Skip to content

Olympus A22043ARecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Resection sheath, 28 Fr. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04042761021005), received between Jun 24, 2021 and Oct 17, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 11, 2025Open, Classified

    Complaints of the ceramic tip of the resection sheath breaking have been received.

    Z-1445-2026 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Break2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Burn(s)1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 17, 2024Break; Break

    IT WAS REPORTED THE CERAMIC TIP OF THE RIGID ENDOSCOPE SHEATH WAS BROKEN. IT IS UNKNOWN WHEN THE ISSUE OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20469343 ↗
  • MalfunctionSep 12, 2023Break; Break

    THE CUSTOMER REPORTED THAT THE CERAMIC TIPS ON SEVERAL OF HIS OLYMPUS RESECTION SHEATHS WERE BREAKING DURING PROCEDURE PREPARATION. SPECIFIC DETAILS REGARDING THESE EVENTS WERE REQUESTED, BUT NO ANSWERS HAVE BEEN PROVIDED AT THIS TIME. THERE WAS NO PATIENT HARM REPORTED AS A RESULT OF THESE EVENTS. THIS REPORT IS RELATED TO REPORTS WITH PATIENT IDENTIFIERS (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17729444 ↗
  • InjuryJun 24, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    UPDATE: THE CUSTOMER CLARIFIED 02JUL2021 THAT THE CUSTOMER DOES NOT USE AN OLYMPUS GENERATOR OR OLYMPUS GROUNDING PADS. THE BURN EXPERIENCED BY THE PATIENT IS NOT RELATED TO THE USE OF AN OLYMPUS DEVICE.

    Patient problems: Burn(s)

    MAUDE report 12063800 ↗

Other Resectoscope devices

All Resectoscope devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page