Accessories Bubble Sensor 3/8“ x 3/32“Recalls & reported problems
LivaNova Deutschland GmbH · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04033817902447), received between May 7, 2024 and May 16, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Low Readings” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K232291); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Low Readings3
- Break1
- Failure to Auto Stop1
- Under-Sensing1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 16, 2025Failure to Auto Stop; Failure to Auto Stop
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A BUBBLE SENSOR DETECTOR ALARMED AND STOPPED THE PUMP, CORRECTLY. HOWEVER, THE SECOND TIME IT ALARMED, A FEW MINUTES AFTER THE PREVIOUS ALARM, IT SOUNDED BUT DID NOT STOP THE PUMP. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22036143 ↗MalfunctionFeb 18, 2025Low Readings; Low Readings
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A BUBBLE SENSOR DETECTOR DID NOT WORK DURING PROCEDURE. NO ADDITIONAL INFORMATION IS AVAILABLE. IT CANNOT BE EXCLUDED THAT BUBBLE SENSOR DID NOT DETECT AIR BUBBLE. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21393398 ↗MalfunctionSep 10, 2024Break; Low Readings; Break; Low Readings
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A BUBBLE SENSOR DETECTOR WAS DEFECTIVE DURING PROCEDURE. NO ADDITIONAL INFORMATION IS AVAILABLE. IT CANNOT BE EXCLUDED THAT BUBBLE SENSOR DID NOT DETECT AIR BUBBLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20178094 ↗MalfunctionMay 7, 2024Under-Sensing; Low Readings; Under-Sensing; Low Readings
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT, THE 3/8 ARTERIAL BUBBLE SENSOR WAS NOT TRIGGERING ALARM WHEN TESTED (AIR IN TUBE) DURING IN SERVICE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19257756 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K232291. One submission often covers many models, so check the product and lot details in each record.
- Z-0571-2025 · Oct 18, 2024 · Open, ClassifiedEssenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software
Other CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS devices
- Terumo® Advanced Perfusion System 1 801188Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass3 recalls447 reports
- MAQUET CARDIOHELP CARDIOHELP-iMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 recalls2 deaths64 reports
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls18 deaths107 reports
- Terumo® Advanced Perfusion System 1 816300Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass300 reports
- Terumo® Advanced Perfusion System 1 816571Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass230 reports
- Terumo® Advanced Perfusion System 1 801763Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 death230 reports
- ESSENZ HLMLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall1 death110 reports
- Terumo® Advanced Perfusion System 1 5773Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 recall78 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.