Skip to content

Olympus A42011ARecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Resection sheath, 8 mm, for 8.5 mm/26 Fr. outer sheath, ABS · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus resectoscope sheath is a reusable part used in endoscopic procedures. Most of its 958 reports describe parts breaking, and some describe parts detaching. The reports mention no deaths and 21 injuries. One recall, started in November 2025 and still open, concerns complaints that the ceramic tip breaks.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
963
Deaths reported
0
Injuries reported
20
Recalls naming it
1

Summary

The FDA's MAUDE database contains 963 adverse event reports that list this exact device (DI 04042761023658), received between May 4, 2018 and Sep 24, 2026 — classified as 20 injuries, 943 malfunctions.

The most frequently reported specific device problem is “Break” (706 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 16 reportsDec 2023: 39 reportsJan 2024: 20 reportsFeb 2024: 16 reportsMar 2024: 19 reportsApr 2024: 33 reportsMay 2024: 16 reportsJun 2024: 32 reportsJul 2024: 38 reportsAug 2024: 31 reportsSep 2024: 35 reportsOct 2024: 17 reportsNov 2024: 13 reportsDec 2024: 6 reportsJan 2025: 24 reportsFeb 2025: 17 reportsMar 2025: 15 reportsApr 2025: 13 reportsMay 2025: 18 reportsJun 2025: 24 reportsJul 2025: 26 reportsAug 2025: 25 reportsSep 2025: 17 reportsOct 2025: 12 reportsNov 2025: 30 reportsDec 2025: 25 reportsJan 2026: 19 reportsFeb 2026: 10 reportsMar 2026: 28 reportsApr 2026: 16 reportsMay 2026: 32 reportsJun 2026: 6 reportsJul 2026: 6 reportsAug 2026: 16 reportsSep 2026: 14 reportsOct 2026: 0 reports20242025202639
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 11, 2025Open, Classified

    Complaints of the ceramic tip of the resection sheath breaking have been received.

    Z-1448-2026 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Break706
  • Detachment of Device or Device Component128
  • Loose or Intermittent Connection41
  • Crack29
  • Material Fragmentation26
  • Fracture23
  • Mechanical Problem22
  • Material Integrity Problem16

Patient problems

  • No Clinical Signs, Symptoms or Conditions947
  • No Known Impact Or Consequence To Patient4
  • Blood Loss3
  • Hemorrhage/Blood Loss/Bleeding3
  • Device Embedded In Tissue or Plaque2
  • Insufficient Information2
  • Laceration(s)2
  • Unspecified Tissue Injury2

Recent adverse event reports

The 25 most recent of 963 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Break; Detachment of Device or Device Component

    The customer reported the ceramic head damaged during reprocessing involving an Olympus Endoscope sheath, reusable. There was no patient injury reported. During the device evaluation the investigator indicated the ceramic tip is missing.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883709 ↗
  • MalfunctionSep 22, 2026Break

    It was reported that during set up / inspection for use, the resection sheath ceramic head was damaged. There was no harm or injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50849759 ↗
  • MalfunctionSep 21, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50848030 ↗
  • MalfunctionSep 15, 2026Break

    It was reported that during set up / inspection for use, the resection sheath ceramic head was damaged. There was no harm or injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50766464 ↗
  • MalfunctionSep 15, 2026Crack

    The device was returned to the service center. During inspection of the device, the resection sheath ceramic tip had cracks. There was no harm or injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50766348 ↗
  • MalfunctionSep 10, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50719992 ↗
  • MalfunctionSep 10, 2026Use of Device Problem

    The customer returned the reusable Endoscope sheath for a separate issue. During the device analysis, it was observed that the ceramic tip was broken was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50717551 ↗
  • MalfunctionSep 9, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50699224 ↗
  • MalfunctionSep 8, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50680424 ↗
  • MalfunctionSep 8, 2026Unstable

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50676275 ↗
  • MalfunctionSep 2, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50572021 ↗
  • MalfunctionSep 1, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50551253 ↗
  • MalfunctionSep 1, 2026Detachment of Device or Device Component

    The customer returned the reusable Endoscope sheath, for a separate issue. During the initial evaluation of the returned device, loose Ceramic tip was observed. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50535884 ↗
  • MalfunctionSep 1, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50535139 ↗
  • MalfunctionAug 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50489926 ↗
  • MalfunctionAug 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50488995 ↗
  • MalfunctionAug 27, 2026Break

    The customer returned the reusable endoscope sheath to the service center for a separate issue. During the device analysis, the service repair technician indicated that the ceramic tip (insulation beak) was broken. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50477517 ↗
  • MalfunctionAug 27, 2026Break

    The customer returned the reusable endoscope sheath to the service center for a separate issue. During the device analysis, the service repair technician indicated that the ceramic tip was broken. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50476031 ↗
  • MalfunctionAug 26, 2026Detachment of Device or Device Component

    It was reported that the ceramic tip of the inner sheath of the resectoscope fell off on its own into the patient¿s uterine cavity. The issue was found during a hysteroscopic endometrial polypectomy procedure. It was eventually removed using multiple sets of instruments. There were no reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50459353 ↗
  • MalfunctionAug 26, 2026Break

    The customer returned the reusable endoscope sheath for a separate issue. During the initial evaluation of the returned device, it was observed the ceramic tip was broken. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50457980 ↗
  • MalfunctionAug 24, 2026Crack

    No new additional information received from the customer.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50421584 ↗
  • MalfunctionAug 21, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50373084 ↗
  • MalfunctionAug 21, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50371636 ↗
  • MalfunctionAug 17, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50286436 ↗
  • MalfunctionAug 14, 2026Break; Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50262324 ↗

Other Resectoscope devices

All Resectoscope devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.