER WAS 4.2 NG/ML. THIS REPEAT RESULT WAS DEEMED CORRECT. THE SECOND REPEAT RESULT FROM THEIR OTHER E 411 WAS 4.5 NG/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18261803 ↗MalfunctionAug 29, 2023Incorrect Measurement; Non Reproducible Results; Incorrect Measurement; Non Reproducible Results
THE INITIAL REPORTER QUESTIONED THE RESULTS FOR 3 PATIENTS TESTED FOR ELECSYS TOTAL PSA (TOTAL PSA) THAT DID NOT CORRESPOND TO THEIR CLINICAL HISTORY ON A COBAS E 411 IMMUNOASSAY ANALYZER. ON (B)(6) 2023 PATIENT 1 RESULT WAS 1.3 NG/ML. ON (B)(6) 2023 A NEW SAMPLE WAS OBTAINED AND THE RESULT WAS 229 NG/ML. ON (B)(6) 2023 PATIENT 2 RESULT WAS LESS THAN 0.01 NG/ML. ON (B)(6) 2023 A NEW SAMPLE WAS OBTAINED AND THE RESULT WAS 4.69 NG/ML. ON (B)(6) 2023 PATIENT 3 RESULT WAS 0.02 NG/ML. ON (B)(6) 2023 A NEW SAMPLE WAS OBTAINED AND THE RESULT WAS 32.2 NG/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17641224 ↗MalfunctionJan 27, 2023Incorrect Measurement; High Test Results; Non Reproducible Results; Incorrect Measurement; High Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF A QUESTIONABLE HIGH TOTAL PSA ELECSYS RESULT FOR ONE PATIENT FROM THE COBAS E 411 RACK ANALYZER. THE INITIAL RESULT WAS 16.58 NG/ML AND THE REPEAT RESULT ON (B)(6) 2023 AFTER DEFROSTING THE SAMPLE WAS 0.531 NG/ML. THE QUESTIONABLE RESULT WAS REPORTED OUTSIDE OF THE LABORATORY AND WAS QUESTIONED BY THE LABORATORY DIRECTOR, AS IT DID NOT MATCH THE HISTORY OF THE PATIENT. THE REAGENT LOT NUMBER WAS 59860000 WITH AN EXPIRATION DATE OF 30-JUN-2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16254737 ↗
Other Total, prostate specific antigen (noncomplexed & complexed) for detection of prostate cancer devices
- IMMULITE® 2000 Systems PTS PSA 10706281Siemens Healthcare Diagnostics Products Ltd · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer1 recall1 report
- IMMULITE® 2000 Systems PTS PSA 10706282Siemens Healthcare Diagnostics Products Ltd · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer1 recall0 reports
- Elecsys total PSA 08791732190Roche Diagnostics GmbH · Total, prostate specific antigen (noncomplexed & complexed) for detection of prostate cancer48 reports
- Dimension Vista® System TPSA Flex® Reagent Cartridge 10445090Siemens Healthcare Diagnostics Inc. · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer36 reports
- Dimension® clinical chemistry system TPSA Flex® reagent cartridge 10444889Siemens Healthcare Diagnostics Inc. · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer12 reports
- Elecsys Total PSA 09744860190Roche Diagnostics GmbH · Total, prostate specific antigen (noncomplexed & complexed) for detection of prostate cancer4 reports
- Elecsys total PSA 07027966190Roche Diagnostics GmbH · Total, prostate specific antigen (noncomplexed & complexed) for detection of prostate cancer3 reports
- Access Hybritech PSA 37200Beckman Coulter, Inc. · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page