Skip to content

S5 Console 5-position (UDI 04033817900900)Recalls & reported problems

LivaNova Deutschland GmbH · Class 2

All models of S5 Console (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
2
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 04033817900900), received between Mar 20, 2018 and Feb 20, 2026 — classified as 2 deaths, 14 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (3 reports).

13 other recalls cover devices cleared under the same FDA submission (K071318); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image3
  • No Apparent Adverse Event2
  • Pumping Stopped2
  • Thermal Decomposition of Device2
  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1
  • Device Displays Incorrect Message1
  • Electrical Shorting1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient6
  • Air Embolism2
  • Embolism/Embolus1
  • No Patient Involvement1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 20, 2026Unstable

    see initial report.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24398519 ↗
  • MalfunctionMay 31, 2025Low Readings; Low Readings

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22120428 ↗
  • MalfunctionFeb 20, 2025Pumping Stopped; Pumping Stopped

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21421567 ↗
  • DeathAug 26, 2024No Apparent Adverse Event; No Apparent Adverse Event

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT FROM AEMPS OF A DEATH CASE IN WHICH A S5 SYSTEM WAS USED. CUSTOMER HAS NOT NOTIFIED DIRECTLY LIVANOVA OF THE EVENT WHEN IT OCCURRED, AS HE DID NOT CONSIDER THE S5 DEVICE TO BE INVOLVED.

    Patient problems: Air Embolism

    MAUDE report 20065346 ↗
  • MalfunctionJul 10, 2024Pumping Problem; Pumping Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19708748 ↗
  • MalfunctionJan 16, 2024Pumping Stopped; Pumping Stopped

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A S5 ROLLER PUMP GAVE A MOTOR CONTROL FAILURE ERROR MESSAGE DURING PROCEDURE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18516170 ↗
  • MalfunctionNov 18, 2022Electrical Shorting; Electrical Shorting

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A S5 PUMP STARTED SMOKING DURING A CASE. SIMULTANEOUSLY, A ¿POP¿ WAS HEARD. THE PUMP TRANSITIONED TO BATTERY. THE PERFUSIONIST UNPLUGGED THE PUMP AND RAN ON BATTERY FOR THE REMAINDER OF THE CASE. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15824075 ↗
  • DeathSep 2, 2022Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event

    LIVANOVA RECEIVED REPORT THAT A PATIENT WAS INJURED AND A S5 SYSTEM WAS USED. ACCORDING TO THE MANAGEMENT OF CARDIOTECHNOLOGY, THE TUBE WAS INSERTED IN THE WRONG DIRECTION AND AIR WAS PUMPED INTO THE PATIENT. THE PATIENT DIED. REPORTEDLY, THE MACHINE HAS BEEN TAKEN OUT OF SERVICE. NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND THE EVENT WERE PROVIDED.

    Patient problems: Air Embolism; Embolism/Embolus

    MAUDE report 15347776 ↗
  • MalfunctionDec 17, 2021Failure to Auto Stop; Failure to Auto Stop

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13023921 ↗
  • MalfunctionMay 7, 2019No Display/Image; No Display/Image

    LIVANOVA (B)(4) RECEIVED A REPORT THAT SEVERAL VALUES WHICH WERE DISPLAYED ON THE S5 SYSTEM SUDDENLY DISAPPEARED DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8585964 ↗
  • MalfunctionApr 16, 2019Thermal Decomposition of Device; Appropriate Device Problem Term/Code Not Available; Thermal Decomposition of Device; Appropriate Device Problem Term/Code Not Available

    SEE INITIAL REPORT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8517089 ↗
  • MalfunctionMar 21, 2019Overheating of Device; Overheating of Device

    LIVANOVA (B)(4) RECEIVED A REPORT THAT A BURNING SMELL WAS NOTICED DURING THE USE OF A S5 SYSTEM. IT WAS FURTHER REPORTED THAT AFTER THE BURNING SMELL, AN ERROR MESSAGE WAS DISPLAYED ON THE S5 SYSTEM. THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8439100 ↗
  • MalfunctionNov 2, 2018Loss of Power; Loss of Power

    LIVANOVA (B)(4) RECEIVED A REPORT THAT A S5 SYSTEM STOPPED WORKING DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8031845 ↗
  • MalfunctionJul 16, 2018No Display/Image; No Display/Image

    LIVANOVA (B)(4) RECEIVED A REPORT THAT A S5 SYSTEM PANEL DOES NOT POWER ON DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7688959 ↗
  • MalfunctionMar 20, 2018No Display/Image; No Display/Image

    LIVANOVA (B)(4) RECEIVED A REPORT THAT A ALL DISPLAY AND CONTROL MODULES OF A S5 SYSTEM WENT BLANK DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 7352402 ↗
  • MalfunctionMar 20, 2018Thermal Decomposition of Device; Device Displays Incorrect Message; Thermal Decomposition of Device; Device Displays Incorrect Message

    LIVANOVA (B)(4) RECEIVED A REPORT THAT A S5 SYSTEM DISPLAYED AN ALARM AND THE CUSTOMER REPORTED BURNING SMELL AFTER THE ALARM OCCURRED DURING PROCEDURE. IN ADDITION THE BATTERY LEVEL INDICATOR AND PERCENTAGE BECAME BLANK THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7352385 ↗

FDA clearance and classification

Cleared through Special 510(k) K071318 on Jul 6, 2007 (decision: Substantially Equivalent). Submitted by Sorin Group Deutschland GmbH as “Modification To: Stockert S5 System”. The FDA review took 57 days. FDA record ↗

The FDA regulates product code DTQ (console, heart-lung machine, cardiopulmonary bypass) under 21 CFR 870.4220 as a Class II device. 21 CFR 870.4220 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K071318. One submission often covers many models, so check the product and lot details in each record.

  • Z-1119-2020 · Jan 28, 2019 · Terminated
    LivaNova Sorin Connect DPIII KOMPL. S5 REF 96-415-130, 2017-08-15, IPX1, RXOnly, (01)04033817901594 Product Usage: The S5 System is used to perform, control and monitor extracorp
  • Z-2866-2016 · Aug 25, 2016 · Terminated
    Stckert S5 System, S5 Heart-lung machine, Cardiopulmonary bypass heart-lung machine console Product Usage: The Stckert S5 System is intended to be used during cardiopulmonary bypa
  • Z-0856-2014 · Dec 4, 2013 · Terminated
    S5/C5 Heart-lung machine Product Usage: The StOckert S5/ Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.
  • Z-0142-2014 · Sep 24, 2013 · Terminated
    S5 Heart-lung machine, Stockert S5 System. Catalogue No: 48-30/40/50-00. The Stockert S5 System is intended to be used during cardiopulmonary bypass for procedures lasting six(6)
  • Z-0273-2013 · Sep 19, 2012 · Terminated
    S5 Mast Roller Pump 85, Item Number: 10-88-60 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
  • Z-0275-2013 · Sep 19, 2012 · Terminated
    Sorin C5 Perfusion System, Item number 58-00-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
  • Z-0271-2013 · Sep 19, 2012 · Terminated
    S5 Single Roller Pump 150, Item Number: 10-80-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
  • Z-0272-2013 · Sep 19, 2012 · Terminated
    S5 Double Roller Pump 85, Item Number: 10-85-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
  • Z-0274-2013 · Sep 19, 2012 · Terminated
    S5 Mast Roller Pump 150, Item Number: 10-88-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
  • Z-1328-2012 · Feb 23, 2012 · Terminated
    Control panel for mast roller pump 85 The S5 Perfusion System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six (6) hour
  • Z-1325-2012 · Feb 23, 2012 · Terminated
    S5 Single Roller Pump 150 The S5 Perfusion System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six (6) hours.
  • Z-1326-2012 · Feb 23, 2012 · Terminated
    S5 Double Roller Pump 85 The S5 Perfusion System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six (6) hours.
  • Z-1327-2012 · Feb 23, 2012 · Terminated
    Control panel for mast roller pump 150/85 The S5 Perfusion System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six (6)

Common questions about the S5 Console 5-position

Has the S5 Console 5-position been recalled?

No FDA recall record in our data names this exact device identifier (DI 04033817900900), based on the FDA recall file dated Nov 27, 2026. However, 13 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the S5 Console 5-position?

The FDA's MAUDE database lists 16 adverse event reports for this device, received between Mar 20, 2018 and Feb 20, 2026. In those reports, the most common specific device problems are “No Display/Image” (3), “Pumping Stopped” (2) and “Thermal Decomposition of Device” (2); the most common patient problems are “Air Embolism” (2) and “Embolism/Embolus” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the S5 Console 5-position?

2 reports are classified as deaths, 0 as injuries and 14 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the S5 Console 5-position FDA approved?

It was cleared through Special 510(k) K071318 on Jul 6, 2007. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the S5 Console 5-position, and is it still sold?

The FDA's device database lists LivaNova Deutschland GmbH as the company responsible for it (brand S5 Console, model 5-position). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS devices

All CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page