Skip to content

ROTAFLOW Console, English/ USRecalls & reported problems

Maquet Cardiopulmonary AG · ROTAFLOW English/US US-Plug · Class 2

All models of ROTAFLOW (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04037691562445), received between Apr 22, 2025 and Feb 27, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (2 reports).

1 FDA recall record name this device's identifier.

1 other recall cover devices cleared under the same FDA submission (K991864); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 7, 2020Terminated

    Instruction For Use (IFU) references incorrect serial number of 90437000. It should state This document applies to ROTAFLOW Consoles as of serial number 94175452 affected by the implementation of the new IEC standard 60601 1 2, 4th edition

    Z-2957-2020 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Electrical /Electronic Property Problem2
  • Pumping Stopped2
  • Improper Flow or Infusion1
  • Pumping Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 27, 2026Pumping Stopped

    THE EVENT OCCURRED IN THE US. IT WAS REPORTED THAT AN ERROR MESSAGE OCCURRED ON THE ROTAFLOW CONSOLE AND THE PUMP STOPPED DURING PATIENT TREATMENT. THE CUSTOMER REBOOTED THE DEVICE AND IT STARTED TO WORK. AFTERWARDS, THE ROTAFLOW WAS REPLACED ON THE PATIENT. THE CUSTOMER DID NOT REMEMBER WHICH EXACT ERROR MESSAGE OCCURRED ON THE ROTAFLOW CONSOLE. NO HARM TO ANY PERSON HAS BEEN REPORTED. A REPORT IS REQUIRED DUE TO THE UNEXPECTED PUMP STOP AND THE DEVICE EXCHANGE. COMPLAINT ID: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24467516 ↗
  • MalfunctionDec 1, 2025Electrical /Electronic Property Problem; Pumping Stopped; Electrical /Electronic Property Problem; Pumping Stopped

    COMPLAINT ID: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23677919 ↗
  • MalfunctionSep 5, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE EVENT OCCURRED IN THE UNITED STATES OF AMERICA. IT WAS REPORTED THAT THE ERROR MESSAGE ¿HEAD ERROR¿ OCCURRED ON THE ROTAFLOW CONSOLE. NO MORE INFORMATION PROVIDED. MORE INFORMATION REQUESTED BUT STILL PENDING. NO HARM TO ANY PERSON HAS BEEN REPORTED. THE REPORTED ¿HEAD ERROR¿ COULD LEAD TO A PUMP STOP DURING USE, THEREFORE A REPORT IS REQUIRED. COMPLAINT ID: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22975487 ↗
  • MalfunctionAug 5, 2025Improper Flow or Infusion; Improper Flow or Infusion

    THE EVENT OCCURRED IN THE US. IT WAS REPORTED THAT THE ERROR MESSAGE ¿SIG ERROR¿ OCCURRED ON THE ROTAFLOW CONSOLE. THERE WAS NO PUMP STOP. THE RPMS WERE VISIBLE BUT ON THE FLOW SCREEN JUST DASHES ¿---¿ WERE DISPLAYED. THE CONTACT CREAM WAS REAPPLIED AND THE HAND CRANK WAS USED. THE DEVICE WAS IMMEDIATELY EXCHANGED WITH A BACKUP DEVICE. NO HARM TO ANY PERSON HAS BEEN REPORTED. DUE TO THE REPORTED EXCHANGE OF THE DEVICE DURING PATIENT TREATMENT A REPORT IS REQUIRED. COMPLAINT ID: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22712194 ↗
  • MalfunctionApr 22, 2025Pumping Problem; Pumping Problem

    COMPLAINT ID: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21878911 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K991864. One submission often covers many models, so check the product and lot details in each record.

  • Z-0678-2020 · Dec 3, 2019 · Terminated
    Maquet Cardiopulmonary (MCP)/Getinge ROTAFLOW Drive, Model Number 701022161. The device is a component of the ROTAFLOW System Console.

Other PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE devices

All PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page