Custom Perfusion System HY8L55R5Recalls & reported problems

Medtronic, Inc. · CUSTOM PACK HY8L55R5 FUSION · Class 2

All models of Custom Perfusion System (61)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169572263), received between Jun 8, 2023 and Apr 19, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).

3 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 12, 2022Class IIOpen, Classified

    An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

    Z-1028-2023 · Root cause: Process change control · Recall details · FDA record ↗

  2. Dec 12, 2022Class IIOpen, Classified

    An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

    Z-1027-2023 · Root cause: Process change control · Recall details · FDA record ↗

  3. Nov 16, 2016Class IITerminated

    Medtronic is initiating a voluntary product recall for specific lot numbers of the Affinity Fusion Oxygenators with Balance1 Biosurface. These are distributed as stand-alone devices, or as a combination unit with the Affinity Fusion Cardiotomy/Venous Reservoir, or as part of tubing packs. Medtronic has identified an out-of-specification condition exhibiting additional plastic (flash) in the arterial sampling port adjacent to the arterial outlet port of the oxygenator.

    Z-0826-2017 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak3
  • Crack1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 19, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THERE WAS A LEAK FROM THE LOCATION IN WHICH THE TUBING WAS CONNECTED TO THE LUER PORT. THE LEAK ORIGINATED FROM THE TUBING CONNECTION. THE SAME LEAK DID NOT APPEAR IN MULTIPLE PACKS. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19141549 ↗
  • MalfunctionJul 18, 2023Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THE CUSTOMER WAS PRIMING THE CIRCUIT. WHEN FLUID ENTERED THE CONE DURING GRAVITY PRIMING, IT BEGAN TO LEAK FROM WHAT APPEARED TO BE A CRACK AT THE TOP OF THE INLET CONNECTOR NEAR WHERE THE TUBING WAS CONNECTED. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. THE PACKAGING OF THE PACK WAS INTACT WITH NO DAMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17341434 ↗
  • MalfunctionJun 8, 2023Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM PACK, IT WAS REPORTED THAT THE CUSTOMER WAS PRIMING THE CIRCUIT PRIOR TO PATIENT ENTERING THE ROOM AND FLUID LEAK WAS FOUND COMING FROM THE PUMP HEAD DURING GRAVITY PRIME. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17085301 ↗

Other PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE devices

All PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE devices →