Skip to content

Setrox S 60 350975Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Quick summary

This Biotronik Setrox S 60 heart pacing lead has 344 adverse event reports on file, with no deaths and 107 injuries. Many reports describe electrical noise on the lead, weak or failed pacing, and breaks in the lead, and some mention infections. No recalls name this device. Reports in the last 12 months fell to 8 from 22 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
347
Deaths reported
0
Injuries reported
110
Recalls naming it
0

Summary

The FDA's MAUDE database contains 347 adverse event reports that list this exact device (DI 04035479087968), received between Jun 15, 2018 and Sep 17, 2026 — classified as 110 injuries, 237 malfunctions.

The most frequently reported specific device problem is “Ambient Noise Problem” (77 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 3 reportsMay 2024: 4 reportsJun 2024: 2 reportsJul 2024: 6 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 3 reportsJan 2025: 5 reportsFeb 2025: 3 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 3 reportsJul 2025: 2 reportsAug 2025: 3 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 5 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem95
  • Ambient Noise Problem77
  • High Capture Threshold66
  • High impedance46
  • Failure to Capture42
  • Fracture33
  • Device Dislodged or Dislocated21
  • Impedance Problem20

Patient problems

  • No Clinical Signs, Symptoms or Conditions148
  • No Known Impact Or Consequence To Patient86
  • Unspecified Infection79
  • Bradycardia4
  • Pocket Erosion4
  • Bacteremia3
  • Endocarditis3
  • Fall3

Recent adverse event reports

The 25 most recent of 347 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Sepsis

    MAUDE report 50811930 ↗
  • InjurySep 15, 2026Failure to Capture; High impedance

    RV lead was capped due to high impedance and no capture at max output. No adverse patient events were reported. Should additional information be received this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50779018 ↗
  • InjurySep 11, 2026Device Dislodged or Dislocated

    This lead was explanted due to it dislodged while the RA lead was being removed. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50734315 ↗
  • InjurySep 9, 2026High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50680730 ↗
  • InjurySep 2, 2026Low impedance; Ambient Noise Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50576305 ↗
  • MalfunctionAug 13, 2026Ambient Noise Problem

    RV lead failure was detected on (b)(6) 2026. Review of the bipolar RV lead failure episode IEGM demonstrated noise on the RV channel. The lead remains implanted at this time. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50234275 ↗
  • MalfunctionMay 5, 2026Ambient Noise Problem

    NOISE WAS SEEN ON THE RV CHANNEL IN RECORDINGS STORED ON THE DEVICE. LEAD IS CURRENTLY IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25087196 ↗
  • MalfunctionDec 31, 2025Impedance Problem; Impedance Problem

    THIS LEAD WAS CAPPED DUE TO IMPEDANCE OUT OF RANGE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23935963 ↗
  • MalfunctionDec 17, 2025Failure to Capture; Failure to Capture

    DURING GEN CHANGE, THIS LEAD WAS CAPPED DUE TO INTERMITTENT LOSS OF CAPTURE WHILE ON PSA. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23829758 ↗
  • MalfunctionDec 5, 2025Fracture; High impedance; Over-Sensing; Fracture; High impedance; Over-Sensing

    THIS RV PACE/SENSE LEAD WAS CAPPED DUE TO FRACTURE WITH PT SYNCOPE DUE TO OVERSENSING, AND IMPEDANCE OF NEARLY 3000 OHMS. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Syncope/Fainting

    MAUDE report 23728496 ↗
  • InjuryNov 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING LEAD REVISION AN INFECTION WAS FOUND AND SYSTEM WAS EXPLANTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 23509795 ↗
  • InjuryAug 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 22846922 ↗
  • InjuryAug 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 22794291 ↗
  • MalfunctionAug 4, 2025High Capture Threshold; High Capture Threshold

    THIS RV LEAD WAS CAPPED DUE TO HIGH PACING CAPTURE THRESHOLD WITH SIGNIFICANT RV PACING BURDEN. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22699336 ↗
  • MalfunctionJul 31, 2025High impedance; High Capture Threshold; High impedance; High Capture Threshold

    LEAD HAD HIGH IMPEDANCE AND THRESHOLD. IT WAS CAPPED AND PARTIALLY EXTRACTED (TIP REMAINS IN PATIENT). SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Foreign Body In Patient

    MAUDE report 22680544 ↗
  • MalfunctionJul 16, 2025Intermittent Capture; Intermittent Capture

    THIS RV LEAD WAS EXPLANTED DUE TO INTERMITTENT LOSS OF PACING CAPTURE. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22525015 ↗
  • InjuryJun 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO BACTEREMIA, ENDOCARDITIS, AND A SPINAL ABSCESS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Abscess; Bacteremia; Endocarditis

    MAUDE report 22350253 ↗
  • MalfunctionJun 11, 2025Pacing Intermittently; Ambient Noise Problem; Pacing Intermittently; Ambient Noise Problem

    RV LEAD WAS EXPLANTED DUE TO NOISE CAUSING INHIBITION OF PACING LEADING TO DIZZY SPELLS ON A DEPENDENT PATIENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: Dizziness

    MAUDE report 22190148 ↗
  • MalfunctionJun 4, 2025Failure to Capture; High impedance; Device Sensing Problem; Failure to Capture; High impedance; Device Sensing Problem

    THIS LEAD WAS CAPPED DUE TO >3,000 OHM IMPEDANCE, NO SENSING, AND NO CAPTURE AT MAXIMUM OUTPUT. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22145588 ↗
  • MalfunctionMay 9, 2025Failure to Capture; Failure to Capture

    LEAD WAS CAPPED DUE TO NON-CAPTURE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21996184 ↗
  • InjuryMar 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO ENDOCARDITIS. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Endocarditis

    MAUDE report 21510756 ↗
  • InjuryFeb 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 21354225 ↗
  • MalfunctionFeb 3, 2025Low impedance; Ambient Noise Problem; Low impedance; Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO NOISE AND LOW IMPEDANCE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21300446 ↗
  • MalfunctionFeb 3, 2025High Capture Threshold; High Capture Threshold

    THIS LEAD WAS CAPPED DUE TO HIGH THRESHOLD. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21296437 ↗
  • InjuryJan 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 21224642 ↗

Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices

All Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.