Setrox S 53 350974Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Setrox pacemaker lead appears in 870 adverse event reports, including 1 death and 332 injuries. Many describe electrical noise in heart readings, high stimulation thresholds, and leads moving or breaking, with some mentioning infections that led to removal of the whole system. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
873
Deaths reported
1
Injuries reported
335
Recalls naming it
0

Summary

The FDA's MAUDE database contains 873 adverse event reports that list this exact device (DI 04035479087951), received between Feb 16, 2016 and Sep 17, 2026 — classified as 1 death, 335 injuries, 537 malfunctions.

The most frequently reported specific device problem is “Ambient Noise Problem” (188 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 2 reportsJan 2024: 4 reportsFeb 2024: 4 reportsMar 2024: 5 reportsApr 2024: 4 reportsMay 2024: 5 reportsJun 2024: 5 reportsJul 2024: 2 reportsAug 2024: 6 reportsSep 2024: 2 reportsOct 2024: 3 reportsNov 2024: 7 reportsDec 2024: 4 reportsJan 2025: 5 reportsFeb 2025: 3 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 3 reportsJun 2025: 4 reportsJul 2025: 9 reportsAug 2025: 6 reportsSep 2025: 2 reportsOct 2025: 5 reportsNov 2025: 5 reportsDec 2025: 6 reportsJan 2026: 4 reportsFeb 2026: 4 reportsMar 2026: 7 reportsApr 2026: 1 reportsMay 2026: 3 reportsJun 2026: 4 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem304
  • Ambient Noise Problem188
  • High Capture Threshold97
  • Device Dislodged or Dislocated82
  • High impedance77
  • Fracture72
  • Failure to Capture65
  • Break38

Patient problems

  • No Clinical Signs, Symptoms or Conditions347
  • Unspecified Infection267
  • No Known Impact Or Consequence To Patient202
  • Pocket Erosion16
  • Bacteremia7
  • Endocarditis6
  • Sepsis6
  • Twiddlers Syndrome6

Recent adverse event reports

The 25 most recent of 873 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Sepsis

    MAUDE report 50811937 ↗
  • InjurySep 16, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50796755 ↗
  • InjurySep 11, 2026Fracture; Ambient Noise Problem

    This lead was explanted due to noise and suspected fracture. Should additional information be received this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50733656 ↗
  • InjurySep 2, 2026Adverse Event Without Identified Device or Use Problem

    This system was explanted due to infection. Should additional information become available, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50565834 ↗
  • MalfunctionAug 26, 2026Ambient Noise Problem

    HMSC reported episodes showing noise since generator replacement on (b)(6) 2026. Short interval counter >249 since then. All other lead trends appear unremarkable. Lead currently remains implanted. Should additional information be received this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50455681 ↗
  • MalfunctionAug 20, 2026Fracture; High impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50356172 ↗
  • InjuryAug 10, 2026Device Dislodged or Dislocated

    This lead was explanted due to dislodgement. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50199142 ↗
  • InjuryJun 24, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 25611892 ↗
  • InjuryJun 24, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 25611876 ↗
  • InjuryJun 22, 2026Break; Fracture; Low impedance; High Capture Threshold

    THIS LEAD WAS CAPPED DUE TO HIGH CAPTURE THRESHOLD, LOW IMPEDANCE, FRACTURE AND INSULATION BREAK. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25585606 ↗
  • InjuryJun 15, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 25512578 ↗
  • InjuryMay 27, 2026Fracture

    LEAD WAS CAPPED DUE TO FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25309550 ↗
  • MalfunctionMay 26, 2026Ambient Noise Problem; Unstable Capture Threshold

    DEVICE SHOWED NOISE ON RA LEAD DURING INTERROGATION ON (B)(6) 2026. NOISE WAS REPRODUCIBLE DURING ISOMETRICS. NO STABLE THRESHOLD. LEAD TO BE EVALUATED FURTHER. LEAD CURRENTLY REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25286033 ↗
  • InjuryMay 14, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO BACTEREMIA INFECTION WITH SEPSIS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Bacteremia; Sepsis

    MAUDE report 25177575 ↗
  • MalfunctionApr 28, 2026Ambient Noise Problem; Device Sensing Problem

    HMSC REPORTED EPISODE SHOWING NOISE RESULTING IN AT DETECTION ON (B)(6) 2026. ALL LEAD TRENDS APPEAR STABLE. LEAD CURRENTLY REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25023433 ↗
  • InjuryMar 30, 2026High Capture Threshold

    THIS LEAD WAS CAPPED DUE TO CHRONIC HIGH THRESHOLD. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24736881 ↗
  • InjuryMar 30, 2026Impedance Problem

    RA LEAD WAS CAPPED DUE TO OUT OF RANGE IMPEDANCE. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24736431 ↗
  • MalfunctionMar 25, 2026Ambient Noise Problem

    IT WAS REPORTED THAT WHAT APPEARS TO BE INTERMITTENT NOISE WAS SEEN ON THE ATRIAL CHANNEL. FURTHER EVALUATION WAS RECOMMENDED. LEAD REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24686224 ↗
  • InjuryMar 24, 2026Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24677132 ↗
  • InjuryMar 12, 2026Failure to Capture; Device Sensing Problem

    THIS RV LEAD WAS CAPPED DUE TO LOW SENSING AND NO CAPTURE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24584931 ↗
  • InjuryMar 5, 2026High Capture Threshold

    LEAD WAS CAPPED DUE TO HIGH THRESHOLD. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24526252 ↗
  • MalfunctionMar 5, 2026Ambient Noise Problem

    HMSC REPORTED EPISODES SHOWING NOISE. ADVISED FURTHER LEAD EVALUATION. LEAD CURRENTLY REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24515097 ↗
  • MalfunctionFeb 23, 2026Unstable Capture Threshold

    PATIENT HAS EXPERIENCED TWO SIGNIFICANT JUMPS IN RV THRESHOLD SINCE CHANGEOUT, BOTH RESULTING IN ER VISITS DUE TO DIZZINESS AS SHE IS DEPENDENT. NO UNDERLYING LEAD INTEGRITY ISSUE COULD BE IDENTIFIED FROM LEAD TREND DATA AS IMPEDANCE IS STABLE AND NO NOISE IS PRESENT IN RECORDINGS. PROVOCATIVE MOVEMENTS AND POCKET MANIPULATION REVEALED NO NOISE AND NO CHANGE IN IMPEDANCE MEASUREMENTS. LEAD REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: Dizziness

    MAUDE report 24413695 ↗
  • InjuryFeb 20, 2026Ambient Noise Problem

    RA LEAD WAS CAPPED DUE TO NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24400100 ↗
  • InjuryFeb 9, 2026Break

    THIS LEAD WAS CAPPED DUE TO IT BROKE WHILE PHYSICIAN WAS TRYING TO REMOVE IT FROM THE DEVICE HEADER DURING CHANGE-OUT. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24291902 ↗

Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices

All Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page