Olympus A22021ARecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Outer sheath, 27 Fr., 2 stopcocks, rotatable · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04042761020916), received between Jun 14, 2024 and Jun 26, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical /Electronic Property Problem1
- Fracture1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Internal Organ Perforation1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 26, 2026Fracture
THE CUSTOMER REPORTED A BROKEN TIP DURING AN UNSPECIFIED PROCEDURE INVOLVING AN OLYMPUS OUTER SHEATH. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25634606 ↗MalfunctionSep 17, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT THE ENDOSCOPE SHEATH WAS DAMAGED IN SUCH A WAY IT COULD LEAD TO TRAUMA TO URETHRA RESULTING IN STRICTURES. THE ISSUE WAS FOUND DURING A DEMONSTRATION. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20241664 ↗InjuryJun 14, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THERE WAS A BLADDER PERFORATION. THE PHYSICIAN FELT AS IF IT WAS DUE TO AN ELECTRICAL PROBLEM WITH THE RESECTOSCOPE. THE PATIENT HAD TO UNDERGO A LAPAROTOMY DUE TO THE PERFORATION. ADDITIONAL INFORMATION WAS REQUESTED MULTIPLE TIMES, BUT NOT RECEIVED.
Patient problems: Internal Organ Perforation
MAUDE report 19538647 ↗
Other Endoscope, Ac-Powered And Accessories devices
- Olympus WA22810AOLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories2 recalls8 reports
- Olympus A22201DOLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall7 reports
- Olympus A22205DOLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall5 reports
- Olympus A22257COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall1 report
- Olympus A22207COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall1 report
- Olympus A22221COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall1 report
- Olympus A22222COLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall0 reports
- Olympus A22231DOLYMPUS Winter & Ibe GmbH · Endoscope, Ac-Powered And Accessories1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.