Skip to content

Setrox S 45 350973Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Setrox S 45 pacemaker lead has 200 adverse event reports, including 1 death and 82 injuries. Reports describe electrical noise and lead problems such as dislodgement and failure to pace or sense, and some describe infections that led to removal. There are no recalls that name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
201
Deaths reported
1
Injuries reported
83
Recalls naming it
0

Summary

The FDA's MAUDE database contains 201 adverse event reports that list this exact device (DI 04035479087944), received between Jun 20, 2018 and Sep 25, 2026 — classified as 1 death, 83 injuries, 117 malfunctions.

The most frequently reported specific device problem is “Ambient Noise Problem” (40 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem77
  • Ambient Noise Problem40
  • Device Dislodged or Dislocated29
  • Failure to Capture17
  • High Capture Threshold16
  • Device Sensing Problem10
  • High impedance10
  • Fracture7

Patient problems

  • Unspecified Infection69
  • No Clinical Signs, Symptoms or Conditions64
  • No Known Impact Or Consequence To Patient55
  • Pocket Erosion7
  • Discomfort2
  • Endocarditis2
  • Chest Pain1
  • Fall1

Recent adverse event reports

The 25 most recent of 201 reports, in the reporter's own words as published by the FDA.

  • InjurySep 25, 2026Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905181 ↗
  • InjuryMar 24, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 24673991 ↗
  • InjuryMar 5, 2026Impedance Problem

    THIS RA LEAD WAS EXPLANTED DUE TO STRONG, RAPID DEVIATIONS OF LEAD IMPEDANCE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS REPORT WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24515512 ↗
  • InjuryAug 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 22900020 ↗
  • MalfunctionAug 11, 2025Fracture; Fracture

    THIS LEAD WAS EXPLANTED DUE TO FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22773265 ↗
  • MalfunctionJul 30, 2025Failure to Capture; Device Dislodged or Dislocated; Failure to Capture; Device Dislodged or Dislocated

    THIS LEAD WAS CAPPED DUE TO DISLODGMENT WITH NO CAPTURE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22669464 ↗
  • MalfunctionMay 23, 2025Low impedance; Device Sensing Problem; Unstable Capture Threshold; Low impedance; Device Sensing Problem; Unstable Capture Threshold

    LEAD WAS EXPLANTED DUE TO IMPEDANCE LESS THAN 100 OHMS, DEGRADED SENSING AND THRESHOLD. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22077868 ↗
  • MalfunctionMay 9, 2025Ambient Noise Problem; Ambient Noise Problem

    LEAD WAS EXPLANTED DUE TO NOISE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21997772 ↗
  • MalfunctionJul 31, 2024Ambient Noise Problem; Ambient Noise Problem

    HOME MONITORING REPORTED EPISODE SHOWING NOISE ON ATRIAL LEAD. LEAD TRENDS APPEAR STABLE ALTHOUGH SENSING HAS BEEN TRENDING LOW. ADVISED TO EVALUATE LEAD FURTHER. LEAD CURRENTLY REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19878260 ↗
  • InjuryNov 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 18069761 ↗
  • InjuryApr 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO MYOCARDITIS AND INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Myocarditis; Unspecified Infection

    MAUDE report 16684702 ↗
  • MalfunctionOct 31, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    LEAD WAS CAPPED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15699460 ↗
  • InjurySep 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 15341701 ↗
  • InjuryAug 31, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS LEAD WAS CAPPED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 15326723 ↗
  • MalfunctionAug 8, 2022High Capture Threshold; High Capture Threshold

    THIS RA LEAD WAS EXPLANTED AND REPLACED DUE TO MODERATELY HIGH THRESHOLDS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15183124 ↗
  • MalfunctionJul 21, 2022Over-Sensing; Ambient Noise Problem; Over-Sensing; Ambient Noise Problem

    NOISE AND RV OVERSENSING SEEN ON RA SIGNAL IN RECORDINGS AND AT IN PERSON FOLLOW UP. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15075722 ↗
  • MalfunctionJul 20, 2022Failure to Capture; Failure to Capture

    THIS LEAD WAS CAPPED DUE TO LOSS OF CAPTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15070215 ↗
  • InjuryJul 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 14986758 ↗
  • MalfunctionJun 22, 2022Low impedance; Ambient Noise Problem; Low impedance; Ambient Noise Problem

    IMPEDANCE OF LESS THAN 200 OMS WAS REPORTED VIA THE HMSC. SOME NOISE CAN BE SEEN VERY INTERMITTENTLY ON A PERIODIC IEGM FROM MAY 2022. A FULL LEAD EVALUATION WAS RECOMMENDED. LEAD REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14784006 ↗
  • InjuryJun 22, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 14783495 ↗
  • MalfunctionJun 15, 2022Low impedance; Low impedance

    THIS LEAD WAS EXPLANTED DUE TO SUDDEN DROP IN RA LEAD IMPEDANCES. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14703782 ↗
  • InjuryJun 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 14694652 ↗
  • InjuryMay 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 14402964 ↗
  • MalfunctionApr 29, 2022Fracture; Low impedance; Fracture; Low impedance

    THIS RV LEAD WAS DISPLAYING LOW IMPEDANCES. UPON X-RAY, A FRACTURE WAS CLEARLY VISIBLE. LEAD WAS CAPPED AND REPLACED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14254943 ↗
  • MalfunctionApr 28, 2022Ambient Noise Problem; Ambient Noise Problem

    NOISE REPORTED ON THE ATRIAL CHANNEL IN MODE SWITCH RECORDINGS. NOISE WAS REPRODUCIBLE DURING ISOMETRICS. LEAD CURRENTLY REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14246916 ↗

Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices

All Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page