Safio S 53 370945Recalls & reported problems
BIOTRONIK SE & Co. KG · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Biotronik Safio S pacemaker lead is an implanted wire that connects a pacemaker to the heart. The 23 reports mostly describe electrical sensing problems and rising impedance months or years after implant, with leads capped or replaced. There are no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 23
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 04035479113940), received between Apr 9, 2019 and Mar 9, 2022 — classified as 23 malfunctions.
The most frequently reported specific device problem is “Over-Sensing” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Over-Sensing7
- Failure to Capture5
- High impedance4
- Break3
- Device Dislodged or Dislocated3
- Impedance Problem3
- Fracture2
- High Capture Threshold2
Patient problems
- No Known Impact Or Consequence To Patient14
- No Clinical Signs, Symptoms or Conditions7
- Bradycardia1
- Undesired Nerve Stimulation1
Recent adverse event reports
The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 9, 2022Over-Sensing; Impedance Problem; Over-Sensing; Impedance Problem
AFTER AN IMPLANTATION PERIOD OF APPROXIMATELY 106 MONTHS IT WAS REPORTED THAT THE LEAD WAS CAPPED DUE TO OUT OF RANGE IMPEDANCE MEASUREMENTS AND OVERSENSING. A REVIEW OF DATA SHOWED THAT THIS FILE WAS INADVERTENTLY NOT ENTERED INTO THE SYSTEM AND IS THEREFORE RECORDED NOW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13704387 ↗MalfunctionOct 8, 2021Over-Sensing; Device Sensing Problem; Impedance Problem; Over-Sensing; Device Sensing Problem; Impedance Problem
AFTER AN IMPLANTATION PERIOD OF APPROXIMATELY 88 MONTHS, IT WAS REPORTED THAT OVERSENSING IN THE ATRIAL CHANNEL, SUDDENLY INCREASING PACING IMPEDANCE AND LOSS OF SENSING WERE DETECTED. THE LEAD IS NOW INACTIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12600707 ↗MalfunctionSep 8, 2021Over-Sensing; Over-Sensing
AFTER AN IMPLANTATION PERIOD OF APPROX. 54 MONTHS, OVERSENSING ON THE VENTRICULAR AND FARFIELD CHANNEL WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12440350 ↗MalfunctionMay 24, 2021High impedance; High impedance
AFTER AN IMPLANTATION PERIOD OF APPROX. 31 MONTHS, IT WAS REPORTED THAT AN INCREASING PACING IMPEDANCE WAS OBSERVED. THE LEAD WAS CAPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11869759 ↗MalfunctionMar 9, 2021Extracardiac stimulation; Ambient Noise Problem; Extracardiac stimulation; Ambient Noise Problem
THIS LEAD WAS CAPPED AND REPLACED DUE TO BIPOLAR LEAD FAILURE, NOISE IN BIPOLAR PROGRAMMING, AND PECTORAL STIM IN UNIPOLAR PACING CONFIGURATION. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: Undesired Nerve Stimulation
MAUDE report 11449328 ↗MalfunctionJan 18, 2021Failure to Capture; Failure to Capture
IT WAS REPORTED THAT APPROX. 18 MONTHS AFTER THE IMPLANTATION LOSS OF CAPTURE WAS NOTED. THE PACEMAKER WAS EXPLANTED AND REPLACED, THE LEAD WAS CAPPED AND A NEW ONE WAS IMPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11189481 ↗MalfunctionDec 7, 2020High impedance; High impedance
IT WAS REPORTED, THAT AFTER APPROXIMATELY 67 MONTHS IMPEDANCE INCREASE OF THE RA-LEAD WAS DETECTED. A NEW RA-LEAD WAS IMPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10962559 ↗MalfunctionOct 13, 2020Impedance Problem; Impedance Problem
IT WAS REPORTED THAT APPROX. 48 MONTHS AFTER THE IMPLANTATION, DURING A FOLLOW-UP ERROR MESSAGES WERE SEEN FOR ATRIAL (SAFIO S 53) AND VENTRICULAR (SAFIO S 60) LEADS DUE TO AN IMPEDANCE ISSUE. IT WAS DECIDED TO EXPLANT THE LEADS AND PERFORM A NEW CRT-P IMPLANT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10671624 ↗MalfunctionSep 1, 2020Failure to Capture; Failure to Capture
AFTER AN IMPLANTATION PERIOD OF APPROX. 26 MONTHS, A LOSS OF CAPTURE ON THE ATRIAL CHANNEL WAS OBSERVED. THE LEADS WERE EXPLANTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10475729 ↗MalfunctionJul 22, 2020Break; Break
IT WAS REPORTED THAT APPROX. 61 MONTHS AFTER IMPLANTATION, THE RV (LINOX SMART) AND RA (SAFIO) LEADS WERE EXPLANTED. INSULATION DAMAGE WAS OBSERVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10311861 ↗MalfunctionJun 18, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON (B)(6) 2020, A PATIENT WAS IMPLANTED WITH A NEW PACEMAKER. HOWEVER, DURING FOLLOW-UP BEFORE DISCHARGE ON (B)(6), IT WAS NOTED THAT THE RA AND RV LEADS WERE DISLODGED. ON MAY 28, THE DOCTOR PERFORMED LEAD RE-POSITIONING SURGERY FOR THE PATIENT, BUT THE LEADS DISLODGED AGAIN. THEN THE TWO LEADS WERE UNABLE TO BE FIXED. AT LAST, THE DOCTOR IMPLANTED OTHER 2 NEW LEADS FOR THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10171740 ↗MalfunctionJun 18, 2020Break; Fracture; Naturally Worn; Break; Fracture; Naturally Worn
IT WAS REPORTED THAT THE ATRIAL LEAD SHOWED INSULATION DAMAGE AND COIL FRACTURE DUE TO SUBCLAVIAN CRUSH SYNDROME APPROX. 29 MONTHS AFTER IMPLANTATION. THE LEAD WAS EXPLANTED AND REPLACED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10171407 ↗MalfunctionMay 28, 2020Intermittent Capture; Intermittent Capture
IT WAS REPORTED THAT APPROX. 72 MONTHS AFTER THE IMPLANTATION INTERMITTENT LOSS OF CAPTURE (ATRIAL) WAS NOTED. PATIENT RECEIVED A COMPLETE NEW SYSTEM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10099332 ↗MalfunctionApr 8, 2020Over-Sensing; Over-Sensing
IT WAS REPORTED THAT BOTH LEADS WERE EXPLANTED APPROX. 54 MONTHS AFTER IMPLANTATION. THE ATRIAL LEAD SHOWED OVERSENSING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9941668 ↗MalfunctionMar 11, 2020Over-Sensing; Over-Sensing
AFTER AN IMPLANTATION PERIOD OF APPROX. 67 MONTHS, OVERSENSING ON THE ATRIAL AS WELL AS ON THE VENTRICULAR CHANNEL WAS REPORTED. THE PATIENT WAS HOSPITALIZED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9820932 ↗MalfunctionJan 8, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
THIS RA LEAD WAS EXPLANTED DUE TO DISLODGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9566813 ↗MalfunctionOct 28, 2019Over-Sensing; Over-Sensing
AFTER AN IMPLANTATION PERIOD OF APPROXIMATELY 43 MONTHS, OVERSENSING ON THE ATRIAL CHANNEL WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9248022 ↗MalfunctionOct 24, 2019Failure to Capture; Over-Sensing; Failure to Capture; Over-Sensing
IT WAS REPORTED THAT 59 MONTHS AFTER THE IMPLANT OF THIS SYSTEM OVERSENSING ON VENTRICULAR AND ATRIAL CHANNELS AND LOSS OF CAPTURE WERE NOTED. THE PACEMAKER AND THE LEADS REMAIN IMPLANTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9235425 ↗MalfunctionOct 10, 2019Failure to Capture; High impedance; High Capture Threshold; Failure to Capture; High impedance; High Capture Threshold
IT WAS REPORTED THAT APPROX. 58 MONTHS AFTER THE IMPLANTATION THE LEAD WAS CAPPED DUE TO HIGH PACING IMPEDANCES, HIGH THRESHOLDS AND LOSS OF CAPTURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9175405 ↗MalfunctionOct 1, 2019Break; Break
IT WAS REPORTED THAT THE LEAD SHOWED AN INSULATION DAMAGE DURING THE EXTRACTION OF THE ASSOCIATED ICD LEAD.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9141719 ↗MalfunctionAug 12, 2019Fracture; High Capture Threshold; Fracture; High Capture Threshold
IT WAS REPORTED, THAT APPROXIMATELY 16 MONTHS AFTER IMPLANTATION, THE THRESHOLD VALUE OF THIS LEAD WAS INCREASED. AFTER FURTHER CHECKING BY X-RAY, IT WAS NOTED THAT THE LEAD WAS FRACTURED. THE LEAD WAS EXPLANTED AND REPLACED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8885996 ↗MalfunctionJul 18, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
THIS LEAD WAS EXPLANTED AND REPLACED DUE TO DISLODGEMENT. THE LEAD WAS IMPLANTED WHILE THE PATIENT WAS ON VACATION OUTSIDE OF THE US.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8803061 ↗MalfunctionApr 9, 2019Failure to Capture; High impedance; Failure to Capture; High impedance
OUS MDR - THIS RA LEAD HAD NO CAPTURE AT MAX OUTPUT AND IMPEDANCES OF GREATER THAN 2500 OHMS. THIS LEAD WAS EXPLANTED ON (B)(6) 2019.
Patient problems: Bradycardia
MAUDE report 8496928 ↗
Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes438 deaths19,632 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes200 deaths12,708 reports
- INGEVITY™ MRI 7741Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes36 deaths7,874 reports
- INGEVITY™+ 7841Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes15 deaths7,419 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes190 deaths6,776 reports
- INGEVITY™+ 7842Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes21 deaths5,995 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes50 deaths5,027 reports
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes40 deaths5,000 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page