Skip to content

S5 Console CP5 ConsoleRecalls & reported problems

LivaNova Deutschland GmbH · Class 2

All models of S5 Console (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04033817903123), received between Oct 26, 2022 and Nov 14, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Pumping Stopped2
  • Decreased Pump Speed1
  • Failure to Pump1
  • Pumping Problem1
  • Unexpected Shutdown1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2025Pumping Problem; Pumping Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23556377 ↗
  • MalfunctionDec 7, 2023Decreased Pump Speed; Decreased Pump Speed

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT DURING PROCEDURE SPEED OF A CP5 CONSOLE DROPPED TO 1000 RPM, WHICH IS THE MINIMUM THRESHOLD WITH NO ALARM SHOWING. SPEED WAS THEN TRIED TO BE INCREASED, HOWEVER THERE WAS NO RESPONSE FROM THE CP5 CONTROL PANEL. SYSTEM WAS REBOOTED AND THE ISSUE WAS STILL PRESENT. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18281758 ↗
  • MalfunctionApr 17, 2023Pumping Stopped; Unexpected Shutdown; Pumping Stopped; Unexpected Shutdown

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16750551 ↗
  • MalfunctionNov 22, 2022Pumping Stopped; Pumping Stopped

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15843950 ↗
  • MalfunctionOct 26, 2022Failure to Pump; Failure to Pump

    LIVANOVA RECEIVED REPORT OF PUMP MOTOR HEAD MALFUNCTION ON A CP5 CONSOLE. THE ISSUE OCCURRED DURING SETUP AND THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15673235 ↗

FDA clearance and classification

Cleared through Special 510(k) K210130 on Apr 5, 2021 (decision: Substantially Equivalent). Submitted by Livanova Deutschland, GmbH as “Stockert S5 System”. The FDA review took 76 days. FDA record ↗

The FDA regulates product code DTQ (console, heart-lung machine, cardiopulmonary bypass) under 21 CFR 870.4220 as a Class II device. 21 CFR 870.4220 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the S5 Console CP5 Console

Has the S5 Console CP5 Console been recalled?

No FDA recall record in our data names this exact device identifier (DI 04033817903123), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the S5 Console CP5 Console?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Oct 26, 2022 and Nov 14, 2025. In those reports, the most common specific device problems are “Pumping Stopped” (2), “Decreased Pump Speed” (1) and “Failure to Pump” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the S5 Console CP5 Console?

None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the S5 Console CP5 Console FDA approved?

It was cleared through Special 510(k) K210130 on Apr 5, 2021. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the S5 Console CP5 Console, and is it still sold?

The FDA's device database lists LivaNova Deutschland GmbH as the company responsible for it (brand S5 Console, model CP5 Console). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS devices

All CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page