AESCULAP FF904SBRecalls & reported problems

Aesculap AG · CASPAR DISTRACTION PIN 14MM · Class 1

Adverse event reports
10
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04038653364312), received between Feb 24, 2021 and Jan 9, 2025 — classified as 7 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation6
  • Break3
  • Contamination1
  • Fracture1
  • Microbial Contamination of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Foreign Body In Patient4
  • Failure of Implant2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 9, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED TO AESCULAP INC. THAT A CASPAR DISTRACTION PIN 14MM. (PART # FF904SB) WAS USED DURING REMOVAL OF DISC/ARTHROPLASTY AND ANTERIOR CERVICAL DISCECTOMY AND FUSION (ACDF) C4-5 SURGERY. ACCORDING TO THE COMPLAINANT THE PIN BROKE AND THE PIECE COULD NOT BE EXPLANTED. ADDITIONAL INFORMATION WAS NOT PROVIDED BUT HAS BEEN REQUESTED. THE ADVERSE EVENT IS FILED UNDER AIC REFERENCE (B)(4).

    Patient problems: Foreign Body In Patient

    MAUDE report 21114457 ↗
  • InjuryJan 10, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED TO AESCULAP INC. THAT A CASPAR DISTRACTION PIN 14MM (ITEM # FF904SB) WAS USED DURING AN ANTERIOR CERVICAL DISCECTOMY AND FUSION (ACDF) PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE DISTRACTION PIN BROKE OFF INTO THE CERVICAL VERTEBRAE AND COULD NOT BE REMOVED. THE PROCEDURE WAS COMPLETED WITH ANOTHER AESCULAP DISTRACTION PIN. THE TOP TIP OF THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. A FRAGMENT REMAINED IN SITU; THE TOP TIP OF PIN WAS AVAILABLE FOR RETURN. THE ADVERSE EVENT IS FILED UNDER AIC REFERENCE (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16132017 ↗
  • MalfunctionJan 9, 2023Material Fragmentation; Material Fragmentation

    UPDATE: THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16125125 ↗
  • InjuryFeb 11, 2022Contamination; Microbial Contamination of Device; Contamination; Microbial Contamination of Device

    IT WAS REPORTED TO AESCULAP AG THAT A CASPAR DISTRACTION PIN 14MM (PART # FF904SB) WAS USED DURING AN ANTERIOR CERVICAL SPINAL PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, UPON OPENING THE PACKET, FOREIGN MATTER WAS NOTED TO BE ADHERED TO THE PIN. AN ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED. THERE WAS A SURGICAL DELAY DURING WHICH THE STERILE AREA WAS REESTABLISHED. A 12MM SCREW WAS USED INSTEAD. ADDITIONAL INFORMATION WAS NOT AVAILABLE. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13518573 ↗
  • InjurySep 9, 2021Material Fragmentation; Material Fragmentation

    NO CHANGES REQUIRED.

    Patient problems: Failure of Implant

    MAUDE report 12446829 ↗
  • MalfunctionSep 8, 2021Fracture; Material Fragmentation; Fracture; Material Fragmentation

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH FF904SB - CASPAR DISTRACTION PIN 14MM VIA INFORMATION RECEIVED FROM MW 0504140000-2021-8002. ACCORDING TO THE COMPLAINT DESCRIPTION, THE PIN BROKE DURING SURGERY AND REMAINED IN SITU. DISTRACTION PIN SHEARED WHEN SURGEON ATTEMPTED TO REMOVE IT FROM C3. (IT SHEARED WITHOUT ANY ABOVE AVERAGE AMOUNT OF FORCE). A PORTION OF THE PIN WAS FULLY EMBEDDED IN BONE WITHOUT ANY POTENTIAL TO CAUSE HARM. THE PIECE BROKE OFF IN THE BONE. IT WOULD NOT HAVE BEEN SAFE TO REMOVE IT. NO SURGERY DELAY. A FRAGMENT REAMINED IN SITU. ADDITIONAL DETAILS WERE NOT PROVIDED. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12442381 ↗
  • InjuryJun 23, 2021Break; Break

    NO UPDATES.

    Patient problems: Foreign Body In Patient

    MAUDE report 12050806 ↗
  • MalfunctionJun 21, 2021Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH FF904SB - CASPAR DISTRACTION PIN 14MM. ACCORDING TO THE COMPLAINT DESCRIPTION, THE TIP OF THE DEVICE BROKE DURING SURGERY. WHILE TRYING TO REMOVE THE PIN, THE TIP BROKE OFF INTO THE ANTERIOR VERTREBAL BODY. THE PIECE WAS LEFT INSITU AS THE SURGEON STATED THAT TRYING TO REMOVE IT WOULD CAUSE MORE HARM TO PATIENT. THERE WAS NO PATIENT HARM BUT A SURGICAL DELAY OF 30 MINUTES OCCURRED. ADDITIONAL INFORMATION WAS NOT PROVIDED. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12033772 ↗
  • InjuryMar 25, 2021Break; Break

    KW 05-06-2021. IT WAS REPORTED AESCULAP AG THAT A CASPAR DISTRACTION PIN 14MM (PART # FF904SB) WAS USED DURING AN UNKNOWN PROCCEDURE PERFORMED ON (B)(6), 2021. ACCORDING TO THE COMPLAINANT, THE SCREW BROKE DURING SURGERY. RUPTURE WITH INTRAOSSEOUS SEQUESTRATION OF A CASPAR DISTRACTION SCREW DURING THE INTERVENTION OF A CURE FOR C6-C7 PSEUDO-ARTHROSIS AND INSTALLATION OF OSD PLATE. CONSEQUENCE: PERSISTENCE OF MATERIAL AT LEVEL C6. A FRAGMENT OF THE BROKEN DEVICE REAMINED IN SITU. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4).

    Patient problems: Foreign Body In Patient

    MAUDE report 11571959 ↗
  • InjuryFeb 24, 2021Break; Break

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH FF904SB-CASPAR DISTRACTION PIN 14MM. ACCORDING TO THE COMPLAINT DESCRIPTION, THE CASPAR SCREW BROKE DURING SURGERY ON (B)(6). THE BROKEN PART REMAINED IN THE VERTEBRA, BY MEDICAL CHOICE. ADDITIONAL INFORMATION WAS NOT PROVIDED NOR AVAILABLE / WAS NOT AVAILABLE. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).

    Patient problems: Foreign Body In Patient

    MAUDE report 11373042 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.