Skip to content

Entovis DR-T ProMRI 371992Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BIOTRONIK Entovis DR-T ProMRI is an implanted pacemaker that keeps the heart beating steadily and is built to work in an MRI machine. The 50 reports mostly describe infection around the device that led to removing the pacemaker, plus battery running down sooner than expected and the device failing to pace the heart;33 reports involved injuries and no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
52
Deaths reported
0
Injuries reported
35
Recalls naming it
0

Summary

The FDA's MAUDE database contains 52 adverse event reports that list this exact device (DI 04035479115173), received between Jul 27, 2018 and Sep 17, 2026 — classified as 35 injuries, 17 malfunctions.

The most frequently reported specific device problem is “Battery Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem34
  • Battery Problem6
  • Premature Elective Replacement Indicator5
  • Failure to Capture2
  • Failure to Interrogate1
  • Impedance Problem1
  • Over-Sensing1
  • Overheating of Device1

Patient problems

  • Unspecified Infection27
  • No Clinical Signs, Symptoms or Conditions12
  • No Known Impact Or Consequence To Patient4
  • Abscess1
  • Asystole1
  • Bacteremia1
  • Convulsion/Seizure1
  • Discomfort1

Recent adverse event reports

The 25 most recent of 52 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Sepsis

    MAUDE report 50811934 ↗
  • InjurySep 2, 2026Adverse Event Without Identified Device or Use Problem

    This system was explanted due to infection. Should additional information become available, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50565928 ↗
  • MalfunctionJun 16, 2026Premature Elective Replacement Indicator

    DEVICE IS AT ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. DEVICE CURRENTLY REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25522009 ↗
  • InjuryJun 1, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 25348922 ↗
  • InjuryDec 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. THE EXPLANT DATE IS UNKNOWN. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 23849433 ↗
  • InjuryNov 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING LEAD REVISION AN INFECTION WAS FOUND AND SYSTEM WAS EXPLANTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 23509796 ↗
  • InjuryAug 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 22794293 ↗
  • InjuryJun 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO BACTEREMIA, ENDOCARDITIS, AND A SPINAL ABSCESS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Abscess; Bacteremia; Endocarditis

    MAUDE report 22350262 ↗
  • MalfunctionApr 29, 2025Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE WAS EXPLANTED DUE TO ERI WITH UNUSUAL BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21927967 ↗
  • InjuryFeb 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 21351877 ↗
  • InjuryDec 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO MRSA INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Drug Resistant Bacterial Infection

    MAUDE report 20854693 ↗
  • MalfunctionDec 5, 2024Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE IS AT ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20853870 ↗
  • InjuryJun 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 19593931 ↗
  • InjuryAug 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 17445742 ↗
  • InjuryMar 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 16533464 ↗
  • InjuryNov 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE EXPLANTED DUE TO INFECTION. EXPLANT DATE IS UNKNOWN. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 15875146 ↗
  • InjuryNov 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT HAD A SYNCOPAL EPISODE OCCUR WHILE RV THRESHOLD TESTING WAS PERFORMED DURING A ROUTINE FOLLOW-UP INTERROGATION. RV THRESHOLD TEST WAS SET TO A STARTING AMPLITUDE OF 4.0V AND A PULSE WIDTH OF 0.4MS. PATIENTS PERMANENTLY PROGRAMMED SETTINGS FOR RV WERE AN AMPLITUDE OF 3.6V AND PULSE WIDTH OF 0.4MS. THE RV THRESHOLD TEST WAS SET TO DECREMENT BY ONE PULSE AND THEN RETURN TO STARTING AMPLITUDE OF 4.0V AT PULSE WIDTH OF 0.4MS. DURING THE RV THRESHOLD TEST THE RENAMIC PROGRAMMER FROZE WHILE AT AN OUTPUT LESS THAN CAPTURE THRESHOLD AND PATIENT HAD A SUDDEN LOSS OF CONSCIOUSNESS AND BEGAN TO SEIZE. THE SCREEN ON THE RENAMIC PROGRAMMER WAS UNRESPONSIVE TO INPUT UNTIL THE PHYSICIAN REMOVED THE PROGRAMMING HEAD AND BROKE TELEMETRY WITH THE DEVICE. THE PROGRAMMER WAS SET TO PRINT RESULTS DURING TESTING AND WAS ALSO CONNECTED TO AN EXTERNAL PRINTER. AFTER REBOOTING THE PROGRAMMER THE DEVICE WAS RE-INTERROGATED AND ALL PERMANENTLY PROGRAMMED SETTINGS AND PARAMETERS WERE NORMAL. DEVICE REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Convulsion/Seizure; Syncope/Fainting

    MAUDE report 15798447 ↗
  • InjuryAug 16, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 15233812 ↗
  • InjuryAug 3, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 15161208 ↗
  • MalfunctionJul 28, 2022Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE SHOWS ERI WITH UNEXPECTED BATTERY BEHAVIOR. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15126726 ↗
  • InjuryMay 5, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 14286054 ↗
  • MalfunctionDec 14, 2021Premature End-of-Life Indicator; Premature End-of-Life Indicator

    THIS DEVICE WAS EXPLANTED DUE TO EOS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13002063 ↗
  • MalfunctionNov 15, 2021Adverse Event Without Identified Device or Use Problem; Unstable Capture Threshold; Adverse Event Without Identified Device or Use Problem; Unstable Capture Threshold

    PATIENT WAS AT A FOLLOW UP AND WHEN THE DEVICE WAS INTERROGATED THE PATIENT WENT ASYSTOLIC. WHEN THE WAND WAS REMOVED THE PATIENT WAS OKAY. THE PHYSICIAN DID THE SAME THING WITH THE SAME RESULTS. THE DEVICE THRESHOLD REPORT WAS 2.5 VOLTS AND THE DEVICE WAS SET TO 1.3 VOLTS. THE DEVICE REMAINS IMPLANTED AT THIS TIME.

    Patient problems: Asystole

    MAUDE report 12817588 ↗
  • InjuryOct 7, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. NEW SYSTEM WAS NOT IMPLANTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 12594511 ↗
  • InjuryAug 13, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: Unspecified Infection

    MAUDE report 12319273 ↗

Other Pulse generator, permanent, implantable devices

All Pulse generator, permanent, implantable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page