Olympus WA22810ARecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Outer sheath, 22.5 Fr. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04042761051729), received between Mar 11, 2024 and Apr 22, 2024 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Thermal Decomposition of Device” (6 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Sep 13, 2024Open, Classified
Olympus is removing the statement of compatibility with a GreenLight Laser for BPH therapy from the Instruction for Use (IFU) due to the potential for a damaged tip during use of a laser probe.
Z-0351-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Sep 13, 2024Open, Classified
Product being removed due to incompatibility when used in conjunction with a GreenLight Laser for BPH therapy which could damage the tip of the device.
Z-1437-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Thermal Decomposition of Device6
- Degraded1
- Sparking1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 22, 2024Degraded; Degraded
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE OUTER SHEATH EXHIBITED NUMEROUS SPARK MARKS AT THE TIP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19160087 ↗MalfunctionMar 26, 2024Thermal Decomposition of Device; Thermal Decomposition of Device
THE ISSUE WAS FOUND DURING REPROCESSING FOR A THERAPEUTIC LASER VAPORIZATION OF PROSTATE PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED AND THERE WAS NO DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18979464 ↗MalfunctionMar 26, 2024Thermal Decomposition of Device; Thermal Decomposition of Device
THE ISSUE WAS FOUND DURING REPROCESSING FOR A THERAPEUTIC LASER VAPORIZATION OF PROSTATE PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED AND THERE WAS NO DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18979356 ↗MalfunctionMar 26, 2024Thermal Decomposition of Device; Thermal Decomposition of Device
IT WAS REPORTED THE RIGID ENDOSCOPE SHEATH'S DISTAL TIP WAS BURNT. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18979305 ↗MalfunctionMar 26, 2024Thermal Decomposition of Device; Thermal Decomposition of Device
THE ISSUE WAS FOUND DURING REPROCESSING FOR A THERAPEUTIC LASER VAPORIZATION OF PROSTATE PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED AND THERE WAS NO DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18978795 ↗MalfunctionMar 26, 2024Thermal Decomposition of Device; Thermal Decomposition of Device
THE ISSUE WAS FOUND DURING REPROCESSING FOR A THERAPEUTIC LASER VAPORIZATION OF PROSTATE PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED AND THERE WAS NO DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18978704 ↗MalfunctionMar 18, 2024Thermal Decomposition of Device; Thermal Decomposition of Device
IT WAS REPORTED, THE OUTER SHEATH, 22.5 FR. EXHIBITED A BURNT DISTAL TIP. THE ISSUE OCCURRED DURING REPROCESSING FOR A THERAPEUTIC LASER VAPORIZATION OF PROSTATE PROCEDURE AND THE PROCEDURE WAS COMPLETED. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18923518 ↗MalfunctionMar 11, 2024Sparking; Sparking
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE OUTER SHEATH EXHIBITED MANY SPARK MARKS WERE CONFIRMED AT THE TIP. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18881743 ↗
Other Endoscope, Ac-Powered And Accessories devices
- Olympus A22201DOLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall7 reports
- Olympus A22205DOLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall5 reports
- Olympus A22221COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall1 report
- Olympus A22257COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall1 report
- Olympus A22207COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall1 report
- Olympus A22203COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall0 reports
- Olympus A22223COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall0 reports
- Olympus A22222COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page