Siello T 60Recalls & reported problems
BIOTRONIK SE & Co. KG · Class 3 · Implant
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04035479106263), received between May 9, 2019 and Apr 26, 2021 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Failure to Capture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Failure to Capture2
- High Capture Threshold2
- High impedance2
Patient problems
- No Known Impact Or Consequence To Patient4
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 26, 2021High Capture Threshold; High Capture Threshold
AFTER AN IMPLANTATION PERIOD OF APPROX. 63 MONTHS, AN INCREASING THRESHOLD WAS OBSERVED. THE LEAD WAS CAPPED AND THE PACEMAKER WAS EXPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11729604 ↗MalfunctionFeb 6, 2020Failure to Capture; High impedance; Failure to Capture; High impedance
AFTER AN IMPLANTATION PERIOD OF APPROX. 57 MONTHS, IT WAS REPORTED THAT THERE WAS NO CAPTURE ON THE VENTRICULAR CHANNEL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9676452 ↗MalfunctionJan 15, 2020High Capture Threshold; High Capture Threshold
IT WAS REPORTED THE PATIENT HAD A VERY HIGH AND FLUCTUATING VENTRICULAR THRESHOLD UPON NEW LEAD INSERTION. THE LEAD WAS REPOSITIONED WITH NO SUCCESS AND FINALLY IT WAS EXPLANTED THREE DAYS AFTER THE IMPLANTATION. A NEW LEAD WITH ACTIVE FIXATION WAS IMPLANTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9593259 ↗MalfunctionMay 22, 2019Failure to Capture; Failure to Capture
APPROXIMATELY 7 DAYS AFTER IMPLANTATION, A LOSS OF CAPTURE ON THE VENTRICULAR CHANNEL WAS REPORTED. THIS LEAD WAS EXPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8633778 ↗MalfunctionMay 9, 2019High impedance; High impedance
OUS MDR - IT WAS REPORTED THAT APPROX. 72 MONTHS AFTER THE IMPLANT THE BIPOLAR PACING IMPEDANCE WAS OUT OF RANGE (GREATER THAN 2500 OHMS). THE DEVICE SETTINGS WERE CHANGED TO UNIPOLAR. TWO WEEKS LATER THE LEAD AND THE ENTIRE PACEMAKER SYSTEM WAS EXPLANTED AND REPLACED. THE LEAD WAS DESTROYED DURING THE EXTRACTION PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8597526 ↗
Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes438 deaths19,632 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes200 deaths12,708 reports
- INGEVITY™ MRI 7741Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes36 deaths7,874 reports
- INGEVITY™+ 7841Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes15 deaths7,419 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes190 deaths6,776 reports
- INGEVITY™+ 7842Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes21 deaths5,995 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes50 deaths5,027 reports
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes40 deaths5,000 reports