Custom Perfusion System BB6U40R11Recalls & reported problems
Medtronic, Inc. · CUSTOM PACK BB6U40R11 FUSION · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169449442), received between Apr 23, 2024 and Apr 23, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 12, 2022Open, Classified
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
Z-1027-2023 · Root cause: Process change control · FDA record ↗
- Feb 4, 2016Terminated
Medtronic is notifying customers who have or may receive specific lots of Medtronic Perfusion Tubing Packs that include the potentially affected cuvettes manufactured by Terumo Cardiovascular Systems (Terumo CVS). Terumo Cardiovascular Systems (Terumo CVS) has distributed an Urgent Safety Alert regarding the CD I H/S Cuvette. The cuvette is packaged into Medtronic Perfusion Tubing Packs and only sold as part of the pack.
Z-1023-2016 · Root cause: Component design/selection · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak1
- Material Separation1
Patient problems
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2024Fluid/Blood Leak; Material Separation; Fluid/Blood Leak; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THE QUICK CONNECTOR ON THE ARTERIAL LINE WAS LOOSE/SEPARATED. THE LOOSE/SEPARATED COMPONENT DID NOT FALL INTO THE PATIENT. THERE WAS MINIMAL BLOOD LOSS THAT WAS ESTIMATED AT 100ML. THERE WAS NO TRANSFUSION REQUIRED. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE COMPONENT SEPARATED DURING THE PROCEDURE.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 19163104 ↗
Other PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE devices
- CENTRIMAG MOTOR 102956Thoratec Switzerland GmbH · Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type2 recalls4 deaths591 reports
- Impella Impella 5.0 AIC Pump SetAbiomed, Inc. · Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type3 recalls56 deaths256 reports
- CentriMag™ 102953Thoratec Switzerland GmbH · Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type1 recall35 deaths209 reports
- Impella Impella LD AIC Pump SetAbiomed, Inc. · Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type3 recalls9 deaths57 reports
- NA HY8M65R2Medtronic, Inc. · PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE4 recalls0 reports
- Custom Perfusion System HY8L55R5Medtronic, Inc. · PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE3 recalls3 reports
- NA HY9D36R1Medtronic, Inc. · Pump, blood, cardiopulmonary bypass, non-roller type2 recalls0 reports
- NA BB8U52R8Medtronic, Inc. · Pump, blood, cardiopulmonary bypass, non-roller type2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.