Skip to content

Endo-Model Knee System - Instruments 17-5220/01Recalls & reported problems

WALDEMAR LINK GmbH & Co. KG · Surgical instrument handle, non-torque-limiting - Endo Model Knee System · Class 1

All models of Endo-Model Knee System - Instruments (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04026575284658), received between Jul 27, 2023 and Jul 27, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Solder Joint Fracture” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Jul 26, 2023Open, Classified

    A crack in the weld of the rasp handle may lead to a fracture of the frontal sleeve when the rasp is further impacted or extracted. The rasp remains intact. The rasp would need to be pulled out by a clamp leading to a prolongation of surgery or modified surgery

    Z-2481-2023 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Solder Joint Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 27, 2023Solder Joint Fracture; Solder Joint Fracture

    THE RASPHANDLE BROKE WHEN IT WAS BEING HAMMERED IN.[CUSTOMER].

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17403278 ↗

Other Orthopedic manual surgical instrument devices

All Orthopedic manual surgical instrument devices →

Sources and updates