Olympus A22257CRecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 14042761030448), received between Jun 10, 2019 and Jun 10, 2019 — classified as 1 malfunction.
The most frequently reported specific device problem is “Unintended Electrical Shock” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jun 8, 2023Open, Classified
Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment, implants or stents. As a result, fragments can fall inside the patient and must be retrieved. (2) The second Caution refers to the risk of sparkover when getting in contact with metal parts as this can cause uncontrolled heating of the HF-resection electrode and may result in damage and breakage to the HF-resection electrode tip
Z-2413-2023 · Root cause: Process design · FDA record ↗
Most reported problems
Device problems
- Unintended Electrical Shock1
Patient problems
- Shock1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 10, 2019Unintended Electrical Shock; Unintended Electrical Shock
THE SERVICE CENTER WAS INFORMED THAT DURING UNSPECIFIED PROCEDURE, WHILE USING THE ELECTRODE THE DOCTOR FELT A SHOCK. THERE WAS NO TREATMENT REQUIRED. THE INTENDED PROCEDURE WAS COMPLETED. THERE WAS NO EFFECT TO THE PATIENT.
Patient problems: Shock
MAUDE report 8685488 ↗
Other Endoscope, Ac-Powered And Accessories devices
- Olympus WA22810AOLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories2 recalls8 reports
- Olympus A22201DOLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall7 reports
- Olympus A22205DOLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall5 reports
- Olympus A22207COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall1 report
- Olympus A22221COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall1 report
- Olympus A22222COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall0 reports
- Olympus A22231DOLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall0 reports
- Olympus A22253COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.