ROTAFLOW Console, English/ USRecalls & reported problems
Maquet Cardiopulmonary AG · ROTAFLOW English/US US-Plug · Class 2
All models of ROTAFLOW (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04037691562445), received between Apr 22, 2025 and Feb 27, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (2 reports).
1 FDA recall record name this device's identifier.
1 other recall cover devices cleared under the same FDA submission (K991864); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Aug 7, 2020Terminated
Instruction For Use (IFU) references incorrect serial number of 90437000. It should state This document applies to ROTAFLOW Consoles as of serial number 94175452 affected by the implementation of the new IEC standard 60601 1 2, 4th edition
Z-2957-2020 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Electrical /Electronic Property Problem2
- Pumping Stopped2
- Improper Flow or Infusion1
- Pumping Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 27, 2026Pumping Stopped
THE EVENT OCCURRED IN THE US. IT WAS REPORTED THAT AN ERROR MESSAGE OCCURRED ON THE ROTAFLOW CONSOLE AND THE PUMP STOPPED DURING PATIENT TREATMENT. THE CUSTOMER REBOOTED THE DEVICE AND IT STARTED TO WORK. AFTERWARDS, THE ROTAFLOW WAS REPLACED ON THE PATIENT. THE CUSTOMER DID NOT REMEMBER WHICH EXACT ERROR MESSAGE OCCURRED ON THE ROTAFLOW CONSOLE. NO HARM TO ANY PERSON HAS BEEN REPORTED. A REPORT IS REQUIRED DUE TO THE UNEXPECTED PUMP STOP AND THE DEVICE EXCHANGE. COMPLAINT ID: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24467516 ↗MalfunctionDec 1, 2025Electrical /Electronic Property Problem; Pumping Stopped; Electrical /Electronic Property Problem; Pumping Stopped
COMPLAINT ID: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23677919 ↗MalfunctionSep 5, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE EVENT OCCURRED IN THE UNITED STATES OF AMERICA. IT WAS REPORTED THAT THE ERROR MESSAGE ¿HEAD ERROR¿ OCCURRED ON THE ROTAFLOW CONSOLE. NO MORE INFORMATION PROVIDED. MORE INFORMATION REQUESTED BUT STILL PENDING. NO HARM TO ANY PERSON HAS BEEN REPORTED. THE REPORTED ¿HEAD ERROR¿ COULD LEAD TO A PUMP STOP DURING USE, THEREFORE A REPORT IS REQUIRED. COMPLAINT ID: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22975487 ↗MalfunctionAug 5, 2025Improper Flow or Infusion; Improper Flow or Infusion
THE EVENT OCCURRED IN THE US. IT WAS REPORTED THAT THE ERROR MESSAGE ¿SIG ERROR¿ OCCURRED ON THE ROTAFLOW CONSOLE. THERE WAS NO PUMP STOP. THE RPMS WERE VISIBLE BUT ON THE FLOW SCREEN JUST DASHES ¿---¿ WERE DISPLAYED. THE CONTACT CREAM WAS REAPPLIED AND THE HAND CRANK WAS USED. THE DEVICE WAS IMMEDIATELY EXCHANGED WITH A BACKUP DEVICE. NO HARM TO ANY PERSON HAS BEEN REPORTED. DUE TO THE REPORTED EXCHANGE OF THE DEVICE DURING PATIENT TREATMENT A REPORT IS REQUIRED. COMPLAINT ID: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22712194 ↗MalfunctionApr 22, 2025Pumping Problem; Pumping Problem
COMPLAINT ID: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21878911 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K991864. One submission often covers many models, so check the product and lot details in each record.
- Z-0678-2020 · Dec 3, 2019 · TerminatedMaquet Cardiopulmonary (MCP)/Getinge ROTAFLOW Drive, Model Number 701022161. The device is a component of the ROTAFLOW System Console.
Other PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE devices
- CENTRIMAG MOTOR 102956Thoratec Switzerland GmbH · Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type2 recalls4 deaths591 reports
- Impella Impella 5.0 AIC Pump SetAbiomed, Inc. · Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type3 recalls56 deaths256 reports
- CentriMag™ 102953Thoratec Switzerland GmbH · Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type1 recall35 deaths209 reports
- Impella Impella LD AIC Pump SetAbiomed, Inc. · Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type3 recalls9 deaths57 reports
- NA HY8M65R2Medtronic, Inc. · PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE4 recalls0 reports
- Custom Perfusion System HY8L55R5Medtronic, Inc. · PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE3 recalls3 reports
- Custom Perfusion System BB6U40R11Medtronic, Inc. · PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE2 recalls1 report
- Custom Perfusion System BB8T05R1Medtronic, Inc. · PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page