Olympus A20976ARecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Bridge, one way · Class 2
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04042761020398), received between Jun 17, 2022 and Jun 17, 2022 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Contamination” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Contamination7
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 17, 2022Contamination; Contamination
THE OLYMPUS REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER, DURING A PROCEDURE, THE BRIDGE WAS SWIVELING DURING A CASE. AFTER THE PROCEDURE, AT REPROCESSING, IT WAS NOTED THE BRIDGE UNSCREWED APART. THE TWO PIECES OF THE BRIDGE WERE DISMANTLED AND BIOBURDEN, POSSIBLY CRUSTED BLOOD, WAS FOUND IN THE BRIDGE. THE FACILITY TOOK APART THE OTHER SIX (6) BRIDGES AND BIOBURDEN WAS FOUND ON THE OTHER SIX (6) BRIDGES AS WELL. NO PATIENT HARM REPORTED. THIS COMPLAINT IS RELATED TO PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14736940 ↗MalfunctionJun 17, 2022Contamination; Contamination
THE OLYMPUS REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER, DURING A PROCEDURE, THE BRIDGE WAS SWIVELING DURING A CASE. AFTER THE PROCEDURE, AT REPROCESSING, IT WAS NOTED THE BRIDGE UNSCREWED APART. THE TWO PIECES OF THE BRIDGE WERE DISMANTLED AND BIOBURDEN, POSSIBLY CRUSTED BLOOD, WAS FOUND IN THE BRIDGE. THE FACILITY TOOK APART THE OTHER SIX BRIDGES AND BIOBURDEN WAS FOUND ON THE OTHER SIX BRIDGES AS WELL. NO PATIENT HARM REPORTED. THIS COMPLAINT IS RELATED TO PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14736586 ↗MalfunctionJun 17, 2022Contamination; Contamination
THE OLYMPUS REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER, DURING A PROCEDURE, THE BRIDGE WAS SWIVELING DURING A CASE. AFTER THE PROCEDURE, AT REPROCESSING, IT WAS NOTED THE BRIDGE UNSCREWED APART. THE TWO PIECES OF THE BRIDGE WERE DISMANTLED AND BIOBURDEN, POSSIBLY CRUSTED BLOOD, WAS FOUND IN THE BRIDGE. THE FACILITY TOOK APART THE OTHER SIX (6) BRIDGES AND BIOBURDEN WAS FOUND ON THE OTHER SIX (6) BRIDGES AS WELL. NO PATIENT HARM REPORTED. THIS COMPLAINT IS RELATED TO PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14736213 ↗MalfunctionJun 17, 2022Contamination; Contamination
THE OLYMPUS REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER, DURING A PROCEDURE, THE BRIDGE WAS SWIVELING DURING A CASE. AFTER THE PROCEDURE, AT REPROCESSING, IT WAS NOTED THE BRIDGE UNSCREWED APART. THE TWO PIECES OF THE BRIDGE WERE DISMANTLED AND BIOBURDEN, POSSIBLY CRUSTED BLOOD, WAS FOUND IN THE BRIDGE. THE FACILITY TOOK APART THE OTHER SIX BRIDGES AND BIOBURDEN WAS FOUND ON THE OTHER SIX BRIDGES AS WELL. NO PATIENT HARM REPORTED. THIS COMPLAINT IS RELATED TO PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14735493 ↗MalfunctionJun 17, 2022Contamination; Contamination
THE OLYMPUS REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER, DURING A PROCEDURE, THE BRIDGE WAS SWIVELING DURING A CASE. AFTER THE PROCEDURE, AT REPROCESSING, IT WAS NOTED THE BRIDGE UNSCREWED APART. THE TWO PIECES OF THE BRIDGE WERE DISMANTLED AND BIOBURDEN, POSSIBLY CRUSTED BLOOD, WAS FOUND IN THE BRIDGE. THE FACILITY TOOK APART THE OTHER SIX BRIDGES AND BIOBURDEN WAS FOUND ON THE OTHER SIX BRIDGES AS WELL. NO PATIENT HARM REPORTED. THIS COMPLAINT IS RELATED TO PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14735404 ↗MalfunctionJun 17, 2022Contamination; Contamination
THE OLYMPUS REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER, DURING A PROCEDURE, THE BRIDGE WAS SWIVELING DURING A CASE. AFTER THE PROCEDURE, AT REPROCESSING, IT WAS NOTED THE BRIDGE UNSCREWED APART. THE TWO PIECES OF THE BRIDGE WERE DISMANTLED AND BIOBURDEN, POSSIBLY CRUSTED BLOOD, WAS FOUND IN THE BRIDGE. THE FACILITY TOOK APART THE OTHER SIX BRIDGES AND BIOBURDEN WAS FOUND ON THE OTHER SIX BRIDGES AS WELL. NO PATIENT HARM REPORTED. THIS COMPLAINT IS RELATED TO PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14735323 ↗MalfunctionJun 17, 2022Contamination; Contamination
THE OLYMPUS REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER, DURING A PROCEDURE, THE BRIDGE ON THE RIGID CYSTOSCOPE WAS SWIVELING DURING A CASE. AFTER THE PROCEDURE, AT REPROCESSING, IT WAS NOTED THE BRIDGE UNSCREWED APART. THE TWO PIECES OF THE BRIDGE WERE DISMANTLED AND BIOBURDEN, POSSIBLY CRUSTED BLOOD, WAS FOUND IN THE BRIDGE. THE FACILITY TOOK APART THE OTHER SIX (6) BRIDGES AND BIOBURDEN WAS FOUND ON THE OTHER SIX (6) BRIDGES AS WELL. NO PATIENT HARM REPORTED. THIS COMPLAINT IS RELATED TO PATIENT IDENTIFIERS: (B)(6), (B)(6), (B)(6), (B)(6), (B)(6), (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14735262 ↗
Other Endoscope, Ac-Powered And Accessories devices
- Olympus WA22810AOLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories2 recalls8 reports
- Olympus A22201DOLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall7 reports
- Olympus A22205DOLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall5 reports
- Olympus A22257COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall1 report
- Olympus A22207COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall1 report
- Olympus A22221COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall1 report
- Olympus A22222COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall0 reports
- Olympus A22231DOLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.