Skip to content

Olympus A70960ARecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Telescope "TrueView II", 2.7 mm, 0°, autoclavable · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus arthroscope is a small telescope used to look inside joints. The 34 reports mostly describe inner parts breaking or chipping, sometimes with poor image quality, with no deaths and no injuries. No recalls name this device. Reports fell from 18 in the previous 12 months to 5 in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 04042761010627), received between Sep 13, 2023 and Apr 2, 2026 — classified as 34 malfunctions.

The most frequently reported specific device problem is “Break” (25 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 4 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 4 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 3 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break25
  • Defective Component2
  • No Display/Image2
  • Pitted2
  • Component Missing1
  • Contamination1
  • Crack1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions34

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 2, 2026Break; Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID ARTHROSCOPE EXHIBITED BROKEN AND CHIPPED INNER BARREL ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24764185 ↗
  • MalfunctionMar 23, 2026Material Twisted/Bent

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID ARTHROSCOPE HAD BROKEN OR CHIPPED INNER BARREL ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24660907 ↗
  • MalfunctionJan 20, 2026Break; Defective Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID ARTHROSCOPE HAD A BROKEN DISTAL END OF THE INSERTION TUBE, AND A COMPONENT FAILURE OF INSERTION TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24114869 ↗
  • MalfunctionJan 16, 2026Break

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24099621 ↗
  • MalfunctionJan 5, 2026Break

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23962256 ↗
  • MalfunctionSep 15, 2025Component Missing; Component Missing

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23050743 ↗
  • MalfunctionSep 9, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD POOR VISIBILITY IMAGE QUALITY. THE EVENT HAD OCCURRED DURING SERVICE / MAINTENANCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23004023 ↗
  • MalfunctionJul 17, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22531453 ↗
  • MalfunctionJun 7, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED DISTAL END OF THE TELESCOPE WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22162554 ↗
  • MalfunctionJun 5, 2025Break; Break

    IT WAS REPORTED THAT THE SUBJECT DEVICE'S END IS NOT WORKING, BROKEN. THE ISSUE WAS FOUND DURING SET UP. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22153283 ↗
  • MalfunctionMay 29, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SCOPE EXHIBITED INTERNAL LENS DAMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22109234 ↗
  • MalfunctionMay 29, 2025Break; Break

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD A BROKEN LENS. THIS ISSUE OCCURRED DURING REPROCESSING OF THE DEVICE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22105505 ↗
  • MalfunctionApr 25, 2025Break; Break

    IT WAS REPORTED THE SUBJECT DEVICE WAS BROKE DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21907125 ↗
  • MalfunctionApr 23, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED BROKEN AND DISPLACED LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21895363 ↗
  • MalfunctionMar 24, 2025Break; Contamination; Break; Contamination

    IT WAS REPORTED THE RIGID VIDEOSCOPE HAD STAINS THAT ADHERED TO THE SCOPE AND COULD NOT BE REMOVED EVEN AFTER CLEANING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21679679 ↗
  • MalfunctionFeb 25, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID SCOPE EXHIBITED A BROKEN LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21459195 ↗
  • MalfunctionFeb 20, 2025No Display/Image; No Display/Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID SCOPE EXHIBITED BROKEN OPTICS LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21423970 ↗
  • MalfunctionJan 9, 2025Break; Break

    IT WAS REPORTED THAT THE SUBJECT DEVICE WAS BROKEN. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21119074 ↗
  • MalfunctionDec 27, 2024Break; Break

    IT WAS REPORTED THE RIGID SCOPE EXHIBITED A CHIP FOUND ON THE TIP OF THE PROBE, IT WAS FOUND DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21029264 ↗
  • MalfunctionNov 26, 2024Break; Break

    IT WAS REPORTED THE SUBJECT DEVICE HAD CHIPPED LENS. THE ISSUE OCCURRED DURING THE PRE-USE INSPECTION. THERE WERE NO REPORTS OF PATIENT HARM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20779384 ↗
  • MalfunctionNov 26, 2024Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20777945 ↗
  • MalfunctionNov 25, 2024Crack; Crack

    WHILE EVALUATED A RETURNED OLYMPUS TRUEVIEW TELESCOPE, IT WAS DISCOVERED THAT THE INTERNAL LENS WAS CRACKED. THERE WAS NO PATIENT INVOLVEMENT DURING THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20773765 ↗
  • MalfunctionNov 1, 2024Break; Break

    IT WAS REPORTED THE RIGID ARTHROSCOPE EXPERIENCED CONCEALED BREAKS IN THE LIGHT GUIDE BUNDLE. THE ISSUE OCCURRED DURING A DIAGNOSTIC NASAL EXAMINATION. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM OR PROCEDURAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20587872 ↗
  • MalfunctionSep 13, 2024Break; Break

    IT WAS REPORTED THE DISTAL TIP OF THE RIGID ARTHROSCOPE WAS BROKEN. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR AN UNSPECIFIED DIAGNOSTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20211376 ↗
  • MalfunctionJun 5, 2024Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE RIGID ARTHROSCOPE EXHIBITED LIGHT GUIDE END FACE DAMAGE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19472338 ↗

Other Arthroscope devices

All Arthroscope devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page