Solia T 60Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Solia T 60 is a drug-eluting pacemaker lead (electrode) placed in the heart's right ventricle or atrium. The 75 reports mostly describe leads that move out of position, fail to capture or sense the heartbeat, or show high electrical resistance; several mention infection that led to removing the system. There is 1 death and 27 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
74
Deaths reported
1
Injuries reported
26
Recalls naming it
0

Summary

The FDA's MAUDE database contains 74 adverse event reports that list this exact device (DI 04035479118297), received between May 8, 2019 and Jun 15, 2026 — classified as 1 death, 26 injuries, 47 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem26
  • Device Dislodged or Dislocated19
  • High Capture Threshold12
  • High impedance9
  • Failure to Capture8
  • Break4
  • Low impedance3
  • Ambient Noise Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions33
  • Unspecified Infection22
  • No Known Impact Or Consequence To Patient12
  • Perforation2
  • Asystole1
  • Cardiac Perforation1
  • Chest Pain1
  • Death1

Recent adverse event reports

The 25 most recent of 74 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 15, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 25512525 ↗
  • InjuryMar 23, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 24661546 ↗
  • MalfunctionNov 19, 2025Low impedance; Device Sensing Problem; Low impedance; Device Sensing Problem

    THE LEAD WAS SUCCESSFULLY IMPLANTED, AND INITIAL BIPOLAR MEASUREMENTS WERE WITHIN ACCEPTABLE PARAMETERS. AFTER THE DEVICE WAS SUTURED IN PLACE AND THE PATIENT WAS CLOSED, SUBSEQUENT TESTING REVEALED THE ABSENCE OF BIPOLAR MEASUREMENTS, NO SENSING, AND IMPEDANCE READINGS BELOW 100 OHMS. UNIPOLAR MEASUREMENTS REMAINED ACCEPTABLE; HOWEVER, THE IMPLANTING PHYSICIAN DETERMINED THAT A NEW LEAD WAS REQUIRED. THE LEAD WAS EXPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23600731 ↗
  • InjuryOct 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 23402144 ↗
  • InjurySep 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 23063085 ↗
  • MalfunctionJul 8, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22443781 ↗
  • MalfunctionJun 5, 2025High impedance; High impedance

    LEAD WAS EXPLANTED DUE TO HIGH IMPEDANCE NOTED GREATER THAN 2K OHMS. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22154691 ↗
  • InjuryDec 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 20853950 ↗
  • InjuryOct 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT HAD A HEMATOMA AND INFECTION. SYSTEM WAS EXPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Hematoma; Unspecified Infection

    MAUDE report 20557559 ↗
  • InjuryAug 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 20089910 ↗
  • MalfunctionApr 25, 2024High Capture Threshold; High Capture Threshold

    THIS LEAD WAS SUCCESSFULLY REPOSITIONED THE DAY AFTER IMPLANT DUE TO THRESHOLD INCREASED TO 5.8 VOLTS AT 0.4 MS. POSSIBLY DUE TO PATIENT ANATOMY POST MITRAL VALVE REPAIR. LEAD REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19182985 ↗
  • MalfunctionApr 24, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19177201 ↗
  • InjuryApr 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 19112408 ↗
  • InjuryMar 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 18975990 ↗
  • MalfunctionFeb 27, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18793817 ↗
  • MalfunctionDec 28, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS RV LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18410248 ↗
  • InjuryDec 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 18314559 ↗
  • MalfunctionSep 12, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17730691 ↗
  • MalfunctionJul 11, 2023Break; Ambient Noise Problem; High Capture Threshold; Break; Ambient Noise Problem; High Capture Threshold

    THIS RV LEAD WAS EXPLANTED DUE TO NOISE AND THRESHOLDS GREATER THAN 2500 OHMS LIKELY CAUSED BY SUBCLAVIAN CRUSH. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17300504 ↗
  • MalfunctionJan 11, 2023Break; Failure to Capture; Impedance Problem; Break; Failure to Capture; Impedance Problem

    LEAD WAS CAPPED DUE TO POSSIBLE INSULATION DAMAGE, DROP IN IMPEDANCE VALUES AND POSSIBLE LOSS OF CAPTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16141407 ↗
  • InjuryOct 19, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 15631471 ↗
  • InjuryOct 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 15620954 ↗
  • MalfunctionOct 3, 2022Fracture; High impedance; Fracture; High impedance

    THIS LEAD WAS CAPPED DUE TO SUSPICION OF FRACTURE. BOTH BIPOLAR AND UNIPOLAR IMPEDANCES HIGHER THAN 2900-3000. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15533114 ↗
  • MalfunctionAug 15, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    LEAD EXPLANTED DUE TO MICRODISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15223948 ↗
  • MalfunctionJul 29, 2022High impedance; High Capture Threshold; High impedance; High Capture Threshold

    IT WAS REPORTED THAT PACING IMPEDANCE WAS TOO HIGH AND OUT OF RANGE AND THE PACING THRESHOLDS INCREASED. THE LEAD WAS EXPLANTED AND REPLACED APPROX. 38 MONTHS AFTER THE IMPLANTATION. PLEASE NOTE THAT THE INCIDENT WAS ALREADY REPORTED BY THE HEALTH FACILITY. THE SERIAL NUMBER IN THAT REPORT WAS MISTAKEN. SOLIA T 60, SN 8000428942 WAS IMPLANTED DURING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15127346 ↗

Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices

All Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices →