S5 Console 4-position (UDI 04033817900894)Recalls & reported problems

LivaNova Deutschland GmbH · Class 2

Quick summary

The S5 Console is a heart-lung machine (cardiopulmonary bypass console) made by LivaNova, which pumps and manages blood circulation and gases during open-heart surgery. Most of the 73 reports describe the pump stopping or pumping incorrectly during procedures, along with covers or clamps that would not open or close, blank displays, and unexpected shutdowns; several mention tubing clamps and a gas blender not delivering CO2 as expected. There is 1 death and 1 injury reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
65
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 65 adverse event reports that list this exact device (DI 04033817900894), received between Oct 13, 2016 and Mar 4, 2026 — classified as 1 death, 1 injury, 62 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (15 reports).

13 other recalls cover devices cleared under the same FDA submission (K071318); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 3 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 3 reportsJun 2025: 2 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Pumping Stopped15
  • Pumping Problem9
  • Difficult to Open or Close5
  • No Display/Image5
  • Appropriate Device Problem Term/Code Not Available4
  • Unexpected Shutdown3
  • Device Displays Incorrect Message2
  • Device Operational Issue2

Patient problems

  • No Clinical Signs, Symptoms or Conditions36
  • No Consequences Or Impact To Patient18
  • No Patient Involvement6
  • Insufficient Information2
  • No Known Impact Or Consequence To Patient2
  • Air Embolism1
  • Brain Injury1
  • Death1

Recent adverse event reports

The 25 most recent of 65 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 4, 2026Mechanical Jam

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24510882 ↗
  • MalfunctionMar 4, 2026Pumping Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24510619 ↗
  • MalfunctionJan 22, 2026Pumping Problem

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT AN S5 HLM ROLLER PUMP WAS NOT FUNCTIONING. THE ISSUE OCCURRED DURING PROCEDURE THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24144350 ↗
  • MalfunctionDec 19, 2025Use of Device Problem; Use of Device Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23851728 ↗
  • MalfunctionNov 25, 2025Pumping Problem; Pumping Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23645249 ↗
  • MalfunctionNov 7, 2025Low Readings; Low Readings

    LIVANOVA DEUTSCHLAND RECEIVED REPORT OF A DEFECTIVE BUBBLE SENSOR, DETECTING BUBBLES INTERMITTENTLY, WITH ERROR MESSAGE DISPLAYED ONTO THE UNIT. REPORTEDLY, ALL CONNECTIONS AND BUBBLE MODULE WERE CHECKED WITH NO ISSUE. EVEN INSTALLING THE SENSOR ON ANOTHER HLM MACHINE, THE PROBLEM PERSISTED. THE EVENT OCCURRED DURING PROCEDURE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23496794 ↗
  • MalfunctionAug 6, 2025Unexpected Shutdown; Unexpected Shutdown

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A S5 CONSOLE SWITCHED OFF SUDDENLY DURING THE PRIMING SET UP. AFTER SOMETIMES THE MACHINE WAS POWERED ON BUT IT WAS NOT WORKING ON MAINS POWER SUPPLY. WHILE RUNNING ON BATTERY THE CHARGING % WAS ALSO NOT SHOWING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22726493 ↗
  • MalfunctionAug 5, 2025Pumping Stopped; Pumping Stopped

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A S5 SYSTEM WAS DEFECTIVE DURING PRIMING. NO FURTHER INFORMATION IS AVAILABLE AT THE MOMENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22714401 ↗
  • MalfunctionJul 10, 2025Difficult to Open or Close; Difficult to Open or Close

    LIVANOVA DEUTSCHLAND HAS RECEIVED A REPORT THAT, THE S5 HLM DISPLAYED A DEFECTIVE ARTERIAL CLAMP ALARM THAT CANNOT BE CLEARED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22468412 ↗
  • MalfunctionJul 8, 2025Pumping Problem; Pumping Problem

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT RELATED TO A S5 CONSOLE. WHEN ON PUMP WITH THE ARTERIAL LINE CLAMPLED, USER WAS LOSING VOLUME AND THERE WAS NO LEAK. THEREFORE, PERFUSIONIST TOOK OFF BYPASS AND CHANGED DISPOSABLES. NO CHANGE OCCURRED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22434692 ↗
  • MalfunctionJun 20, 2025Pumping Stopped; Pumping Stopped

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT, DURING PROCEDURE, THE S5 ROLLER PUMPS USED AS ASPIRATORS TURNED OFF AND BACK ON SEVERAL TIMES, NOT SIMULTANEOUSLY, WITH THE CONSEQUENT PUMP STOP. THE CP5 DRIVER USED ON THE S5 SYSTEM AS ARTERIAL PUMP DID NOT RESPOND TO THE SET PRESSURE THRESHOLD, MAKING IT IMPOSSIBLE TO USE THE AUTOMATIC FLOW REGULATION FUNCTION. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22298910 ↗
  • MalfunctionJun 5, 2025Difficult to Open or Close; Difficult to Open or Close

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT AN OCCLUDER MOUNTED ON A S5 CONSOLE GAVE AN ERROR MESSAGE DURING PROCEDURE. NO FURTHER INFORMATION IS AVAILABLE AT THE MOMENT. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22147830 ↗
  • MalfunctionMay 26, 2025Complete Loss of Power; Complete Loss of Power

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22084732 ↗
  • MalfunctionMay 15, 2025Restricted Flow rate; Use of Device Problem; Restricted Flow rate; Use of Device Problem

    SEE INITIAL REPORT.

    Patient problems: Insufficient Information

    MAUDE report 22025592 ↗
  • InjuryMay 15, 2025Restricted Flow rate; Restricted Flow rate

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A CENTRIFUGAL PUMP (CP5) USED ON THE S5 SYSTEM SLOWED DOWN DUE TO A BIG BLOOD CLOT PRESENT INSIDE THE REVOLUTION HEAD, DURING PROCEDURE. MEDICAL TEAM THEN ELECTED TO SWITCH THE PATIENT TO ANOTHER S5 SYSTEM. REPORTEDLY, THE PATIENT OUTCOME DOES NOT LOOK GOOD. NO ADDITIONAL DETAILS ON PATIENT CONDITIONS WERE PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 22025557 ↗
  • MalfunctionDec 2, 2024Difficult to Open or Close; Difficult to Open or Close

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20818160 ↗
  • MalfunctionNov 7, 2024Unexpected Shutdown; Unexpected Shutdown

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20628688 ↗
  • MalfunctionOct 28, 2024Pumping Stopped; Pumping Stopped

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A S5 PUMP STOPPED FUNCTIONING WITHOUT TRIGGERING ANY ALARM ALERTS. THE ISSUE OCCURRED DURING A PROCEDURE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20544702 ↗
  • MalfunctionOct 18, 2024Pumping Stopped; Pumping Stopped

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT, DURING A PROCEDURE, THREE (3) DIFFERENT ROLLER PUMPS USED AS SUCKER DISPLAYED MOTOR CONTROL ERROR (E441). THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20480318 ↗
  • MalfunctionOct 4, 2024Pumping Stopped; Pumping Stopped

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20366102 ↗
  • MalfunctionSep 12, 2024Electrical Shorting; Electrical Shorting

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT WHEN THE SORIN S5 SYSTEM IS PLUGGED IN, IT TRIGGER THE OPERATION ROOM ALARM. THE ISSUE OCCURRED DURING MAINTENANCE AND THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20201489 ↗
  • MalfunctionAug 21, 2024Pumping Problem; Pumping Problem

    LIVANOVA RECEIVED A REPORT ABOUT A PUMP ALARM "FAULT IN MOTOR CONTROLLER (431)". EVEN AFTER SWITCHING OFF/ON MULTIPLE TIMES - PUMP NOT REVOLVING. NO PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20023638 ↗
  • MalfunctionJun 18, 2024Defective Alarm; Defective Alarm

    LIVANOVA DEUTSCHLAND HAS RECEIVED A REPORT THAT A S5 PUMP DISPLAYED AN ERROR RELATED TO ENCODER MALFUNCTION. THE EVENT OCCURRED DURING A PROCEDURE. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19557045 ↗
  • MalfunctionMay 7, 2024Pumping Problem; Pumping Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19257353 ↗
  • MalfunctionMay 6, 2024Pumping Stopped; Pumping Stopped

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT THE S5 CONSOLE FOR 4 PUMP WAS NOT WORKING PROPERLY DURING SET UP. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19246684 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K071318. One submission often covers many models, so check the product and lot details in each record.

  • Z-1119-2020 · Jan 28, 2019 · Terminated
    LivaNova Sorin Connect DPIII KOMPL. S5 REF 96-415-130, 2017-08-15, IPX1, RXOnly, (01)04033817901594 Product Usage: The S5 System is used to perform, control and monitor extracorp
  • Z-2866-2016 · Aug 25, 2016 · Terminated
    Stckert S5 System, S5 Heart-lung machine, Cardiopulmonary bypass heart-lung machine console Product Usage: The Stckert S5 System is intended to be used during cardiopulmonary bypa
  • Z-0856-2014 · Dec 4, 2013 · Terminated
    S5/C5 Heart-lung machine Product Usage: The StOckert S5/ Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.
  • Z-0142-2014 · Sep 24, 2013 · Terminated
    S5 Heart-lung machine, Stockert S5 System. Catalogue No: 48-30/40/50-00. The Stockert S5 System is intended to be used during cardiopulmonary bypass for procedures lasting six(6)
  • Z-0274-2013 · Sep 19, 2012 · Terminated
    S5 Mast Roller Pump 150, Item Number: 10-88-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
  • Z-0275-2013 · Sep 19, 2012 · Terminated
    Sorin C5 Perfusion System, Item number 58-00-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
  • Z-0273-2013 · Sep 19, 2012 · Terminated
    S5 Mast Roller Pump 85, Item Number: 10-88-60 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
  • Z-0271-2013 · Sep 19, 2012 · Terminated
    S5 Single Roller Pump 150, Item Number: 10-80-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
  • Z-0272-2013 · Sep 19, 2012 · Terminated
    S5 Double Roller Pump 85, Item Number: 10-85-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
  • Z-1327-2012 · Feb 23, 2012 · Terminated
    Control panel for mast roller pump 150/85 The S5 Perfusion System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six (6) ho
  • Z-1328-2012 · Feb 23, 2012 · Terminated
    Control panel for mast roller pump 85 The S5 Perfusion System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six (6) hours.
  • Z-1326-2012 · Feb 23, 2012 · Terminated
    S5 Double Roller Pump 85 The S5 Perfusion System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six (6) hours.
  • Z-1325-2012 · Feb 23, 2012 · Terminated
    S5 Single Roller Pump 150 The S5 Perfusion System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six (6) hours.

Other CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS devices

All CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.