S5 Double Roller Pump 85 DRP 85 (UDI 04033817900405)Recalls & reported problems
LivaNova Deutschland GmbH · Class 2
Quick summary
This LivaNova S5 Double Roller Pump 85 is the blood pump used on a heart-lung machine during cardiopulmonary bypass surgery. The 54 reports mostly describe the pump stopping or not pumping smoothly during a procedure, along with error messages on the display and a cockpit screen that briefly blacked out. There are no deaths and no injuries, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 54
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 04033817900405), received between Jun 8, 2016 and Mar 17, 2026 — classified as 52 malfunctions.
The most frequently reported specific device problem is “Pumping Stopped” (27 reports).
13 other recalls cover devices cleared under the same FDA submission (K071318); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Pumping Stopped27
- Device Displays Incorrect Message10
- Pumping Problem7
- Appropriate Device Problem Term/Code Not Available3
- Failure to Auto Stop2
- Image Display Error/Artifact2
- No Display/Image2
- Activation, Positioning or Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions30
- No Patient Involvement18
- No Consequences Or Impact To Patient6
Recent adverse event reports
The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 17, 2026Pumping Stopped
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT, DURING PROCEDURE, A S5 DOUBLE ROLLER PUMP HAD AN INTERNAL ALARM. THE MESSAGE DISPLAY READ "INTERNAL TEMPERATURE IS TOO HIGH". REPORTEDLY, PUMP WAS NOT ROTATING SMOOTHLY BUT JERKY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24616385 ↗MalfunctionMar 10, 2026Application Program Problem
LIVANOVA DEUTSCHLAND HAS RECEIVED A REPORT CONCERNING AN ESSENZ HEART-LUNG MACHINE (HLM) COCKPIT THAT BRIEFLY BLACKED OUT AFTER THE PERFUSIONIST REVERSED THE PUMP DIRECTION DURING A PROCEDURE. NO PATIENT INJURY HAS BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24556530 ↗MalfunctionFeb 27, 2026Pumping Problem
SEE INITIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24471746 ↗MalfunctionSep 3, 2025Failure to Auto Stop; Failure to Auto Stop
SEE INITIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22950145 ↗MalfunctionMay 7, 2025Pumping Stopped; Pumping Stopped
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT, DURING A PROCEDURE, THE S5 DOUBLE ROLLER PUMP 85 FAILED AND SHUT OFF 10 SECOND AFTER BEING TURNED ON (NO FLOW). THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21976561 ↗MalfunctionNov 29, 2024Pumping Problem; Pumping Problem
SEE INITIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20809263 ↗MalfunctionOct 9, 2024Pumping Stopped; Pumping Stopped
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A S5 DOUBLE ROLLER PUMP 85 HEAD MALFUNCTIONED WHEN IT WAS IN SERVICE. IT WAS THE BLOOD PUMP ON THE CARDIOPLEGIA DOUBLE HEADER. THE AFFECTED PUMP HAS BEEN REPLACED WITH A SPARE ONE. NO ADDITIONAL INFORMATION IS AVAILABLE SO FAR. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20407313 ↗MalfunctionOct 9, 2024Pumping Stopped; Pumping Stopped
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A S5 DOUBLE ROLLER PUMP 85 HEAD MALFUNCTIONED WHEN IT WAS IN SERVICE. IT WAS THE BLOOD PUMP ON THE CARDIOPLEGIA DOUBLE HEADER. THE AFFECTED PUMP HAS BEEN REPLACED WITH A SPARE ONE. NO ADDITIONAL INFORMATION IS AVAILABLE SO FAR. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20407264 ↗MalfunctionJul 8, 2024Pumping Stopped; Pumping Stopped
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A S5 DOUBLE HEAD PUMP GAVE A FAULT IN MOTOR CONTROLLER ERROR MESSAGE (CODE 442) DURING PROCEDURE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19687681 ↗MalfunctionJul 5, 2024Pumping Problem; Pumping Problem
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A S5 DOUBLE HEAD PUMP GAVE AN INTERMITTENT ERROR (ROTATION DETECTOR ERROR VALUE) DURING PROCEDURE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19678251 ↗MalfunctionMar 6, 2024Pumping Problem; Pumping Problem
SEE INITIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18844770 ↗MalfunctionJan 12, 2024Pumping Stopped; Pumping Stopped
SEE INTIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18500849 ↗MalfunctionJul 27, 2023Pumping Stopped; Pumping Stopped
SEE INITIAL REPORT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17404798 ↗MalfunctionMay 12, 2023Pumping Stopped; Pumping Stopped
SEE INITIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16917933 ↗MalfunctionApr 21, 2023Pumping Problem; Pumping Problem
SEE INTIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16783509 ↗MalfunctionApr 5, 2023Pumping Stopped; Pumping Stopped
SEE INITIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16681784 ↗MalfunctionFeb 23, 2023Pumping Stopped; Pumping Problem; Pumping Stopped; Pumping Problem
MalfunctionFeb 23, 2023Pumping Stopped; Pumping Problem; Pumping Stopped; Pumping Problem
SEE INTIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16427041 ↗MalfunctionJan 19, 2023Pumping Stopped; Pumping Stopped
SEE INITIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16204963 ↗MalfunctionNov 23, 2022Pumping Stopped; Pumping Stopped
LIVANOVA RECEIVED A REPORT THAT A S5 ROLLER PUMP GAVE A MOTOR CONTROL FAULT ERROR MESSAGE DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15852321 ↗MalfunctionOct 27, 2022Pumping Stopped; Pumping Stopped
LIVANOVA RECEIVED REPORT THAT A S5 DOUBLE HEAD PUMP GAVE A MOTOR CONTROL FAILURE ERROR MESSAGE DURING A PROCEDURE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15679559 ↗MalfunctionOct 17, 2022Material Rupture; Material Rupture
LIVANOVA RECEIVED REPORT THAT A S5 DOUBLE HEAD PUMP HAD A BLOOD LEAK DUE TO HEMOCON TUBE BREAKAGE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15615191 ↗MalfunctionOct 12, 2022Pumping Stopped; Pumping Stopped
LIVANOVA RECEIVED REPORT THAT A S5 DOUBLE ROLLER PUMP GAVE A MOTOR CONTROL FAILURE ERROR MESSAGE DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15585429 ↗MalfunctionSep 2, 2022Pumping Stopped; Pumping Stopped
LIVANOVA RECEIVED REPORT OF MULTIPLE MOTOR CONTROL FAILURE ERROR MESSAGES FOLLOWED BY ANOTHER MESSAGE ¿INTERNAL MALFUNCTION - REMAINS OPERATIONAL¿. THE S5 DOUBLE HEAD PUMP HAS BEEN TAKEN OUT OF SERVICE. REPORTEDLY, CARDIOPLEGIA PUMP STOPPED WORKING, AFTER SWITCHING PUMP OFF AND THEN ON AGAIN THE PUMP STARTED TO WORK AGAIN AS EXPECTED. THIS HAS HAPPENED ON A FEW OCCASIONS DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15347773 ↗MalfunctionAug 26, 2022Pumping Stopped; Pumping Stopped
LIVANOVA RECEIVED REPORT THAT A S5 DOUBLE ROLLER PUMP GAVE A MOTOR CONTROL FAILURE ERROR MESSAGE DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15301061 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K071318. One submission often covers many models, so check the product and lot details in each record.
- Z-1119-2020 · Jan 28, 2019 · TerminatedLivaNova Sorin Connect DPIII KOMPL. S5 REF 96-415-130, 2017-08-15, IPX1, RXOnly, (01)04033817901594 Product Usage: The S5 System is used to perform, control and monitor extracorp
- Z-2866-2016 · Aug 25, 2016 · TerminatedStckert S5 System, S5 Heart-lung machine, Cardiopulmonary bypass heart-lung machine console Product Usage: The Stckert S5 System is intended to be used during cardiopulmonary bypa
- Z-0856-2014 · Dec 4, 2013 · TerminatedS5/C5 Heart-lung machine Product Usage: The StOckert S5/ Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.
- Z-0142-2014 · Sep 24, 2013 · TerminatedS5 Heart-lung machine, Stockert S5 System. Catalogue No: 48-30/40/50-00. The Stockert S5 System is intended to be used during cardiopulmonary bypass for procedures lasting six(6)
- Z-0274-2013 · Sep 19, 2012 · TerminatedS5 Mast Roller Pump 150, Item Number: 10-88-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
- Z-0275-2013 · Sep 19, 2012 · TerminatedSorin C5 Perfusion System, Item number 58-00-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
- Z-0272-2013 · Sep 19, 2012 · TerminatedS5 Double Roller Pump 85, Item Number: 10-85-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
- Z-0271-2013 · Sep 19, 2012 · TerminatedS5 Single Roller Pump 150, Item Number: 10-80-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
- Z-0273-2013 · Sep 19, 2012 · TerminatedS5 Mast Roller Pump 85, Item Number: 10-88-60 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
- Z-1325-2012 · Feb 23, 2012 · TerminatedS5 Single Roller Pump 150 The S5 Perfusion System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six (6) hours.
- Z-1328-2012 · Feb 23, 2012 · TerminatedControl panel for mast roller pump 85 The S5 Perfusion System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six (6) hours.
- Z-1326-2012 · Feb 23, 2012 · TerminatedS5 Double Roller Pump 85 The S5 Perfusion System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six (6) hours.
- Z-1327-2012 · Feb 23, 2012 · TerminatedControl panel for mast roller pump 150/85 The S5 Perfusion System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six (6) ho
Other CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS devices
- Terumo® Advanced Perfusion System 1 801188Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass3 recalls447 reports
- MAQUET CARDIOHELP CARDIOHELP-iMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 recalls2 deaths64 reports
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls18 deaths107 reports
- Terumo® Advanced Perfusion System 1 816300Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass300 reports
- Terumo® Advanced Perfusion System 1 816571Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass230 reports
- Terumo® Advanced Perfusion System 1 801763Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 death230 reports
- ESSENZ HLMLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall1 death110 reports
- Terumo® Advanced Perfusion System 1 5773Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 recall78 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.