Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04042409363221), received between Oct 7, 2022 and Oct 7, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth1
- Misassembled During Installation1
Patient problems
- Nonunion/Delayed-union Bone Fracture1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryOct 7, 2022Loosening of Implant Not Related to Bone-Ingrowth; Misassembled During Installation; Loosening of Implant Not Related to Bone-Ingrowth; Misassembled During Installation
AFTER TREATMENT OF A DISTAL RADIUS FRACTURE, THE PLATE OSTEOSYNTHESIS FAILED WITHIN 4 WEEKS, RESULTING IN DISLOCATION OF THE SHAFT SCREWS. THE OSTEOSYNTHESIS WAS RESTORED IN A REVISION OPERATION USING A LONGER PLATE. THE TREATED PATIENT (90 YEARS OLD) HAS MULTIPLE COMORBIDITIES (PRONOUNCED OSTEOPOROSIS AND DEMENTIA). THE SURGEON COULD NOT EXCLUDE A FALL EVENT. ON X-RAY ON (B)(6) 2022 LOOSENING OF THE SHAFT SCREWS COULD BE SEEN. THE REVISION SURGERY WAS PERFORMED ON (B)(6) 2022.
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 15562787 ↗
Other Plate, fixation, bone devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-MLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-MRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page