Skip to content

Olympus A3765ARecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Telescope, 1.9 mm, 30°, 210mm, autoclavable · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04042761005739), received between Feb 23, 2023 and Feb 23, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Detachment of Device or Device Component2
  • Loose or Intermittent Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 23, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID ENDOSCOPE TELESCOPE HAD A LOOSE OUTER TUBE DIRECTION PIN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24413339 ↗
  • MalfunctionJul 11, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED BROKEN LENSES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22478732 ↗
  • MalfunctionMay 23, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22079349 ↗
  • MalfunctionAug 22, 2024Break; Break

    IT WAS REPORTED THE RIGID ENDOSCOPE TELESCOPE HAD A BROKEN LOCK PIN. THE ISSUE OCCURRED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20031570 ↗
  • MalfunctionAug 20, 2024Break; Loose or Intermittent Connection; Break; Loose or Intermittent Connection

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE AUTOCLAVABLE TELESCOPE EXHIBITED A BROKEN/LOOSE COMPONENT ON THE MAIN BODY. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20006248 ↗
  • MalfunctionAug 14, 2024Break; Break

    IT WAS REPORTED THE AUTOCLAVABLE TELESCOPE TIP WAS CHIPPED. THERE WERE NO REPORTS OF PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19985891 ↗
  • MalfunctionAug 22, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED TO OLYMPUS, THE TELESCOPE, HAD A PEDIATRIC LENS PIN BREAK. THE ISSUE WAS FOUND DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17600565 ↗
  • MalfunctionFeb 23, 2023Break; Break

    IT WAS REPORTED TO OLYMPUS THAT A TELESCOPE SCOPE WAS NOT WORKING. THE FIELD SERVICE ENGINEER (FSE) CHECKED THE SCOPE AND FOUND THE LOCK PIN BROKEN. THE REPORTED ISSUE WAS FOUND DURING MAINTENANCE OF THE DEVICE. THERE WAS NO INJURY OR HEALTH DAMAGE DUE TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16432424 ↗

Other Resectoscope devices

All Resectoscope devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page