Skip to content

Solia T 53Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Quick summary

This Biotronik Solia T 53 is a drug-eluting wire (lead) implanted in the heart to keep a pacemaker working in the right ventricle or right atrium. The 50 reports mostly describe leads that shift out of position or break, along with higher-than-expected pacing thresholds that led to the lead being capped or removed. Twenty-three reports describe injuries and none describe deaths; infection at the implant site and pain are also mentioned. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
54
Deaths reported
0
Injuries reported
27
Recalls naming it
0

Summary

The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 04035479118280), received between May 16, 2019 and Sep 29, 2026 — classified as 27 injuries, 27 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem21
  • Device Dislodged or Dislocated19
  • High Capture Threshold8
  • Failure to Capture5
  • Fracture3
  • High impedance3
  • Migration or Expulsion of Device2
  • Entrapment of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions27
  • Unspecified Infection18
  • No Known Impact Or Consequence To Patient5
  • Pain2
  • Fall1
  • Pocket Erosion1
  • Wound Dehiscence1

Recent adverse event reports

The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2026Failure to Capture; Device Dislodged or Dislocated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926839 ↗
  • InjurySep 16, 2026Device Dislodged or Dislocated

    This lead was explanted due to dislodgement. Should additional information become available, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50794175 ↗
  • InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to infection. Should additional information be received, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50788817 ↗
  • InjurySep 15, 2026Entrapment of Device

    Lead was attempted but it was unable to be removed due to possible entanglement on sub valvular structure so it was capped and another lead placed successfully. Lead remains inactive in the body.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50778705 ↗
  • InjuryMay 13, 2026High Capture Threshold

    LEAD WAS CAPPED DUE TO HIGH RV LEAD THRESHOLD. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25162363 ↗
  • InjuryMay 5, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 25085677 ↗
  • InjuryFeb 17, 2026Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24372667 ↗
  • InjuryFeb 16, 2026Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24344089 ↗
  • MalfunctionDec 22, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23871812 ↗
  • InjuryDec 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE WAS EXPLANTED AND LEADS WERE CAPPED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 23694458 ↗
  • MalfunctionOct 15, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23286141 ↗
  • InjurySep 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO POCKET EROSION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Pocket Erosion

    MAUDE report 23063048 ↗
  • MalfunctionSep 5, 2025Failure to Capture; Migration or Expulsion of Device; Device Dislodged or Dislocated; Failure to Capture; Migration or Expulsion of Device; Device Dislodged or Dislocated

    LEAD WAS CAPPED BECAUSE IT DISLODGED AND MIGRATED RESULTING IN NO CAPTURE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22982505 ↗
  • InjuryAug 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 22790212 ↗
  • MalfunctionMay 8, 2025Failure to Capture; Failure to Capture

    PATIENT SYMPTOMATIC AND INTERROGATION SHOWED LOSS OF CAPTURE. PATIENT IN ER FOR MONITORING. DEVICE OUTPUTS PROGRAMMED TO HIGHER OUTPUT AND SAFETY MARGIN. IMPEDANCE MEASUREMENTS VARIED FROM 600-1400OHMS SINCE IMPLANT. X-RAY APPEARED NORMAL. DEVICE WILL BE MONITORED IN HOSPITAL AND REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21986696 ↗
  • InjuryApr 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 21754643 ↗
  • MalfunctionFeb 12, 2025Fracture; Fracture

    THIS LEAD WAS EXPLANTED DUE TO FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21368609 ↗
  • MalfunctionOct 29, 2024Migration or Expulsion of Device; Device Dislodged or Dislocated; Migration or Expulsion of Device; Device Dislodged or Dislocated

    LEAD WAS EXPLANTED DUE TO DISLODGEMENT AND MIGRATION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20557451 ↗
  • MalfunctionAug 8, 2024Fracture; High impedance; Fracture; High impedance

    POSSIBLE LEAD FRACTURE. IMPEDANCE OVER 2000. THE LEAD WAS EXPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19945338 ↗
  • InjuryMay 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED. EXPLANT DATE UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 19337522 ↗
  • MalfunctionApr 18, 2024High Capture Threshold; High Capture Threshold

    PATIENT PRESENTED TO ER AFTER FALL AND ON TELEMETRY IT WAS NOTED HIS HEART RATE WAS IN THE 30S. INTERROGATION REVEALED HIGH THRESHOLD IN RV LEAD IN BIPOLAR AND UNIPOLAR. PATIENT UNDERWENT RV REPOSITION. LEAD REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Fall

    MAUDE report 19132928 ↗
  • MalfunctionApr 3, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS CAPPED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19032904 ↗
  • InjuryMar 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO POCKET INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 18975993 ↗
  • MalfunctionJan 17, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS CAPPED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18528373 ↗
  • MalfunctionNov 21, 2023High Capture Threshold; High Capture Threshold

    THIS RV LEAD WAS CAPPED DUE TO HIGH THRESHOLDS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18183464 ↗

Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices

All Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.