Solia S 45 377176Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This is the Biotronik Solia S 45 pacemaker lead, a drug-eluting electrode placed in the heart's right chambers. The 1,695 reports mostly describe the lead moving out of place, electrical noise in the signal, and failure to pace the heart reliably, with infections also mentioned. Several describe the lead being removed. The records include 6 deaths, 798 injuries and 891 malfunctions, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,702
Deaths reported
6
Injuries reported
805
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,702 adverse event reports that list this exact device (DI 04035479118259), received between Oct 9, 2017 and Sep 29, 2026 — classified as 6 deaths, 805 injuries, 891 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (706 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 14 reportsDec 2023: 20 reportsJan 2024: 20 reportsFeb 2024: 3 reportsMar 2024: 16 reportsApr 2024: 20 reportsMay 2024: 14 reportsJun 2024: 11 reportsJul 2024: 11 reportsAug 2024: 8 reportsSep 2024: 26 reportsOct 2024: 16 reportsNov 2024: 10 reportsDec 2024: 6 reportsJan 2025: 37 reportsFeb 2025: 9 reportsMar 2025: 15 reportsApr 2025: 7 reportsMay 2025: 19 reportsJun 2025: 14 reportsJul 2025: 22 reportsAug 2025: 28 reportsSep 2025: 14 reportsOct 2025: 13 reportsNov 2025: 18 reportsDec 2025: 17 reportsJan 2026: 15 reportsFeb 2026: 19 reportsMar 2026: 19 reportsApr 2026: 25 reportsMay 2026: 12 reportsJun 2026: 7 reportsJul 2026: 0 reportsAug 2026: 11 reportsSep 2026: 13 reportsOct 2026: 0 reports20242025202637
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem713
  • Device Dislodged or Dislocated706
  • Ambient Noise Problem93
  • Failure to Capture86
  • High Capture Threshold74
  • High impedance52
  • Device Sensing Problem35
  • Fracture29

Patient problems

  • No Clinical Signs, Symptoms or Conditions669
  • Unspecified Infection585
  • No Known Impact Or Consequence To Patient245
  • Pocket Erosion67
  • Twiddlers Syndrome46
  • Cardiac Perforation14
  • Endocarditis13
  • Implant Pain12

Recent adverse event reports

The 25 most recent of 1,702 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2026Patient Device Interaction Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940891 ↗
  • InjurySep 28, 2026Device Dislodged or Dislocated

    This lead was repositioned due to dislodgement. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916369 ↗
  • InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem

    System explanted due to pocket erosion. Should additional information be received, this file will be updated.

    Patient problems: Pocket Erosion

    MAUDE report 50899157 ↗
  • InjurySep 23, 2026Device Dislodged or Dislocated

    This lead was explanted due to dislodgement. Should additional information become available, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50870617 ↗
  • InjurySep 16, 2026Device Dislodged or Dislocated

    This lead was explanted due to dislodgement. Should additional information become available, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50794168 ↗
  • InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to infection. Should additional information be received, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50788205 ↗
  • InjurySep 15, 2026Device Dislodged or Dislocated

    Lead dislodged shortly after implant procedure was completed. After dislodgement was discovered, patient developed pericardial effusion. MD opted to explant the lead and replace with a different type of lead. Should additional information be received this file will be updated.

    Patient problems: Pericardial Effusion

    MAUDE report 50773265 ↗
  • InjurySep 15, 2026Failure to Capture; Device Sensing Problem; Impedance Problem

    This lead was explanted due to loss of capture, sensing and impedance issues. No adverse patient events were reported. Should additional information become available, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50772447 ↗
  • InjurySep 11, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to infection. Should additional information be received, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50735010 ↗
  • InjurySep 11, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to infection. Should additional information be received, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50734847 ↗
  • InjurySep 9, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to infection. Should additional information become available, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50682759 ↗
  • InjurySep 8, 2026Fracture; Ambient Noise Problem

    This lead was capped due to noise and suspected insulation damage. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50670145 ↗
  • InjurySep 3, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to Vegetation noted in Echocardiogram of RV lead. Should additional information be received, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50590285 ↗
  • InjuryAug 19, 2026Device Dislodged or Dislocated

    This lead was explanted due to dislodgement as a result of patient non-compliance. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50331023 ↗
  • InjuryAug 19, 2026Device Dislodged or Dislocated

    This RA lead was explanted due to dislodgement. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50330285 ↗
  • InjuryAug 17, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to infection. Should additional information be received, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50278577 ↗
  • InjuryAug 13, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to infected endocarditis. Should additional information be received, this file will be updated.

    Patient problems: Endocarditis; Unspecified Infection

    MAUDE report 50233906 ↗
  • InjuryAug 12, 2026Device Dislodged or Dislocated

    This lead was explanted due to dislodgement. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50230293 ↗
  • MalfunctionAug 12, 2026Ambient Noise Problem; Device Sensing Problem

    Mode switch recording transmitted to Home Monitoring shows atrial oversensing of what appears to be noise. Other recent IEGMs show a clean atrial channel. Postural testing and isometrics were recommended. Lead remains implanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50230089 ↗
  • InjuryAug 12, 2026Fracture

    Lead explanted due to fracture. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50217177 ↗
  • InjuryAug 10, 2026Device Dislodged or Dislocated

    This lead was explanted due to dislodgement as a result of twiddling. Should additional information be received, this file will be updated.

    Patient problems: Twiddlers Syndrome

    MAUDE report 50198772 ↗
  • InjuryAug 7, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to infection. Should additional information be received, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50179021 ↗
  • InjuryAug 7, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to infection. Should additional information be received, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50178996 ↗
  • InjuryAug 4, 2026Failure to Capture; Device Sensing Problem

    This lead was explanted due to loss of sensing and no capture. No adverse patient events were reported. Should additional information become available, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50134830 ↗
  • MalfunctionJun 25, 2026Ambient Noise Problem

    NOISE IS PRESENT ON RA AND RV LEAD. ALL OTHER VALUES APPEAR WITHIN NORMAL LIMITS. PATIENT IS NONCOMPLIANT WITH FOLLOW-UPS. RULED OUT EXTERNAL PROCEDURE DURING TIME OF RECORDING. LEAD REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25630548 ↗

Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices

All Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices →