SCPC w Level/Bubble & PressureRecalls & reported problems

LivaNova Deutschland GmbH · Class 2

All models of SCPC (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
20
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 04033817901020), received between Jan 5, 2017 and Sep 5, 2025 — classified as 1 death, 1 injury, 18 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 3 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Pumping Stopped5
  • Unexpected Shutdown4
  • No Display/Image3
  • Pumping Problem3
  • Decreased Pump Speed1
  • Device Sensing Problem1
  • Display or Visual Feedback Problem1
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • No Consequences Or Impact To Patient3
  • No Patient Involvement3
  • Cardiac Arrest1
  • Nervous System Injury1
  • No Information1

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 5, 2025Display or Visual Feedback Problem; Pumping Stopped; Display or Visual Feedback Problem; Pumping Stopped

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A SCPC GAVE BLANK BLUE SCREEN AND A ALARMING TONE DURING PRIMING AND SET UP. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22979351 ↗
  • MalfunctionMay 20, 2025Low Readings; Low Readings

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A PATIENT SENSOR OF A SCPC DID NOT READING ANYTHING. NO FURTHER INFORMATION IS AVAILABLE. IT CANNOT BE EXCLUDED THAT THE AFFECTED SENSOR IS A BUBBLE ONE AND THAT DID NOT DETECT AIR BUBBLE. THE ISSUE OCCURRED DURING PROCEDURE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22050829 ↗
  • MalfunctionMay 16, 2025Pumping Stopped; Pumping Stopped

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22036607 ↗
  • MalfunctionMay 7, 2025Unexpected Shutdown; Unexpected Shutdown

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT RELATED TO A SCPC DOWN DURING PROCEDURE. NO FURTHER INFORMATION IS AVAILABLE AT THE MOMENT. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21976991 ↗
  • MalfunctionApr 30, 2025Pumping Problem; Pumping Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21932974 ↗
  • MalfunctionMar 24, 2025Unexpected Shutdown; Unexpected Shutdown

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A SCPC (STOCKERT CENTRIFUGAL PUMP SYSTEM) DID TURN OFF WHEN UNPLUGGED FROM THE MAINS. THE BATTERY DID NOT WORK. THE ISSUE OCCURRED POST PROCEDURE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21675232 ↗
  • DeathMay 3, 2024Pumping Stopped; Pumping Stopped

    LIVANOVA HAS RECEIVED A REPORT THAT THE SCPC PUMP STOPPED DURING THE SERVICE, THE SCREEN PRESENTED AS THE INTIAL SCREEN. AFTER RESTARTED, THE PUMP RUN FOR A WHILE AND THE EQUIPMENT STOPPED AGAIN. THE OPERATION WAS COMPLETED WITH CP5 AS A REPLACEMENT. ACCORIND TO INFORMATION, THE EQUIPMENT FAILURE CAUSED ARTERIAL BLOOD REGURGITATION, AND CARDIOPULMONARY RESUSCITATION WAS PERFORMED. PLEASE CREATE A NEW CUSTOMER IN ETQ. THANK YOU! HOSPITAL: (B)(6), ADDRESS: (B)(6).

    Patient problems: Cardiac Arrest; Nervous System Injury

    MAUDE report 19236608 ↗
  • MalfunctionAug 8, 2023Device Sensing Problem; Pumping Problem; Device Sensing Problem; Pumping Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17480825 ↗
  • MalfunctionSep 29, 2022Pumping Stopped; Pumping Stopped

    LIVANOVA RECEIVED REPORT THAT A SCP DRIVE UNIT COULD BE CONTROLLED BY THE SCP CONTROL PANEL FOR ABOUT 20 SECONDS AND THEN THE DRIVE UNIT WOULD ROTATE TO MAXIMUM RPM, THEN DROP TO 0 RPM. IN ADDITION AN ERROR CODE ASSOCIATED WITH A DIFFERENCE BETWEEN THE SET VALUE AND THE ACTUAL VALUE MORE THAN 1000 RPM WAS DISPLAYED. THE ISSUE WAS IDENTIFIED DURING MAINTENANCE AND THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15508886 ↗
  • MalfunctionSep 1, 2022Pumping Stopped; Pumping Stopped

    LIVANOVA RECEIVED REPORT OF A MECHANICAL FAILURE OF A CENTRIFUGAL PUMP CONSOLE NOT GENERATING FORWARD FLOW DURING ECMO SUPPORT. THERE WAS NO PATIENT INJURY. ON AUGUST 23, 2022 LIVANOVA RECEIVED A MEDWATCH NOTIFICATION (B)(4). WITHIN THE REPORT IT WAS STATED THE FOLLOWING: PATIENT ON VA ECMO FOR PULMONARY HYPERTENSION (PHTN) LISTED AND DELISTED FOR HEART/LUNG TRANSPLANT DURING HER COURSE OF ECMO RUN. ECMO CANNULAS PLACED IN RIGHT INTERNAL JUGULAR (RIJ) AND RIGHT COMMON CAROTID ARTERY (RCCA). SHE IS ACTIVE AND VERBAL. SHE REFUSES NECK COLLAR WHILE IN BED. SHE SEEMS AGITATED IN THE LAST COUPLE OF DAYS AND HASN'T BEEN ABLE TO SLEEP. SHE HAD TWO PRIOR CIRCUIT EVENTS WHICH SHE REQUIRED A PART OF HER CIRCUIT REPLACED ONCE AND A WHOLE CIRCUIT CHANGE FOR THE SECOND EVENT BOTH RELATED TO AIR BUBBLE ALARM THAT HAS INTERRUPTED FORWARD FLOW. ON BOTH EVENTS, NO CIRCUIT AIR WERE NOTED BY ECMO SPECIALISTS. BEDSIDE RN HAD NOTED DAMPENING OF BLOOD PRESSURE TRACING AND WHILE ADJUSTING ARTERIAL LINE POSITION ON HER WRIST, SHE FOUND PATIENT UNRESPONSIVE. ECMO SPECIALISTS, AT THE SAME TIME DURING HER HOURLY CIRCUIT ASSESSMENT AND CANNULA CHECK NOTICED PATIENT UNRESPONSIVE AS WELL. THIS HAD OCCURRED AFTE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15335763 ↗
  • MalfunctionJul 7, 2022Unexpected Shutdown; Unexpected Shutdown

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A CENTRIFUGAL PUMP CONSOLE DID NOT DISPLAY BATTERY INFORMATION AND SHUT DOWN WHEN UNPLUGGED. THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14942276 ↗
  • MalfunctionJan 5, 2022Pumping Problem; Pumping Problem

    LIVANOVA RECEIVED REPORT ABOUT A CONTROL KNOB NOT WORKING SOMETIMES. NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13168178 ↗
  • MalfunctionDec 13, 2021Decreased Pump Speed; Decreased Pump Speed

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12983546 ↗
  • MalfunctionJun 25, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    LIVANOVA (B)(4) RECEIVED A REPORT THAT A CENTRIFUGAL PUMP CONSOLE STOPPED MEASURING THE PRESSURE AND AN ERROR MESSAGE WAS DISPLAYED DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8729955 ↗
  • MalfunctionJun 7, 2019No Display/Image; No Display/Image

    SEE INITIAL REPORT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8678256 ↗
  • InjuryMay 28, 2019Unexpected Shutdown; Unexpected Shutdown

    SEE INITIAL REPORT.

    Patient problems: No Information

    MAUDE report 8645323 ↗
  • MalfunctionMar 13, 2018Loss of Power; Loss of Power

    LIVANOVA (B)(4) RECEIVED A REPORT THAT A CENTRIFUGAL PUMP CONSOLE LOST POWER DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7333856 ↗
  • MalfunctionDec 20, 2017No Display/Image; No Display/Image

    LIVANOVA (B)(4) RECEIVED A REPORT THAT THE DISPLAY OF THE CENTRIFUGAL PUMP CONSOLE BECAME BLANK AND THE BACKGROUND TURNED BLUE DURING SETUP. THE DEVICE WAS RESTARTED AND BEGAN TO FUNCTION AS EXPECTED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 7135926 ↗
  • MalfunctionMay 5, 2017No Display/Image; No Display/Image

    LIVANOVA (B)(4) RECEIVED A REPORT THAT THE DISPLAY OF THE CENTRIFUGAL PUMP CONSOLE WAS BLANK DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 6549754 ↗
  • MalfunctionJan 5, 2017Overheating of Device; Overheating of Device

    SORIN GROUP (B)(4) RECEIVED A REPORT THAT THE CHASSIS OF THE SORIN CENTRIFUGAL PUMP CONSOLE OVERHEATED DURING THE BATTERY DISCHARGE TEST DURING MAINTENANCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 6227450 ↗

FDA clearance and classification

Cleared through 510(k) K020571 on Sep 23, 2002 (decision: Substantially Equivalent). Submitted by Stockert Instrumente GmbH as “Stockert Centrifugal Pump Console”. The FDA review took 214 days. FDA record ↗

The FDA regulates product code DTQ (console, heart-lung machine, cardiopulmonary bypass) under 21 CFR 870.4220 as a Class II device. 21 CFR 870.4220 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the SCPC w Level/Bubble & Pressure

Has the SCPC w Level/Bubble & Pressure been recalled?

No FDA recall record in our data names this exact device identifier (DI 04033817901020), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the SCPC w Level/Bubble & Pressure?

The FDA's MAUDE database lists 20 adverse event reports for this device, received between Jan 5, 2017 and Sep 5, 2025. In those reports, the most common specific device problems are “Pumping Stopped” (5), “Unexpected Shutdown” (4) and “No Display/Image” (3); the most common patient problems are “Cardiac Arrest” (1) and “Nervous System Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SCPC w Level/Bubble & Pressure?

1 report is classified as a death, 1 as an injury and 18 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SCPC w Level/Bubble & Pressure FDA approved?

It was cleared through 510(k) K020571 on Sep 23, 2002. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the SCPC w Level/Bubble & Pressure, and is it still sold?

The FDA's device database lists LivaNova Deutschland GmbH as the company responsible for it (brand SCPC, model w Level/Bubble & Pressure). Its record shows it as in commercial distribution.