ORTHOPILOT NP621RRecalls & reported problems

Aesculap AG · ORTHOPILOT BICORTICAL SCREW 35MM · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04038653331727), received between Nov 5, 2020 and Nov 5, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Patient-Device Incompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Patient-Device Incompatibility1

Patient problems

  • Bone Fracture(s)1
  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 5, 2020Patient-Device Incompatibility; Patient-Device Incompatibility

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH NP621R - ORTHOPILOT BICORTICAL SCREW 35MM. ACCORDING COMPLAINT DESCRIPTION A (B)(6) FEMALE PATIENT WHO UNDERWENT TKA USING ORTHOPILOT ON (B)(6) 2020, FELL AND SUFFERED AN OBLIQUE FRACTURE OF THE FEMORAL SHAFT. THE ONSET OF THE FRACTURE WAS OBSERVED IN THE VICINITY OF THE SCREW HOLE IN THE FEMEUR THAT WAS MADE USING A BICORTICAL SCREW. CT SCAN EXAMINATION REVEALED A FRACTURE LINE RUNNING FROM THE SCREW HOLE . THE SURGEON STATED THAT CAUSE OF THE POST OPERATIVE FRACTURE IS OBVIOUSLY THE PATIENT'S FALL, BUT THAT THE FEMEUR MIGHT HAVE BEEN MORE VULNERABLE TO FRACTURE AS A RESULT OF THE HOLE. REVISION IS SCHEDULE TO REINFORCE THE FEMEUR USING A PLATE. ADDITIONAL INFORMATION WAS NOT PROVIDED NOR AVAILABLE / WAS NOT AVAILABLE. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).

    Patient problems: Bone Fracture(s); Insufficient Information

    MAUDE report 10792890 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates