Accessories Bubble Sensor 3/16“ x 1/16”Recalls & reported problems
LivaNova Deutschland GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04033817902423), received between Aug 21, 2024 and Sep 3, 2024 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Low Readings” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K232291); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Low Readings2
- Defective Alarm1
- Defective Device1
Patient problems
- Cardiovascular Dysfunction/Insufficiency2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjurySep 3, 2024Low Readings; Low Readings
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A CARDIOPLEGIA BUBBLE SENSOR ALARMED FIVE (5) DIFFERENT TIMES DURING THE INITIAL DOSE OF CARDIOPLEGIA. NO AIR WAS OBSERVED IN ALL INSTANCES AND THE ALARM WAS CLEARED. THE PATIENT WAS PLACED ON A RVAD DEVICE POST CARDIOPULMONARY BYPASS, PURPORTEDLY DUE TO MYOCARDIAL PROTECTION.
Patient problems: Cardiovascular Dysfunction/Insufficiency
MAUDE report 20125726 ↗InjurySep 3, 2024Low Readings; Low Readings
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A CARDIOPLEGIA BUBBLE SENSOR ALARMED FIVE (5) DIFFERENT TIMES DURING THE INITIAL DOSE OF CARDIOPLEGIA. NO AIR WAS OBSERVED IN ALL INSTANCES AND THE ALARM WAS CLEARED. THE PATIENT WAS PLACED ON A RVAD DEVICE POST CARDIOPULMONARY BYPASS, PURPORTEDLY DUE TO MYOCARDIAL PROTECTION.
Patient problems: Cardiovascular Dysfunction/Insufficiency
MAUDE report 20125721 ↗MalfunctionAug 21, 2024Defective Alarm; Defective Alarm
LIVANOVA RECEIVED A REPORT ABOUT BUBBLE SENSOR ALARM NOT TRIGGERED. AT THE REMOVAL OF CROSS CLAMP AND DECANNULATING OF THE AORTIC ROOT VENT, THE CARDIOPLEGIA LINE WAS DRAINED BACK OF BLOOD VIA GRAVITY. WHEN THE AIR/BLOOD INTERFACE PASSED THROUGH THE CARDIOPLEGIA BUBBLE SENSOR, THE ALARM DID NOT ACTIVATE. THE SENSOR WAS INSPECTED, WITH THE LATCH OF THE SENSOR THEN OPENED, BUT STILL, THE ALARM DID NOT GO OFF. IT WAS NOT UNTIL THE BUBBLE SENSOR WAS TURNED OFF AND BACK ON, THAT THE ALARM ACTIVATED AND THE ORANGE AUDIBLE AND VISUAL ALARM WAS OCCURRED. THE PRE-BYPASS CHECKS OF ALL SENSORS WAS COMPLETED, AND THE ALARM PERFORMED AS IT SHOULD. THERE WAS NOT PATIENT IMPACT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20019977 ↗MalfunctionAug 21, 2024Defective Device; Defective Device
SEE INITIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20018793 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K232291. One submission often covers many models, so check the product and lot details in each record.
- Z-0571-2025 · Oct 18, 2024 · Open, ClassifiedEssenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software
Other CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS devices
- Terumo® Advanced Perfusion System 1 801188Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass3 recalls447 reports
- MAQUET CARDIOHELP CARDIOHELP-iMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 recalls2 deaths64 reports
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls18 deaths107 reports
- Terumo® Advanced Perfusion System 1 816300Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass300 reports
- Terumo® Advanced Perfusion System 1 801763Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 death230 reports
- Terumo® Advanced Perfusion System 1 816571Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass230 reports
- ESSENZ HLMLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall1 death110 reports
- Terumo® Advanced Perfusion System 1 5773Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 recall78 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page