Skip to content

Resilience 75TRRecalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

All models of Resilience (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04035479170349), received between May 29, 2024 and May 12, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Break1
  • Fracture1

Patient problems

  • Foreign Body In Patient2
  • Low Blood Pressure/ Hypotension2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 12, 2026Adverse Event Without Identified Device or Use Problem

    LEAD WAS ATTEMPTED BUT NOT IMPLANTED. PROCEDURE WAS ABORTED. PATIENTS BLOOD PRESSURE FELL BELOW 90/40. MD TOOK NEEDLES AND LEADS OUT AND FLIPPED THE PATIENT OVER. EMERGENT SITUATION WHICH WAS RESOLVED AFTER FLIPPING THE PATIENT OVER.

    Patient problems: Low Blood Pressure/ Hypotension

    MAUDE report 25148744 ↗
  • InjuryMay 12, 2026Adverse Event Without Identified Device or Use Problem

    LEAD WAS ATTEMPTED BUT NOT IMPLANTED. PROCEDURE WAS ABORTED. PATIENTS BLOOD PRESSURE FELL BELOW 90/40. MD TOOK NEEDLES AND LEADS OUT AND FLIPPED THE PATIENT OVER. EMERGENT SITUATION WHICH WAS RESOLVED AFTER FLIPPING THE PATIENT OVER.

    Patient problems: Low Blood Pressure/ Hypotension

    MAUDE report 25148743 ↗
  • MalfunctionApr 23, 2026Break

    LEAD ATTEMPTED BUT NOT IMPLANTED. MD WAS ATTEMPTING TO PLACE LEAD AT LEVEL T5-6 WHERE NEEDLE WAS PLACED WHEN THE LEAD SEEMED TO HAVE GOTTEN IMPACTED BY THE LIGAMENT THEN THE PHYSICIAN ATTEMPTED TO REMOVE THE NEEDLE AND THE LEAD. ONE CONTACT OF THE LEAD REMAINED IN THE PATIENT EPIDURAL SPACE IN THE T5 AREA WHILE THE REMAINING LEAD AND NEEDLE WERE REMOVED. NO COMPLICATION OCCURRED DUE TO THIS INCIDENT. THERAPY WAS ACTIVATED WITH GOOD PAIN RELIEF AS TRIAL CONTINUES. PATIENT WILL BE SENT TO NEUROSURGERY FOR REMOVAL OF THE BROKEN CONTACT WHEN THE PERMANENT SYSTEM IS IMPLANTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 24967127 ↗
  • MalfunctionMay 29, 2024Fracture; Fracture

    TRAUMA WAS CAUSED TO THE LEAD WHEN THE PHYSICIAN WAS TRYING TO ADVANCE THE NEEDLE INTO THE EPIDURAL SPACE WHILE THE LEAD WAS PARTIALLY ADVANCED BEYOND THE NEEDLE TIP. ELECTRODE 1 AND 2 BROKE OFF THE LEAD WHILE THE PHYSICIAN WAS REMOVING THE LEAD AND NEEDLE. THE TWO ELECTRODES ARE OUTSIDE THE EPIDURAL SPACE AND THE PHYSICIAN PLANS ON REMOVING THEM WHEN THE PATIENT HAS THE PERMANENT IMPLANT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Foreign Body In Patient

    MAUDE report 19416741 ↗

Other Stimulator, spinal-cord, totally implanted for pain relief devices

All Stimulator, spinal-cord, totally implanted for pain relief devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page