Olympus A0358Recalls & reported problems

OLYMPUS Winter & Ibe GmbH · HF-cable, monopolar, 3.5 m, UES-30, Erbe Intl. and Valleylab (new) HF-unit to HiQ, HiQ+ · Class 2

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04042761001557), received between Nov 17, 2021 and May 19, 2025 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Thermal Decomposition of Device2
  • Electrical Shorting1
  • Failure to Conduct1
  • Mechanical Problem1
  • Output Problem1
  • Peeled/Delaminated1
  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Full thickness (Third Degree) Burn1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2025Failure to Conduct; Failure to Conduct

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED COMPONENT FAILURE OF THE HF-CABLE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22048823 ↗
  • MalfunctionJul 15, 2024Break; Break

    IT WAS REPORTED, THE MONOPOLAR HIGH FREQUENCY CABLE WAS BROKEN. THE USER WAS IMAGING INSIDE THE PATIENT AND THE CAUTERY SCREEN GAVE A CABLE WARNING. THE ISSUE OCCURRED DURING AN UNKNOWN THERAPEUTIC PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19740759 ↗
  • MalfunctionDec 1, 2023Peeled/Delaminated; Peeled/Delaminated

    THE CUSTOMER REPORTED TO OLYMPUS, A PIECE OF METAL HF-CABLE, MONOPOLAR HAD FALLEN ON THE MACHINE TABLE DURING THE SURGERY. IT WAS SUSPECTED THAT THE COATING INSIDE AND AT THE CONTACT POINT WITH THE FORCEPS MAY HAVE PEELED OFF. THE TECHNIQUE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18249470 ↗
  • MalfunctionNov 28, 2023Break; Electrical Shorting; Break; Electrical Shorting

    THE CUSTOMER INITIALLY REPORTED TO OLYMPUS THAT WHEN CUTTING THE GALLBLADDER OF THE PATIENT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, AN ELECTRIC KNIFE WAS USED TO PERFORM ELECTROCOAGULATION FOR HEMOSTASIS, AND A SUDDEN ELECTRICAL CONDUCTIVITY AT THE ELECTROCOAGULATION WIRE HANDLE LED TO THE SURGEON'S GLOVE BREAKAGE, CAUSING A BURN OF APPROXIMATELY 0.5X0.3 MM ON THE OPERATOR'S RIGHT THUMB. THERE WAS NO PATIENT INJURY REPORTED. OLYMPUS THEN RECEIVED FURTHER INFORMATION CLARIFYING THAT THE HIGH-FREQUENCY CABLE (A0358) WAS DAMAGED, AND THE SURGEON TOUCHED THE DAMAGED AREA WITH THEIR HANDS DURING USE, RESULTING IN BURNS. THERE WAS NO MALFUNCTION OF THE ESG-400 THAT WAS ALSO USED IN THE PROCEDURE. THE CUSTOMER FURTHER CLARIFIED THAT THE SURGEON SUSTAINED A THIRD DEGREE BURN WITH A SMALL AREA OF 2X5 MM TREATED WITH COLD APPLICATION. THERE WERE NO OTHER SYMPTOMS OR COMPLICATIONS. THE SURGEON CURRENTLY DOES NOT REPORT ANY DISCOMFORT.

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 18219819 ↗
  • MalfunctionOct 31, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS, THE TELESCOPE, 10 MM, 0°, QUICK LOCK, AUTOCLAVABLE HAD A BROKEN LIGHT POST. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18040016 ↗
  • MalfunctionFeb 13, 2023Thermal Decomposition of Device; Thermal Decomposition of Device

    OLYMPUS (OCI) WAS INFORMED BY THE NURSE AT THE USER FACILITY, THE CUSTOMER HIGH-FREQUENCY MONOPOLAR CABLE MALFUNCTIONED, ¿THE CABLE STARTED TO HUM (AUDITORY) AND THEN BECAME RED AND SNAPPED IN TWO.¿ THE CUSTOMER REPORTED EVENT WAS FOUND DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE. ANOTHER CABLE WAS USED TO COMPLETE THE INTENDED PROCEDURE. NO OTHER EQUIPMENT WAS REPLACED AND NO DEATH OR INJURY AND NO IMPACT TO PATIENT OR OTHER WAS REPORTED TO OLYMPUS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16368038 ↗
  • MalfunctionAug 25, 2022Sparking; Output Problem; Temperature Problem; Sparking; Output Problem; Temperature Problem

    OLYMPUS WAS INFORMED THAT DURING AN UNSPECIFIED THERAPEUTIC LAPAROSCOPIC SURGERY, THE AREA NEAR THE CONNECTION BETWEEN THE HF-CABLE AND THE HF-ELECTRODE POPPED AND A SPARK OCCURRED. BEFORE THIS HAPPENED THE SURGEON HAD FELT THE HANDLE OF THE HF-ELECTRODE TO BE HOT AND THE OUTPUT PERFORMANCE OF THE ELECTRODE TO BE WEAKER THAN NORMAL. THE SURGEON THEN REPLACED BOTH THE HF ELECTRODE AND THE HF-CABLE. THE INTENDED PROCEDURE WAS COMPLETED AND THERE WAS NO REPORT ABOUT AN ADVERSE EVENT OR PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15292496 ↗
  • MalfunctionMay 20, 2022Break; Break

    OLYMPUS WAS INFORMED THAT DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE THE HF CABLE BROKE. NO FURTHER INFORMATION WAS PROVIDED BUT THE INTENDED PROCEDURE WAS COMPLETED USING ANOTHER SIMILAR DEVICE AND THERE WAS NO REPORT ABOUT AN ADVERSE EVENT OR PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14460678 ↗
  • MalfunctionDec 6, 2021Thermal Decomposition of Device; Thermal Decomposition of Device

    THE OLYMPUS TERRITORY SALES MANAGER WAS INFORMED BY THE STERILE PROCESSING MANAGER AT THE USER FACILITY THAT THE HIGH FREQUENCY MONOPOLAR CABLE REPORTEDLY ¿EXPLODED, WITH A SMALL FLAME¿ DURING AN UNSPECIFIED EVENT. THERE WAS NO INJURY OR HARM REPORTED. NO DEVICE WILL BE RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12935044 ↗
  • MalfunctionNov 17, 2021Break; Mechanical Problem; Break; Mechanical Problem

    OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC LAPAROSCOPIC APPENDIX REMOVAL PROCEDURE, THE HF-CABLE BROKE AND MADE AN EXPLOSION SOUND. THE INTENDED PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A SIMILAR CABLE AND THERE WAS NO REPORT ABOUT AN ADVERSE EVENT OR PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12826430 ↗

Other Laparoscope, General & Plastic Surgery devices

All Laparoscope, General & Plastic Surgery devices →