Olympus A60201ARecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Handle "HiQ+", Ergo S, unipolar · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04042761009799), received between Jan 14, 2019 and May 11, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Material Deformation1
- Material Separation1
Patient problems
- Device Embedded In Tissue or Plaque1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 11, 2023Material Separation; Material Deformation; Material Separation; Material Deformation
A USER FACILITY REPORTED TO OLYMPUS THAT DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE USING THE JAWS "HIQ+", 5 X 330 MM, GRASPING FORCEPS, CROCEOLMI, IT BECAME IMPOSSIBLE TO GRASP. "ABNORMAL NOISE" WAS REPORTED. THE SHAFT OF THE UNIT SEPARATED FROM THE MOVING PART OF THE JAW, AND DID NOT INTERLOCK WITH THE HANDLE OPERATION. IT WAS STATED THAT "FRACTURE SURFACE" WAS NOT VISIBLE AND DEFORMATION OF THE COMBINED SHEATHS WAS CONFIRMED. THE PROCEDURE WAS COMPLETED BY REPLACING THE EQUIPMENT WITH SIMILAR EQUIPMENT. IT WAS ALSO STATED THAT THE AT THE TIME OF THE REPORT, THE HOSPITAL COULD NOT CONFIRM THE PARTS HAD NOT BEEN RETAINED IN THE PATIENT, AND IT WAS BEING TREATED AS IF THE DEVICES HAD FALLEN INTO A BODY CAVITY. ADDITIONAL INFORMATION WAS REQUESTED MULTIPLE TIMES, BUT WAS NOT RECEIVED. THIS EVENT INVOLVED 3 DEVICES. PATIENT IDENTIFIER (B)(6) IS FOR THE JAWS "HIQ+", 5 X 330 MM, GRASPING FORCEPS, CROCEOLMI PATIENT IDENTIFIER (B)(6) IS FOR THE SHAFT "HIQ+", 5 X 330 MM PATIENT IDENTIFIER (B)(6) IS FOR THE HANDLE "HIQ+", ERGO S, UNIPOLAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16911611 ↗InjuryJan 14, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
OLYMPUS WAS INFORMED THAT AFTER A THERAPEUTIC VIDEO-ASSISTED THORACOSCOPIC SURGERY (VATS) PROCEDURE HAD BEEN SUCCESSFULLY COMPLETED, IT WAS NOTICED THAT THE SCREW OF THE HINGE AXIS OF THE UNIPOLAR HANDLE WAS MISSING. THE PATIENT WAS THEN RE-OPENED ON THE SAME DAY AND THE COMPONENT RETRIEVED FROM THE PATIENT¿S THORACIC CAVITY. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8243591 ↗
Other Laparoscope, General & Plastic Surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page